journal article·2009-06

Differences in wear between load and displacement control tested total knee replacements

T. Schwenke, D. Orozco, E. Schneider, M.A. Wimmer

Wear· Volume 267· Issue 5-8
Publisher page

Abstract

Abstract Two identical sets of posterior cruciate retaining total knee replacements were tested for wear on one contemporary knee joint simulator according to the current ISO standards. The first set of implants was tested in load control mode (ISO 14243-1), while the second set was tested in displacement control mode (ISO 14243-3). The type of control mode alludes to the feedback signal of anterior–posterior and internal–external degrees of freedom. Resulting wear scars and rates between the different control mode tests were compared. Significant differences in wear rates were found: the wear rates in displacement control were approximately half of the wear rates in load control mode. This result was mirrored in the wear scars, which were significantly different in size on the medial plateau. Comparing wear rate with wear scar size (area) yielded a positive correlation. Analyzing the motion data of the two tests revealed similar ranges of movement in anterior–posterior translation and internal–external rotation, however the time-scale of the rotational movements differed between the two control types, explaining the differences in wear. This study demonstrated that substantial variability in wear outcome is possible depending on the type of control mode. Future revisions of testing directions should incorporate more implant specific test parameters to produce clinically relevant wear results.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Total Knee Arthroplasty Outcomes · Orthopaedic implants and arthroplasty · Tribology and Wear Analysis

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (15)

Show source-provided references
1. The future burden of hip and knee revisions: US projections from 2005 to 2030
2. The swedish knee arthroplasty register 1975–1997: an update with special emphasis on 41,223 knees operated on in 1988–199710.1080/000164701753532853 ↗
3. ISO 14243-1, Implants for surgery—wear of total knee-joint prostheses. Part 1. Loading and displacement parameters for wear-testing machines with load control and corresponding environmental conditions for test, 2002.
4. ISO 14243-2, Implants for surgery—wear of total knee joint prosthesis. Part 3. Loading and displacement parameters for wear testing machines with displacement control and corresponding environmental conditions for tests, 2001.
5. Stresses in polyethylene components of contemporary total knee replacements
6. The effects of conformity and load in total knee replacement10.1097/00003086-200006000-00036 ↗
7. Finite element analysis of TKR captures device-dependent kinematics and wear potential
8. Computer model to predict subsurface damage in tibial inserts of total knees10.1002/jor.1100160507 ↗
9. Analysis and recommendations for the optimum spring configurations for soft tissue restraint in force-control knee simulator testing
10. ISO 14243-2, Implants for surgery—wear of total knee-joint prostheses. Part 2. Methods of measurement, 2000.
11. Comparison of wear in a total knee replacement under different kinematic conditions10.1023/A:1012894023657 ↗
12. Comparison between force-controlled in-vivo wear testing on a widely used TKR implant
13. Differences in damage between revision and postmortem retrieved TKA implants
14. Effect of stair descent loading on ultra-high molecular weight polyethylene wear in a force-controlled knee simulator10.1243/095441102321032201 ↗
15. Implementation of multiple activities of daily living for knee wear testing