Quantifying the wear of acetabular cups using coordinate metrology
Source abstract
Abstract
Abstract In an ageing population where individuals demand a pain free and active existence, hip arthroplasty has become one of the most important operations of later life, with developments increasing the longevity of this prosthesis. Accurate assessment of wear is an important way to assess how well hip prostheses perform over time. This paper describes a new protocol to accurately locate and map the wear of acetabular cups. 16 metal-on-metal acetabular cups were tested in a hip simulator for up to 6 million cycles at a range of angles of inclination in order to create a large range of wear volumes. Unworn and worn regions of acetabular cups were digitised in two separate scans using a CMM. The two scans were superimposed in order to allow calculation of maximum depth, wear volume loss and rim damage. Values obtained from the CMM protocol (CMM) were compared to the wear volume determined gravimetrically (G); a strong correlation (CMM = 0.992G−0.504) was established. The CMM protocol established was shown to be a methodology more powerful than many existing geometric protocols which could be used to assess explants and implants tested in vitro.
Source: openaire · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Hip disorders and treatments · Orthopedic Infections and Treatments
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (28)
Show source-provided references
1. Outcomes after hip or knee replacement surgery for osteoarthritis. A prospective cohort study comparing patients’ quality of life before and after surgery with age-related population norms10.5694/j.1326-5377.1999.tb123628.x ↗
2. Replacement of arthritic hips by the McKee–Farrar prosthesis10.1302/0301-620X.48B2.245 ↗
3. Young adults with arthritic hips10.1136/bmj.315.7103.265 ↗
4. Pseudotumours associated with metal-on-metal hip resurfacings10.1302/0301-620X.90B7.20213 ↗
5. National Joint Registry for England and Wales. 6th Annual Report, 2009.
6. Department of Health. Hospital Episode Statistics, 2008. http://hesonline.nhs.uk (accessed 3.2.10).
7. ‘Severe’ wear challenge to ‘as-cast’ and ‘double heat-treated’ large-diameter metal-on-metal hip bearings10.1243/09544119JEIM81 ↗
8. Design considerations and life prediction of metal-on-metal bearings: the effect of clearance10.2106/JBJS.H.00610 ↗
9. Development and problems of metal-on-metal hip arthroplasty10.1243/095441105X63264 ↗
10. Metal ion release from bearing wear and corrosion with 28mm and large-diameter metal-on-metal bearing articulations: a follow-up study10.1302/0301-620X.92B1.22226 ↗
11. Revision Arthroplasty: an update10.1007/s00256-009-0781-z ↗
12. The wear pattern in metal-on-metal hip prostheses10.1002/jbm.1068 ↗
13. A hip joint simulator study of the performance of metal-on-metal joints: part II: design
14. Investigation on stick phenomena in metal-on-metal hip joints after resting periods
15. The effects of proteins on the friction and lubrication of artificial joints10.1243/09544119JEIM21 ↗
16. R.J. Underwood, P.M. Cann, A. Hart, K. Ilo, J. Skinner, C.R. Wagner. Development of protocol for measuring wear in explanted hip joints using roundness machine. 56th Annual Meeting of the Orthopaedic Research Society, Poster No. 2322.
17. Wear behaviour and histopathology of classic cemented metal on metal hip endoprostheses10.1097/00003086-199608001-00016 ↗
18. X.Q. Hu, M. Tuke, A. Taylor. The wear measurement of the hip joint – a geometrical methodology. 53rd Annual Meeting of the Orthopaedic Research Society, Poster No: 1646.
19. In vivo wear of three types of metal on metal hip prothesis during 2 decades of use10.1097/00003086-199608001-00013 ↗
20. Tribology and total hip joint replacement: current concepts in mechanical simulation10.1016/j.medengphy.2008.07.006 ↗
21. 3D linear and volumetric wear measurement on artificial hip joints—Validation of a new methodology10.1016/j.precisioneng.2010.06.001 ↗
22. Cobalt chromium molybdenum metal combination for modular hip prostheses10.1097/00003086-199608001-00004 ↗
23. Geometrics of tooth wear10.1016/j.wear.2008.04.062 ↗
24. Ceramic bearing surfaces10.1097/00003086-199912000-00009 ↗
25. Becker A, Schollhorn K, Dirix Y, Schomtzer H. Correlation between linear wear and volumetric waer in vitro and in vivo. 52nd Annual Meeting of the Orthopaedic Research Society, Paper No: 5054.
26. Becker A, Schollhorn K, Dirix Y, Schomtzer H. The influence of 3D measurement accuracy on the calculated wear of a ball head using a new mathematical approach. 52nd Annual Meeting of the Orthopaedic Research Society, Paper No: 5050.
27. A metrology solution for the orthopaedic industry. 7th International Symposium on Measurement Technology and Intelligent Instruments
28. Development of a technique for accurately determining clinical wear in explanted total hip replacements10.1016/j.wear.2007.01.054 ↗