journal article·2010-02

Wear patterns of, and wear volume formulae for, hemispherical acetabular cup liners

James Shih-Shyn Wu, Shu-Ling Hsu, Jian-Horng Chen

Wear· Volume 268· Issue 3-4
Publisher page

Abstract

Abstract This study analyzed wear patterns of, and wear volume formulae for, hemispherical acetabular cup liners used in total hip replacement. The geometric patterns of the wear surface were first classified, then wear volume formulae were derived by integral calculus. Based on the theoretical maximal and minimal wear volumes, three-dimensional wear models constructed using SolidWorks® software or published formulae were used to verify the accuracy of the formulae proposed in this study. Analytical results showed that, with the exception of the pattern of a wear angle of 90°, hemispherical liner wear can be divided into two patterns. Once the wear pattern is determined from the wear angle and depth, a wear formula is applied to calculate the wear volume. The formulae and wear patterns described herein are crucial for more accurate performance evaluation of existing hip components implanted in patients, as well as for designing new hip components.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Advanced materials and composites · Orthopedic Infections and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (41)

Show source-provided references
1. Cemented revision for aseptic acetabular loosening: a review of 387 hips10.1302/0301-620X.77B3.7744914 ↗
2. Total hip arthroplastic patients younger than 45 years. A nine to ten-year follow-up study10.1097/00003086-199011000-00036 ↗
3. Total hip arthoplasty in patients with osteonecrosis10.1097/00003086-199705000-00014 ↗
4. The mechanism of loosening of cemented acetabular components in total hip arthroplasty. Analysis of specimens retrieved at autopsy10.1097/00003086-199201000-00009 ↗
5. Mechanism and clinical significance of wear debris-induced osteolysis10.1097/00003086-199203000-00003 ↗
6. The problem is osteolysis
7. Polyethylene wear in cementless total hip arthroplasty10.1302/0301-620X.79B3.0790361 ↗
8. A literature review of the association between wear rate and osteolysis in total hip arthroplasty10.1054/arth.2002.33664 ↗
9. Relationship of acetabular wear to osteolysis and loosening in total hip arthroplasty10.1097/00003086-199906000-00018 ↗
10. The potential role of phagocytic capacity in the osteolytic process induced by polyethylene wear particles10.1177/147323000603400611 ↗
11. Biological reactions to wear debris in total joint replacement10.1243/0954411001535219 ↗
12. Effect of femoral head size on wear of the polyethylene acetabular component10.2106/00004623-199072040-00008 ↗
13. Polyethylene wear in uncemented acetabular components
14. Measurement accuracy in acetabular cup wear. Three retrospective methods compared with Roentgen stereophotogrammetry10.1016/S0883-5403(05)80208-6 ↗
15. The effect of joint loading on acetabular wear measurement in total hip arthroplasty10.1054/arth.2000.4336 ↗
16. Measurement of polyethylene wear: a new three-dimensional methodology10.1016/S0169-2607(01)00164-X ↗
17. The optimum size of prosthetic heads in relation to the wear of plastic sockets in total replacement of the hip10.1007/BF02474667 ↗
18. In vivo wear of polyethylene acetabular components10.1302/0301-620X.75B2.8444946 ↗
19. Measurement of wear in retrieved acetabular sockets10.1243/PIME_PROC_1995_209_350_02 ↗
20. The estimation of acetabular cup wear volume from two-dimensional measurements: a comprehensive analysis10.1243/0954411981534060 ↗
21. Polyethylene wear in total hip prostheses: the influence of direction of linear wear on volumetric wear determined from radiographic data10.1007/s00256-003-0685-2 ↗
22. Validation of acetabular cup wear volume based on direct and two-dimensional measurements: hip simulator analysis10.1007/s00776-003-0674-2 ↗
23. Estimation of the wear volume after total hip replacement. A simple access to geometrical concepts10.1016/j.medengphy.2007.04.003 ↗
24. Polyethylene wear in total hip arthroplasty: volumetric wear measurement of retrieved acetabular components
25. The origin of submicron polyethylene wear debris in total hip arthroplasty
26. Wear of polyethylene cups in total hip arthroplasty. A study of specimens retrieved post mortem10.2106/00004623-199608000-00009 ↗
27. Dimensional changes in metal-backed polyethylene acetabular cups under cyclic loading
28. Finite element simulation of early creep and wear in total hip arthroplasty10.1016/j.jbiomech.2004.10.022 ↗
29. An in vivo determination of total hip arthroplasty pistoning during activity10.1054/arth.2000.6637 ↗
30. In vivo” determination of hip joint separation and the forces generated due to impact loading conditions10.1016/S0021-9290(00)00239-6 ↗
31. Long-term wear of HIPed alumina on alumina bearings for THR under microseparation conditions10.1023/A:1012802308636 ↗
32. Analysis of micro-separation and contact mechanics between the femoral head and the acetabular cup in artificial hip joint replacements
33. Analysis of wear and oxidation on retrieved bipolar polyethylene liner10.1007/s00776-008-1233-7 ↗
34. The measurement of creep in ultrahigh molecular weight polyethylene10.1016/j.arth.2004.08.008 ↗
35. Dimensional changes in metal-backed polyethylene acetabular cups under cyclic loading
36. Effects of the sterilization method on the wear of UHMWPE acetabular cups tested in a hip simulator10.1016/S0142-9612(01)00265-4 ↗
37. Validation of volumetric and linear wear-measurement in UHMWPE cups—a hip simulator analysis10.1016/S0043-1648(02)00064-9 ↗
38. The effects of irradiation and EtO-treatment on ultrahigh molecular weight polyethylene acetabular cups following accelerated aging: degradation of mechanical properties and morphology changes during hip simulator tests10.1016/j.molstruc.2007.04.042 ↗
39. Influence of polyethylene creep behavior on wear in total hip arthroplasty10.1002/jor.20042 ↗
40. Dioscorea improves the morphometric and mechanical properties of bone in ovariectomised rat10.1002/jsfa.3396 ↗
41. Wear of polyethylene acetabular components in total hip arthroplasty10.2106/00004623-199703000-00005 ↗