journal article·2010-07

Evaluation of the stress distribution in CFR-PEEK dental implants by the three-dimensional finite element method

João Rodrigo Sarot, Cintia Mussi Milani Contar, Ariadne Cristiane Cabral da Cruz, Ricardo de Souza Magini

Publisher page

Abstract

CFR-PEEK (carbon fiber reforced-poly ether ether ketone) has been demonstrated to be excellent substitute titanium in orthopedic applications and can be manufactured with many physical, mechanical, and surface properties, in several shapes. The aim of this study was to compare, using the three-dimensional finite element method, the stress distribution in the peri-implant support bone of distinct models composed of PEEK components and implants reinforced with 30% carbon fiber (30% CFR-PEEK) or titanium. In simulations with a perfect bonding between the bone and the implant, the 30% CFR-PEEK presented higher stress concentration in the implant neck and the adjacent bone, due to the decreased stiffness and higher deformation in relation to the titanium. However, 30% CFR-PEEK implants and components did not exhibit any advantages in relation to the stress distribution compared to the titanium implants and components.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Dental Implant Techniques and Outcomes · Dental materials and restorations

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (34)

Show source-provided references
1. Cook SD, Rust-Dawicki AM. Preliminary evaluation of titanium coated PEEK dental implants. J Oral Implantol. 1995;21:176–81.
2. Fujihara K, Huang ZM, Ramakrishna S, Satknanantham K, Hamada H. Performance study of braided carbon/PEEK composite compression bone plates. Biomaterials. 2003;24:2661–7.10.1016/S0142-9612(03)00065-6 ↗
3. Fujihara K, Huang ZM, Ramakrishna S, Satknanantham K, Hamada H. Feasibility of knitted carbon/PEEK composites for orthopedic bone plates. Biomaterials. 2004;25:3877–85.10.1016/j.biomaterials.2003.10.050 ↗
4. Skinner HB. Composite technology for total hip arthroplasty. Clin Orthop Relat Res. 1988;235:224–36.10.1097/00003086-198810000-00022 ↗
5. Wenz LM, Merritt K, Brown SA, Moet A, Steffee AD. In vitro biocompatibility of polyetheretherketone and polysulfone composites. J Biomed Mater Res. 1990;24:207–15.10.1002/jbm.820240207 ↗
6. Hunter A, Archer CW, Walker PS, Blunn GW. Attachment and proliferation of osteoblasts and fibroblasts on biomaterials for orthopaedic use. Biomaterials. 1995;16:287–95.10.1016/0142-9612(95)93256-D ↗
7. Morrison C, Macnair R, MacDonald C, Wykman A, Goldie I, Grant MH. In vitro biocompatibility testing of polymers for orthopaedic implants using cultured fibroblasts and osteoblasts. Biomaterials. 1995;16:987–92.10.1016/0142-9612(95)94906-2 ↗
8. Lin TW, Corvelli AA, Frondoza CG, Roberts JC, Hungerford DS. Glass PEEK composite promotes proliferation and osteocalcin production of human osteoblastic cells. J Biomed Mater Res. 1997;36:137–44.10.1002/(SICI)1097-4636(199708)36:2<137::AID-JBM1>3.0.CO;2-L ↗
9. Katzer A, Marquardt H, Westendorf J, Wening JV, von Foerster G. Polyetheretherketone—cytotoxicity and mutagenicity in vitro. Biomaterials. 2002;23:1749–59.10.1016/S0142-9612(01)00300-3 ↗
10. Scotchford CA, Garle MJ, Batchelor J, Bradley J, Grant DM. Use of a novel carbon fibre composite material for the femoral stem component of a THR system: in vitro biological assessment. Biomaterials. 2003;24:4871–9.10.1016/S0142-9612(02)00488-X ↗
11. Sheets CG, Earthmann JC. Natural intrusion and reversal in implant assisted prosthesis; evidence of and a hypothesis for the occurrence. J Prosthet Dent. 1993;70:513–20.10.1016/0022-3913(93)90265-P ↗
12. Anusavice KJ. Phillips’ science of dental materials. 11th ed. St Louis: Sanders; 2003.
13. Tillitson EW, Craig RG, Peyton FA. Friction and wear of restorative dental materials. J Dent Res. 1971;50:149–54.10.1177/00220345710500011001 ↗
14. Kitamura E, Stegaroiu R, Nomura S, Miyakawa O. Biomechanical aspects of marginal bone resorption around osseointegrated implants: considerations based on a three-dimensional finite element analysis. Clin Oral Implants Res. 2004;15:401–12.10.1111/j.1600-0501.2004.01022.x ↗
