journal article·2011-05Open access

Direct visualization and quantification of bone growth into porous titanium implants using micro computed tomography

E. Baril, L. P. Lefebvre, S. A. Hacking

Abstract

The utility of porous metals for the integration of orthopaedic implants with host bone has been well established. Quantification of the tissue response to cementless implants is laborious and time consuming process requiring tissue processing, embedding, sectioning, polishing, imaging and image analysis. Micro-computed tomography (μCT) is a promising three dimensional (3D) imaging technique to quantify the tissue response to porous metals. However, the suitability and effectiveness of μCT for the quantification of bone ingrowth remains unknown. The purpose of this study was to evaluate and compare bone growth within porous titanium implants using both μCT and traditional hard-tissue histology techniques. Cylindrical implants were implanted in the distal femora and proximal tibiae of a rabbit. After 6 weeks, bone ingrowth was quantified and compared by μCT, light microscopy and backscattered electron microscopy. Quantification of bone volume and implant porosity as determined by μCT compared well with data obtained by traditional histology techniques. Analysis of the 3D dataset showed that bone was present in the pores connected with openings larger 9.4 μm. For pore openings greater than 28.2 μm, the size of the interconnection had little impact on the bone density within the porosity for the titanium foams.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Bone Tissue Engineering Materials · Advanced X-ray and CT Imaging

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (29)

