journal article·2018-12-01

Application of magneto-rheological fluids for generating a wide range of radial pulse waveforms

Jeong-Hoi Koo, Takdanai Tantiyartyanontha, Young-Min Kim, Heeshin Kang, Tae-Heon Yang

Smart Materials and Structures· Volume 27· Issue 12
Publisher page

Abstract

Abstract To effectively study radial pulses for various applications, there exists a tremendous need for a pulsatile simulator that can generate various radial pulse waveforms cost-effectively and reliably. Such a radial simulator can make significant contributions to the research and development efforts in estimating central blood pressures, validating computational work on arterial pulse mechanics, developing wearable devices, and modernizing pulse diagnosis. The primary goal of this study is to develop a pulsatile simulator that can generate a wide range of radial pulse waveforms. To develop an effective and well-controlled simulator, it proposes to use Magneto-Rheological (MR) fluids whose motion can be controlled instantly based on the applied magnetic field strength. The MR pulsatile simulator is designed to produce various ‘two-peak’ pulse waveform patterns and control the shape of waveforms by adjusting the ratio of the peak amplitudes and the distance or the time delay between the two peaks (effectively changing the augmentation index, AI). To this end, an experimental setup consists of a peristaltic pump, a tubing connected to the MR fluid reservoir, and electromagnets with control electronics is constructed. The Pulse Width Modulation (PWM) technique is employed to control magnetic fields that regulates the MR fluids flow. Using the prototype, a series of experiments are performed to evaluate its effectiveness in generating various radial pulse patterns and controlling the pulse waveforms. The results show that the MR simulator is capable of generating actual human radial pulse waveforms for different ages. The results further show that the MR simulator changed the amplitude of the systolic pressure peak and the time delay of the peaks according to the input PWM signals, demonstrating its controllability for the AI of pressure waveforms.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Cardiovascular Health and Disease Prevention · Cardiovascular Function and Risk Factors · Heart Rate Variability and Autonomic Control

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (24)

Show source-provided references
1. Central blood pressure: current evidence and clinical importance10.1093/eurheartj/eht565 ↗
2. Non-invasive methods and techniques for central blood pressure estimation: procedures, validation, reproducibility and limitations10.2174/138161209787354203 ↗
3. Clinical utility of aortic pulses and pressures calculated from applanated radial-artery pulses10.1161/01.HYP.0000084051.34269.A9 ↗
4. Validation of carotid artery tonometry as a means of estimating augmentation index of ascending aortic pressure10.1161/01.HYP.27.2.168 ↗
5. Influence of ventricular ejection on the relationship between central aortic and brachial pressure pulse in man10.1093/cvr/4.3.291 ↗
6. Expert consensus document on arterial stiffness: methodological issues and clinical applications10.1093/eurheartj/ehl254 ↗
7. Arterial stiffness: a brief review10.1038/aps.2010.123 ↗
8. A high performance pulsatile pump for aortic flow experiments in 3-dimensional models10.1007/s13239-016-0260-3 ↗
9. Radial artery pulse wave simulator using a linear motor10.1109/IEMBS.2008.4650311 ↗
10. Harvard Apparatus
11. KYOTO KAGAKU
12. Tellyes Scientific Co.
13. BDC Laboratories
14. ViVitro Labs Inc.
15. Relationship between radial and central arterial pulse wave and evaluation of central aortic pressure using the radial arterial pulse wave10.1291/hypres.30.219 ↗
16. Radial Augmentation Index: A Useful and Easily Obtainable Parameter for Vascular Aging10.1016/j.amjhyper.2004.10.010 ↗
17. An analysis of the relationship between central aortic and peripheral upper limb pressure waves in man10.1093/eurheartj/14.2.160 ↗
18. Pulse wave analysis and pulse wave velocity—a review of blood pressure interpretation 100 years after korotkov10.1253/circj.70.1231 ↗
19. Arterial stiffness and central blood pressure, as determined by pulse wave analysis, in rheumatoid arthritis10.1136/ard.62.5.414 ↗
20. Pressure wave reflection assessed from the peripheral pulse is a transfer function necessary?10.1161/01.HYP.0000057574.64076.A5 ↗
21. Peaceman · 1977 · 6
22. A review on the magnetorheological fluid preparation and stabilization10.1016/j.jmmm.2014.09.020 ↗
23. Compressibility and thermal expansion of petroleum oils in the range 0 degrees to 300 °C10.6028/jres.005.062 ↗
24. Noninvasive determination of age-related changes in the human arterial pulse10.1161/01.CIR.80.6.1652 ↗