Implementation of the continuous expansion method with a variable conductance in the pressure range 1 × 10−6 to 1 × 10−3 Pa
Source abstract
Abstract
Abstract The continuous expansion method is widely used by National metrology institutes for traceability of vacuum measurements below 10−3 Pa. It is usually implemented using a fixed conductance such as a small circular orifice. This article describes how LCM-LNE has used the method with a variable conductance and includes a detailed uncertainty budget.