journal article·2024-03-08Open access

Development of a Smartphone-linked Optical Biosensing System for Point-of-care Testing Applications

Miku Sarubo, Yuka Numazaki, Hiroyuki Kudo

Sensors and Materials· Volume 36· Issue 3

Abstract

We report a miniaturized batteryless colorimetric biosensing system for measuring saliva components.Our system utilizes a CMOS image sensor and an LED light source embedded in smartphones as a part of the optical system.We measured the concentration of chemical substances by observing color changes of the coloring reagent DA-64, which is specific for hydrogen peroxide.The concentration calculated from the RGB scores of colorimetric images showed strong correlations (r > 0.95) with the concentrations of the samples.This result shows the possibility of biochemical measurements using miniaturized biosensing systems to determine physiological conditions.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Biosensors and Analytical Detection

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References

Reference metadata was not supplied for this record.