Determination of urethral catheter surface lubricity
Source abstract
Abstract
Device for in-vitro measurement of static and kinetic friction coefficient of catheter surface was developed. Tribometer was designed and constructed to work with exchangeable counter-faces (polymers, tissue) and various types of tubes, in wet conditions in order to mimic in-vivo process. Thus seven commercially available urethral catheters, made from vinyl polymers, natural latex with silicone coating, all-silicone or hydrogel coated, and one made from polyvinylchloride with polyurethane/polyvinylpyrrolidone hydrogel coating obtained in our laboratory, were tested against three various counter faces: polymethacrylate (organic glass), inner part of porcine aorta and porcine bladder mucosa. Additionally, the hydrophility/hydrophobity of tested catheters was stated via water wetting contact angle measurement. Super-hydrophilic biomaterials revealed low friction on tissue and hydrophobic counter-face; slightly hydrophobic showed higher friction in both cases, while more hydrophobic manifested low friction on tissue but high on hydrophobic polymer. The smoothest friction characteristic was achieved in all cases on tissue counter-faces. The measured values of the static coefficient of friction of catheters on bladder mucosa counter-face were as follows: the highest (0.15) for vinyl and siliconised latex catheters and 3 folds lower (0.05) for all-silicone ones. Hydrogel coated catheters exhibited the lowest static and kinetic friction factors.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Urinary Bladder and Prostate Research · Central Venous Catheters and Hemodialysis · Pelvic floor disorders treatments
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (25)
Show source-provided references
1. D. T. BEIKO, B. E. KNUDSEN, J. D. WATTERSON, P. A. CADIEUX, G. REID and J. D. DENSTEDT, J. Urol. 171 (2004) 243810.1097/01.ju.0000125001.56045.6c ↗
2. E. L. LAWRENCE and I. G. TURNER, Med. Eng. Phys. 27 (2005) 44310.1016/j.medengphy.2004.12.013 ↗
3. B. H. CHEW, M. DUVDEVANI and J. D. DENSTEDT, Expert Rev. Med. Devices 3 (2006) 39510.1586/17434440.3.3.395 ↗
4. J. C. NICKEL, M. E. OLSON and J. W. COSTERTON, Urology 29 (1987) 50110.1016/0090-4295(87)90037-9 ↗
5. A. E. KHOURY, M. E. OLSON, F. VILLARI and J. W. COSTERTON, J. Urol. 145 (1991) 61010.1016/S0022-5347(17)38405-7 ↗
6. ASTM G115–04 Standard guide for measuring and reporting friction coefficients. ASTM International: www.astm.org
7. B. BHUSHAN and M. NOSONOVSKY, Nanotechnology 15 (2004) 74910.1088/0957-4484/15/7/006 ↗
8. B. N. J. PERSSON U. TARTAGLINO E. TOSSATTI and O. ALBOHR, KGK Kautsch. Gummi Kunstst. 57 (2004) 532
9. ASTM F623–99(2006) Standard performance specification for Foley catheter. ASTM International: www.astm.org
10. M. FADER, K. N. MOORE, A. M. COTTENDEN, L. PETTERSSON, R. BROOKS and J. MALONE-LEE, BJU Int. 88 (2001) 37310.1046/j.1464-410X.2001.02342.x ↗
11. N. TOMITA, S. TAMAI, E. OKAJIMA, Y. HIRAO, K. IKEUCHI and Y. IKADA, J. Appl. Biomater. 5 (1994) 17510.1002/jab.770050211 ↗
12. J. STENSBALLE, D. LOOMS, P. N. NIELSEN and M. TVEDE, Eur. Urol. 48 (2005) 97810.1016/j.eururo.2005.07.009 ↗
13. P. RAMESH, R. JOSEPH and M. C. SUNNY, J. Biomater. Appl. 15 (2001) 34410.1106/NCRR-MDV5-FQFG-L7PG ↗
14. D. GRAIVER, R. L. DURALL and T. OKADA, Biomaterials 14 (1993) 46510.1016/0142-9612(93)90150-Z ↗
15. G. MARMIERI, M. PETTENATI, C. CASSINELLI and M. MORRA, J. Biomed. Mater. Res. 33 (1996) 2910.1002/(SICI)1097-4636(199621)33:1<29::AID-JBM5>3.0.CO;2-O ↗
16. M. M. TUNNEY and S. P. GORMAN, Biomaterials 23 (2002) 460110.1016/S0142-9612(02)00206-5 ↗
17. V. A. CONVENEY and D. GROVER, Physiol. Meas. 22 (2001) 50510.1088/0967-3334/22/3/308 ↗
18. J. PARKIN, J. SCANLAN, W. WOOLLEY, D. GROVER, A. EVANS and R. C. L. FENELEY, BJU Int. 90 (2002) 66610.1046/j.1464-410X.2002.03014.x ↗
19. D. S. JONES, C. P. GARVIN and S. P. GORMAN, J. Mater. Sci. Mater. Med. 12 (2001) 1510.1023/A:1026744732504 ↗
20. D. S. JONES, C. P. GARVIN and S. P. GORMAN, Biomaterials 25 (2004) 142110.1016/j.biomaterials.2003.08.011 ↗
21. M. BIEHL, US patent no. 6460397 (2002)
22. M. J. MICKLUS and D. T. OU-YANG, Raritan, US patent no. 4100309 (1978)
23. L. WALLER, M. TELANDER and L. SULLIVAN, Spinal Cord 35 (1997) 22910.1038/sj.sc.3100390 ↗
24. J. LUNDGREN, O. BENGTSSON, A. ISRAELSSON, A. C. JONSSON, A. S. LINDH and J. UTAS, Spinal Cord 38 (2000) 4510.1038/sj.sc.3100935 ↗
25. E. D’ANGELO, S. H. LORING, M. E. GIOIA, M. PECCHIARI and C. MOSCHENI, Respir. Physiol. Neurobiol. 142 (2004) 5510.1016/j.resp.2004.05.006 ↗