journal article·2004-01

Tensile Performance of Nonsterile Suture Monofilaments Affected by Test Conditions

A.G. Heward, R.M. Laing, D.J. Carr, B.E. Niven

Textile Research Journal· Volume 74· Issue 1
Publisher page

Abstract

Selection of a suture material typically depends on its tensile properties, but in the absence of an international standard test method, a comparison of properties is problem atic. A series of experiments is conducted to determine the tensile performance of non-knotted and knotted non-sterile suture monofilaments, and the effect of test conditions on this performance. Using the ear and cut loop technique, the factorial experiment includes three gauge lengths, three crosshead speeds, three non-sterile suture monofila ments, and five knot configurations. Each combination of variables is replicated five times. Different test conditions affect the ultimate strain and tenacity of both knotted and non-knotted monofilaments. Ultimate strain values are lower with increasing gauge length (when tested at the same crosshead speed) and crosshead speed (when tested at the same gauge length). All monofilaments are characterized by higher tenacity values with in creasing gauge length (at the same crosshead speed) and lower tenacity values with increasing crosshead speed (at the same gauge length). Load-extension curves and visual images of selected monofilament sutures are provided.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Surgical Sutures and Adhesives · Orthopedic Surgery and Rehabilitation · Engineering Technology and Methodologies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

University of Otago, Dunedin, New, Zealand

References (26)

Show source-provided references
1. ADInstruments Pty. Ltd., PowerLab Chart for Windows, Version 3.4, Australia, 1999.
2. 10.1177/00405175010710051110.1177/004051750107100511 ↗
3. Brouwers, J.E. · Surg. Gynecol. Obstet. · 1991 · 173 · 443
4. Ethicon Incorporated, Ethicon Knot Tying Manual, 2000.
5. 10.1302/0301-620X.76B3.817583610.1302/0301-620X.76B3.8175836 ↗
6. Herrmann, J.B. · Surgeon · 1971 · 37 · 209
7. Holmlund, D.E.W. · Acta Chir. Scand. · 1974 · 140 · 355
8. 10.1053/jvet.1999.026010.1053/jvet.1999.0260 ↗
9. International Organization for Standardization, ISO 139: 1973; BS EN 20139:1992. Textiles-Standard atmospheres for conditioning and testing, Geneva, Switzerland. 1992.
10. 10.1016/0010-4361(85)90335-010.1016/0010-4361(85)90335-0 ↗
11. 10.1002/pc.75006011010.1002/pc.750060110 ↗
12. Lewis, D.D. · Vet. J. · 1997 · 75 · 890
13. 10.1016/S0736-4679(97)00081-410.1016/S0736-4679(97)00081-4 ↗
14. 10.1177/03635465970250012210.1177/036354659702500122 ↗
15. 10.1007/BF0098075510.1007/BF00980755 ↗
16. 10.3109/0284431820900657610.3109/02844318209006576 ↗
17. 10.1016/S0736-4679(98)00059-610.1016/S0736-4679(98)00059-6 ↗
18. 10.1111/j.1532-950X.1989.tb01083.x10.1111/j.1532-950X.1989.tb01083.x ↗
19. 10.1111/j.1532-950X.1993.tb00383.x10.1111/j.1532-950X.1993.tb00383.x ↗
20. SPSS Inc. · 2001
21. Tera, H. · Acta Chirurgica Scandinavica · 1976 · 142 · 1
22. Tera, H. · Acta Chirurgica Scandinavica · 1976 · 142 · 301
23. 10.1016/0002-9610(75)90408-010.1016/0002-9610(75)90408-0 ↗
24. The United States Pharmacopoeia, The Pharmacopoeia of the United States of America, NY, 1965.
25. Tukey, J.W. · 1977
26. 10.1016/0736-4679(91)90313-510.1016/0736-4679(91)90313-5 ↗