Factors affecting the mechanical and viscoelastic properties of acrylic bone cement
Source abstract
Abstract
The aim of this paper is to report a series of experiments investigating the factors that influence the viscoelastic properties of acrylic bone cement. The effects of the brand of cement, the length of time since mixing, temperature, the hydration of the cement, and the influence of fat and or blood in the environment on the creep and stress relaxation behavior of the cement have been studied in laboratory-prepared specimens in tension, compression and four point bending. Although there are significant differences in the viscoelastic behavior of some of the different brands of polymethylmethacrylate based cements, these differences are small by comparison with the major effects that can be exerted by the length of time since mixing and some environmental factors. These effects have important practical consequences, especially with regard to the ability of bench top and theoretical studies to predict reliably the mechanical and viscoelastic behavior of acrylic cement in vivo.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Orthopedic Infections and Treatments
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (34)
Show source-provided references
1. S. Saha and S. Pal, J. Biomed. Mat. Res. 18 (1984) 435-462.10.1002/jbm.820180411 ↗
2. W. Krause and R. Mathis, ibid. 22A1 (1988) 37-53.
3. G. Lewis, ibid. 38 (1997) 155-182.10.1002/(SICI)1097-4636(199722)38:2<155::AID-JBM10>3.0.CO;2-C ↗
4. W. L. Bargar, S. A. Brown, H. A. Paul, T. Voegli, Y. Hseih and N. Sharkey. J. Orthop. Res. 4 (1986) 86-89.10.1002/jor.1100040110 ↗
5. S. A. Brown and W. L. Bargar, J. Biomed. Mat. Res. 18 (1984) 523-536.10.1002/jbm.820180506 ↗
6. R. Huiskes, N. Verdonschot and B. Nivbrant, Clin. Orthop. Rel. Res. 355 (1998) 103-112.10.1097/00003086-199810000-00011 ↗
7. G. Shen, J. Bone Joint Surg. Br. 80-B (1998) 754-756.10.1302/0301-620X.80B5.0800754 ↗
8. N. G. Mccrum, C. P. Buckley and C. B. Bucknall, in “Principles of Polymer Engineering” (Oxford University Press, Oxford, New York, Tokyo, 1997) pp. 117-183.
9. W. R. Moore, in “An Introduction to Polymer Chemistry” (University of London Press Ltd., London, 1963) pp. 66-91.
10. A. J. C. Lee, R. D. Perkins and R. S. M. Ling, in “Implant Bone Interface”, edited by J. Older (Springer-Verlag, London, 1990) pp. 85-90.10.1007/978-1-4471-1811-4_12 ↗
11. N. Verdonschot and R. Huiskes, J. Appl. Biomat. 5 (1994) 235-243.10.1002/jab.770050309 ↗
12. N. Verdonschot and R. Huiskes, ibid. 29 (1995) 575-581.
13. N. Hughes, G. A. Gie, A. J. C. Lee and R. S. M. Ling, J. Bone Joint Surg. (Br.) 79-B (1997) 367.
14. J. Davies, M. J. Anderson and W. H. Harris, Trans. O.R.S. 21 (1996) 525.
15. Z. Lu and H. Mckellop, J. Biomed. Mat Res. 34 (1997) 221-226.10.1002/(SICI)1097-4636(199702)34:2<221::AID-JBM11>3.0.CO;2-L ↗
16. N. Verdonschot and R. Huiskes, J. Bone Joint Surg. (Br.) 79-B (1997) 665-669.10.1302/0301-620X.79B4.0790665 ↗
17. D. A. Thomas and S. Turner, in “Testing of Polymers — Volume 4” edited by W. E. Brown (Wiley Interscience, London, 1969) pp. 73-120.
18. J. Thanner, C. Freij-Larsson, J. Karrholm, H. Malchau and B. Wesslen, Acta Orthop. Scand. 66 (1995) 207-214.10.3109/17453679508995525 ↗
19. S. Suominen, ibid. 66 (1995) 13.10.3109/17453679508994630 ↗
20. P. Riegels-Nielsen, L. SØrensen, H. M. Andersen and S. Lindequist, ibid. 66 (1995) 215-217.10.3109/17453679508995526 ↗
21. L. Linder, ibid. 66 (1995) 205.10.3109/17453679508995524 ↗
22. A. R. Nilsen and M. Wiig, ibid. 67 (1996) 57-59.10.3109/17453679608995610 ↗
23. J. P. Walczak, J. C. D. D'Arcy, K. R. Ross, S. E. James, A. V. Bonnici, S. R. Koka and R. W. Morris, J. Arthroplast. 15 (2000) 205-209.10.1016/S0883-5403(00)90264-X ↗
24. K. F. M. Abdel-Kader, S. Allcock, D. I. Walker and S. B. Chaudhry, ibid. 16 (2001) 811-819.10.1054/arth.2001.25561 ↗
25. M. A. R. Freeman, Personal communication (1999).
26. A. S. Litsky, R. M. Rose, C. T. Rubin and E. L. Thrasher, J. Orth. Res. 8 (1990) 623-626.10.1002/jor.1100080420 ↗
27. O. Furnes, S. A. Lie, L. I. Havelin, S. E. Vollset and L. B. Engesoeter, Acta Orthop. Scand. 68 (1997) 515-520.10.3109/17453679708999017 ↗
28. P. B. Thomsen, S. Bovling, B. Jacoby and T. B. Hansen, Hip Int. 10 (2000) 102-107.10.1177/112070000001000205 ↗
29. K. T. Hustosky, T. L. Norman, V. L. Kish, J. D. Blaha and T. A. Gruen, Trans. O.R.S. 21 (1996) 423.
30. A. J. C. Lee, in “Interfaces in Total Hip Arthroplasty”, edited by I. D. Learmonth (Springer-Verlag, Berlin, 1999) pp. 11-19.
31. J. Charnley, in “Acrylic Cement in Orthopaedic Surgery” (E. & S. Livingstone, Edinburgh and London, 1970) pp. 23-32.
32. M. F. Ashby, D. R. H. Jones, in “Engineering Materials 2” (Pergamon Press, Oxford, 1988) pp. 218-232.
33. G. Bergmann, F. Graichen, A. Rohlmann, N. Verdonschot and G. H. Van Lenthe, J. Biomech. 34 (2001) 421-428.10.1016/S0021-9290(00)00188-3 ↗
34. E. A. Turi, in “Thermal analysis in polymer characterisation” (John Wiley & Sons, New York, 1981)