journal article·2017-03Open access

In vitro cytotoxicity and surface topography evaluation of additive manufacturing titanium implant materials

Jukka T. Tuomi, Roy V. Björkstrand, Mikael L. Pernu, Mika V. J. Salmi, Eero I. Huotilainen, Jan E. H. Wolff, Pekka K. Vallittu, Antti A. Mäkitie

Abstract

Custom-designed patient-specific implants and reconstruction plates are to date commonly manufactured using two different additive manufacturing (AM) technologies: direct metal laser sintering (DMLS) and electron beam melting (EBM). The purpose of this investigation was to characterize the surface structure and to assess the cytotoxicity of titanium alloys processed using DMLS and EBM technologies as the existing information on these issues is scarce. "Processed" and "polished" DMLS and EBM disks were assessed. Microscopic examination revealed titanium alloy particles and surface flaws on the processed materials. These surface flaws were subsequently removed by polishing. Surface roughness of EBM processed titanium was higher than that of DMLS processed. The cytotoxicity results of the DMLS and EBM discs were compared with a "gold standard" commercially available titanium mandible reconstruction plate. The mean cell viability for all discs was 82.6% (range, 77.4 to 89.7) and 83.3% for the control reconstruction plate. The DMLS and EBM manufactured titanium plates were non-cytotoxic both in "processed" and in "polished" forms.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Bone Tissue Engineering Materials · Dental Implant Techniques and Outcomes · Dental materials and restorations

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (17)

Show source-provided references
1. Poukens J, Laeven P, Beerens M, Nijenhuis G, Sloten JV, Stoelinga P, Kessler P. A classification of cranial implants based on the degree of difficulty in computer design and manufacture. Int J Med Robot. 2008;4(1):46–50.10.1002/rcs.171 ↗
2. Ciocca L, Fantini M, De Crescenzio F, Corinaldesi G, Scotti R. Direct metal laser sintering (DMLS) of a customized titanium mesh for prosthetically guided bone regeneration of atrophic maxillary arches. Med Biol Eng Comput. 2011;49:1347–52.10.1007/s11517-011-0813-4 ↗
3. Sidambe AT. Biocompatibility of advanced manufactured titanium implants—a review. Materials ISSN 1996-1944. Materials. 2014;7:8168–88. doi: 10.3390/ma7128168 . pp 818410.3390/ma7128168 ↗
4. Salmi M, Tuomi J, Paloheimo K-S, Björkstrand R, Paloheimo M, Salo J, Kontio R, Mesimäki K, Mäkitie AA. Patient-specific reconstruction with 3D modeling and DMLS additive manufacturing. Rapid Prototyping J. 2012;18(3):209–14.10.1108/13552541211218126 ↗
5. Bibb R, Eggbeer D, Evans P, Bocca A, Sugar A. Rapid manufacture of custom-fitting surgical guides. Rapid Prototyping J. 2009;15(5):346 –54.10.1108/13552540910993879 ↗
6. Rouse S. At the speed of light – additive manufacturing of custom medical implants. The TCT Magazine. 2009;17(6):47–50.
7. Stübinger S, Mosch I, Robotti P, Sidler M, Klein K, Ferguson SJ. Rechenberg Brigitte von. histological and biomechanical analysis of porous additive manufactured implants made by direct metal laser sintering: a pilot study in sheep. J Biomed Mater Res Part B. 2013;101B:1154–63.10.1002/jbm.b.32925 ↗
8. Haslauer CM, Springer JC, Harrysson OL, Loboa EG, Monteiro-Riviere NA, Marcellin-Little DJ. In vitro biocompatibility of titanium alloy discs made using direct metal fabrication. Med Eng Phys. 2010;32(6):645–52.10.1016/j.medengphy.2010.04.003 ↗
9. ASTM Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401); ASTM F136–13; American Society for Testing Materials: West Conshohocken, PA, USA, 2013.
10. International Organization for Standardization, ISO 17664:2004 Sterilization of medical devices – Information to be provided by the manufacturer for the processing of resterilizable medical devices.
11. ISO 10993 - 1 (Fourth edition 2009-10-15): Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process.
12. ISO 10993 - 5 (Third edition 2009-06-01): Biological evaluation of medical devices – Part 5: Test for in vitro cytotoxicity.
13. ISO 10993 - 12 (Third edition 2007-11-15; Corrected version 2008-02-15): Biological evaluation of medical devices – Part 12: Sample preparation and reference materials.
14. Murr LE, Gaytan SM, Martinez E, Medina F, Wicker RB. Next generation orthopaedic implants by additive manufacturing using electron beam melting, Int J Bio 2012, 2012, Article ID 245727, doi: 10.1155/2012/245727 .10.1155/2012/245727 ↗
15. Rapani M, Rapani C. Sinus floor lift and simultaneous implant placement: a retrospective evaluation of implant success rate. Indian J Dent. 2012;3(3):132–38.10.1016/j.ijd.2012.05.003 ↗
16. Kaivosoja E, Barreto G, Levón K, Virtanen S, Ainola M, Konttinen YT. Chemical and physical properties of regenerative medicine materials controlling stem cell fate. Ann Med. 2012;44:635–50.10.3109/07853890.2011.573805 ↗
17. Konttinen YT, Milošev I, Trebše R, van der Linden R, Pieper J, Sillat T, Virtanen S, Tiainen VM. Metals for joint replacement. In: Revell P, editor. Joint replacement technology. 2nd edn. Cambridge, England: Woodhead Publishing Limited; 2017. in press.