journal article·2010-06

Reduced tribocorrosion of CoCr alloys in simulated body fluid after nitrogen insertion

J. Lutz, S. Mändl

Surface and Coatings Technology· Volume 204· Issue 18-19
Publisher page

Abstract

Tribological studies in air and simulated body fluid (SBF) after surface modification by nitrogen plasma immersion ion implantation of medical CoCr alloys were compared. The PIII treatment temperatures were varied between 390 and 570 °C. Subsequent wear measurements in air exhibit reduced wear rates for the PIII treated specimens. In SBF, a minimum wear rate was observed for a process temperature near 400 °C. For higher PIII temperatures, fretting corrosion leads to an elevated cobalt ion release. Apparently a weakened chemical bonding of cobalt as well as chromium immobilization caused by the nitrogen insertion and a subsequent reduction of passivating oxide layer causes the selective high Co ion release for samples with high process temperatures. A compromise for the PIII treatment temperature must be found as the minimum particle release rate does not correspond to the minimum Co ion release rate.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Advanced materials and composites · Metal and Thin Film Mechanics

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (21)

Show source-provided references
1. Wintermantel · 2002
2. Sargeant · Mat. Des. · 2007 · 28 · 15510.1016/j.matdes.2005.05.018 ↗
3. Webster · Biomat. · 2004 · 25 · 473110.1016/j.biomaterials.2003.12.002 ↗
4. Long · Toxicol. Pathol. · 2008 · 36 · 8510.1177/0192623307310951 ↗
5. Çelik · Surf. Coat. Technol. · 2008 · 202 · 243310.1016/j.surfcoat.2007.08.030 ↗
6. Chen · Wear · 2008 · 264 · 15710.1016/j.wear.2006.12.012 ↗
7. Wei · Surf. Coat. Technol. · 2004 · 186 · 30510.1016/j.surfcoat.2004.02.052 ↗
8. Eichentopf · Plasma Proc. Polymers · 2007 · 4 · 4410.1002/ppap.200730304 ↗
9. Lutz · Phys. Stat. Sol. (A) · 2008 · 205 · 98010.1002/pssa.200778329 ↗
10. Lutz · Plasma Proc. Polymers · 2009
11. Diaz · Nucl. Instrum. Meth. B · 2009 · 267 · 163010.1016/j.nimb.2009.01.118 ↗
12. Öztürk · Surf. Coat. Technol. · 2006 · 200 · 568710.1016/j.surfcoat.2005.08.113 ↗
13. [13] J. Lutz, Diploma thesis, Universität Leipzig, 2006.
14. Kokubo · Biomater. · 2006 · 27 · 290710.1016/j.biomaterials.2006.01.017 ↗
15. [15] D. Manova, J. Lutz, and S. Mändl, Sputtering Effects During Plasma Immersion Ion Implantation of Metals, these proceedings.
16. Büscher · J. Biomed. Mater. Res. Part B · 2005 · 72B · 20610.1002/jbm.b.30132 ↗
17. Taylor · J. Mater. Sci. · 1983 · 18 · 326510.1007/BF00544151 ↗
18. Messer · Biomaterials · 1999 · 20 · 164710.1016/S0142-9612(99)00044-7 ↗
19. Covington · J. Prosthet. Dent. · 1985 · 54 · 12710.1016/S0022-3913(85)80087-1 ↗
20. Christiansen · Metall. Mater. Trans. A · 2006 · 37 · 67510.1007/s11661-006-0039-5 ↗
21. Milošev · Electrochim. Acta · 2003 · 48 · 276710.1016/S0013-4686(03)00396-7 ↗