15. Kurtz SM, Devine JN. PEEK biomaterials in trauma, orthopedic and spinal implants. Biomaterials. 2007;28:4845–69.10.1016/j.biomaterials.2007.07.013 ↗
16. Toth JM, Wang M, Estes BT, Scifert JL, Seim HB, Turner AS. Polyetheretherketone as a biomaterial for spinal applications. Biomaterials. 2006;27:324–34.10.1016/j.biomaterials.2005.07.011 ↗
17. Brantigan JW, Neidre A, Toohey JS. The Lumbar I/F Cage for posterior lumbar interbody fusion with the variable screw placement system: 10-year results of a Food and Drug Administration clinical trial. Spine J. 2004;4:681–8.10.1016/j.spinee.2004.05.253 ↗
18. Akhavan S, Matthiesen MM, Schulte L, Penoyar T, Kraay MJ, Rimnac CM, Goldberg VM. Clinical and histologic results related to a lowmodulus composite total hip replacement stem. J Bone Joint Surg Am. 2006;88:1308–14.10.2106/JBJS.E.00316 ↗
19. Isidor F. Influence of forces on peri-implant bone. Clin Oral Implants Res. 2006;17(Suppl 2):8–18.10.1111/j.1600-0501.2006.01360.x ↗
20. Williams KR, Edmundson JT, Rees JS. Finite element stress analysis of restored teeth. Dent Mater. 1987;3:200–6.10.1016/S0109-5641(87)80034-9 ↗
21. Misch CE, Suzuki JB, Misch-Dietsh FM, Bidez MW. A positive correlation between occlusal trauma and peri-implant bone loss: literature support. Implant Dent. 2005;14:108–16.10.1097/01.id.0000165033.34294.db ↗
22. Hobkirk JA and Wiskott HW. Biomechanical aspects of oral implants: consensus report of Working Group I. Clin Oral Implants Res. 2006;17(Suppl 2):52–4.10.1111/j.1600-0501.2006.01358.x ↗
23. Petrie CS, Williams JL. Comparative evaluation of implant designs: influence of diameter, length, and taper on strains in the alveolar crest. A three-dimensional finite element analysis. Clin Oral Implants Res. 2005;16:486–94.10.1111/j.1600-0501.2005.01132.x ↗
24. Kitamura E, Stegaroiu R, Nomura S, Miyakawa O. Influence of marginal bone resorption on stress around an implant–a three dimensional finite element analysis. J Oral Rehabil. 2005;32:279–86.10.1111/j.1365-2842.2004.01413.x ↗
25. Bidez MW, Misch CE. Issues in bone mechanics related to oral implants. Implant Dent. 1992;1:289–94.10.1097/00008505-199200140-00011 ↗
26. Bozkaya D, Muftu S, Muftu A. Evaluation of load transfer characteristics of five different implants in compact bone at different load levels by finite elements analysis. J Prosthet Dent. 2004;92:523–30.10.1016/j.prosdent.2004.07.024 ↗
27. Siegele D, Soltesz U. Numerical investigations of the influence of implant shape on stress distribution in the jaw bone. Int J Oral Maxillofac Implants. 1989;4:333–40.
28. Geng JP, Q. Ma QS, Xu W, Tan KB, Liu GR. Finite element analysis of four threat-form configurations in a stepped screw implant. J Oral Rehabil. 2004;31:233–9.10.1046/j.0305-182X.2003.01213.x ↗
29. Meyer U, Vollmer D, Runte C, Bourauel C, Joos U. Bone loading pattern around implants in average and atrophic edentulous maxillae: a finite-element analysis. J Maxillofac Surg. 2001;29:100–5.
30. Koca OL, Eskitascioglu G, Usumez A. Three-dimensional finite-element analysis of functional stresses in different bone locations produced by implants placed in the maxillary posterial region of the sinus floor. J Prosthet Dent. 2005;93:38–44.10.1016/j.prosdent.2004.10.001 ↗
31. Zampelis A, Rangert B, Heijl L. Tilting of splinted implants for improved prosthodontic support: a two-dimensional finite element analysis. J Prosthet Dent. 2007;97:535–43.10.1016/S0022-3913(07)60006-7 ↗
32. Natali AN, Carniel EL, Pavan PG. Investigation of bone inellastic response in interaction phenomena with dental implants. Dent Mater. 2008;24:561–9.10.1016/j.dental.2007.11.024 ↗
33. Cehreli M, Duyck J, De Cooman M, Puers R, Naert I. Implant design and interface force transfer: a photoelastic and strain-gauge analysis. Clin Oral Implants Res. 2004;15:249–57.10.1111/j.1600-0501.2004.00979.x ↗
34. Sutpideler M, Eckert SE, Zobitz M, An KN. Finite element analysis of effect of prosthesis height, angle of force application, and implant offset on supporting bone. Int J Oral Maxillofac Implants. 2004;19:819–25.