Show source-provided references
1. Della Valle CJ, Mesko NW, Quigley L, Rosenberg AG, Jacobs JJ, Galante JO. Primary total hip arthroplasty with a porous-coated acetabular component. J Bone Joint Surg Am. 2009;91(5):1130–5.10.2106/JBJS.H.00168 ↗
2. Levine BR, Sporera S, Poggie RA, Della Vallea CJ, Jacobs JJ. Experimental and clinical performance of porous tantalum in orthopedic surgery. Biomaterials. 2006;27(27):4671–81.10.1016/j.biomaterials.2006.04.041 ↗
3. Sporer S, Paprosky WG. The use of a trabecular metal acetabular component and trabecular metal augment for severe acetabular defects. J Arthroplasty. 2006;21(6-suppl):83–6.10.1016/j.arth.2006.05.008 ↗
4. Ryan G, Pandit A, Apatsidis DP. Fabrication methods of porous metals for use in orthopaedic applications. Biomaterials. 2006;27(13):2651–70.10.1016/j.biomaterials.2005.12.002 ↗
5. Lefebvre LP, Baril E, Bureau MN. Effect of the oxygen content in solution on the static and cyclic deformation of titanium foams. J Mater Sci: Mater Med. 2009;20(11):2223–33.10.1007/s10856-009-3798-x ↗
6. Zhang Y, Ahn PB, Fitzpatrick DC, Heiner AD, Poggie RA, Brown TD. Interfacial frictional behavior: cancellous bone, cortical bone, and a novel porous tantalum biomaterial. J Musculoskeletal Res. 1999;3(4):245–51.10.1142/S0218957799000269 ↗
7. Heiner AD, Brown TD. Frictional coefficients of a new bone ingrowth structure. Trans Orthopaed Res Soc. 2007;32:1623.
8. Boyde A, Jones SJ. Back-scattered electron imaging of skeletal tissues. Metab Bone Dis Relat Res 1983–1984;5(3):145–50.10.1016/0221-8747(83)90016-4 ↗
9. Assad M, Jarzem P, Leroux MA, Couillard C, Chernyshov AV, Charrette S, Rivard CH. Porous titanium-nickel for intervertebral fusion in a sheep model: part 1. Histomorphometric and radiological analyses. J Biomed Mater Res B Appl Biomater. 2003;64B(2):107–20.10.1002/jbm.b.10530 ↗
10. Odgaard A, Gundersen HJG. Quantification of connectivity in cancellous bone, with special emphasis on 3-D reconstruction. Bone. 1993;14(2):173–82.10.1016/8756-3282(93)90245-6 ↗
11. Feldkamp LA, Goldstein SA, Parfitt M, Jesion G, Kleerekoper M. The direct examination of three-dimensional bone architecture in vitro by computed tomography. J Bone Miner Res. 1989;4(1):3–11.10.1002/jbmr.5650040103 ↗
12. Bouxsein ML, Boyd SK, Christiansen BA, Guldberg RE, Jepsen KJ, Müller R. Guidelines for assessment of bone microstructure in rodents using micro-computed tomography. J Bone Miner Res. 2010;25(7):1468–86.10.1002/jbmr.141 ↗
13. Gebauer M, Barvencik F, Mumme M, Beil FT, Vettorazzi E, Rueger JM, et al. Microarchitecture of the radial head and its changes in aging. Calcif Tissue Int. 2010;86(1):14–22.10.1007/s00223-009-9304-0 ↗
14. Zhang ZM, Li ZC, Jiang LS, Jiang SD, Dai LY. Micro-CT and mechanical evaluation of subchondral trabecular bone structure between postmenopausal women with osteoarthritis and osteoporosis. Osteoporos Int. 2009;21(8):1383–90.10.1007/s00198-009-1071-2 ↗
15. Bonse U, Busch F, Günnewig O, Beckmann F, Pahl R, Delling G, et al. 3D computed X-ray tomography of human cancellous bone at 8 microns spatial and 10(-4) energy resolution. Bone Miner. 1994;25(1):25–38.10.1016/S0169-6009(08)80205-X ↗
16. Kirby BJ, Davis JR, Grant JA, Morgan MJ. Monochromatic microtomographic imaging of osteoporotic bone. Phys Med Biol. 1997;42(7):1375–85.10.1088/0031-9155/42/7/010 ↗
17. Bernhardt R, Scharnweber D, Müller B, Thurner P, Schliephake H, Wyss P, et al. Comparison of microfocus and synchrotron X-ray tomography for the analysis of osseontegration around Ti6Al4 V-implants. Euro Cell Mater. 2004;7:42–51.10.22203/eCM.v007a05 ↗
18. Bobyn JD, Pilliar RM, Cameron HU, Weatherly GC. The optimum pore size, for the fixation of porous-surfaced metal implants by the ingrowth of bone. Clin Orthop Relat Res. 1980;150:263–70.10.1097/00003086-198007000-00045 ↗
19. Baril E, Lefebvre LP, Piché N. X-ray microtomographic visualization and quantification of metallic foam structure filled with second phases—examples in biomedical applications. The proceeding of the 6th Int. Conf on Porous Metals and Metallic Foams (Metfoam 2009); 2009 Sep 1–4; Bratislava, Slovakia.
20. Otsuki B, Takemoto M, Fujibayashi S, Neo M, Kokubob T, Nakamura T. Pore throat size and connectivity determine bone and tissue ingrowth into porous implants: three-dimensional micro-CT based structural analyses of porous bioactive titanium implants. Biomaterials. 2006;27(35):5892–900.10.1016/j.biomaterials.2006.08.013 ↗
21. Jones AC, Arns CH, Sheppard AP, Hutmacher DW, Milthorpe K, Knacksted MA. Assessment of bone ingrowth into porous biomaterials using MICRO-CT. Biomaterials. 2007;28(15):2491–504.10.1016/j.biomaterials.2007.01.046 ↗
22. Lefebvre LP, Thomas Y. Method of making open cell material. US Patent No. 6,660,224 B2, 2003.
23. Wazen R, Lefebvre LP, Baril E, Nanci A. Initial evaluation of bone ingrowth into a novel porous titanium coating. J Biomed Mat Res-Part B. 2010;94(1):64–71.10.1002/jbm.b.31624 ↗
24. Gauthier M, Menini R, Bureau MN, So SKV, Dion MJ, Lefebvre LP. Properties of novel titanium foams intended for biomedical applications. ASM Materials and Processes for Medical Devices Conference, 2003 Sep 8–10; Anaheim, California: 382–387.
25. Hacking SA, Bobyn JD, Tanzer M, Krygier JJ. The osseous response to corundum blasted implant surfaces in a canine hip model. Clin Orthop. 1999;364:240–53.10.1097/00003086-199907000-00031 ↗
26. Hacking SA, Harvey EJ, Tanzer M, Bobyn JD, Krygier JJ. Acid etched microtexture for enhancement of bone growth into porous coated implants. J Bone Joint Surg. 2003;85(8):1182–9.10.1302/0301-620X.85B8.14233 ↗
27. Habibovic P, de Groot K. Osteoinductive biomaterials—properties and relevance in bone repair. J Tissue Eng Regen Med. 2007;1(1):25–32.10.1002/term.5 ↗
28. Hsieh J. Computed tomography: principles, design, artifacts, and recent advances. 2nd ed. Bellingham, New York: SPIE Press and Wiley; 2009.
29. Sethian JA. Level set methods and fast marching methods : evolving interfaces in computational geometry, fluid mechanics, computer vision, and materials science. 2nd ed. Cambridge: Cambridge University Press; 1999.