journal article·2012-06Open access

Biotribological study of large diameter ceramic-on-CFR-PEEK hip joint including fluid uptake, wear and frictional heating

Qian Qian Wang, Jun Jie Wu, Anthony Unsworth, Adam Briscoe, Marcus Jarman-Smith, Conor Lowry, David Simpson, Simon Collins

Abstract

A novel material combination of a large diameter Biolox(®) Delta zirconia-toughened-alumina (ZTA) head and a pitch-based carbon fibre reinforced poly ether-ether-ketone (CFR-PEEK) MOTIS(®) cup has been studied. The acetabular cups were inclined at three angles and tested using Durham Hip Simulators. The different inclination angles used did not have a significant effect on the wear rates (ANOVA, p = 0.646). Averaged over all cups, the wear rates were calculated to be 0.551 ± 0.115 mm(3)/10(6) cycles and 0.493 ± 0.107 mm(3)/10(6) cycles taking into account two types of soak controls; loaded at room temperature and unloaded at 37 °C respectively. Averaged across all femoral heads, the wear rate was 0.243 ± 0.031 mm(3)/10(6) cycles. The temperature change of the lubricant caused by the frictional heat was measured in situ. Friction factors measured using the Durham Friction Simulator were lower for the worn CFR-PEEK cups compared with unworn. This correlated with the decreased surface roughness. Even though relatively high friction was observed in these hemispherical hard-on-soft bearings, the wear rate is encouragingly low.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Advanced materials and composites

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (24)

Show source-provided references
1. Clarke MT, Lee PTH, Arora A, Villar RN. Levels of metal ions after small- and large-diameter metal-on-metal hip arthroplasty. J Bone Joint Surg British Vol. 2003;85B(6):913–7. doi: 10.1302/0301-620x.85b6.14166 .10.1302/0301-620X.85B6.14166 ↗
2. MacDonald SJ, Brodner W, Jacobs JJ. A consensus paper on metal ions in metal-on-metal hip arthroplasties. J Arthroplasty. 2004;19(8):12–6. doi: 10.1016/j.arth.2004.09.009 .10.1016/j.arth.2004.09.009 ↗
3. De Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De Smet K. Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Joint Surg British Vol. 2008;90B(10):1291–7.10.1302/0301-620X.90B10.20533 ↗
4. Grover M, Gandhe A. Head size, does it matter? Curr Orthop. 2008;22(3):155–64. doi: 10.1016/j.cuor.2008.05.003 .10.1016/j.cuor.2008.05.003 ↗
5. Kurtz SM, Devine JN. PEEK biomaterials in trauma, orthopedic, and spinal implants. Biomaterials. 2007;28(32):4845–69. doi: 10.1016/j.biomaterials.2007.07.013 .10.1016/j.biomaterials.2007.07.013 ↗
6. Wang A, Lin R, Polineni VK, Essner A, Stark C, Dumbleton JH. Carbon fiber reinforced polyether ether ketone composite as a bearing surface for total hip replacement. Tribol Int. 1998;31(11):661–7.10.1016/S0301-679X(98)00088-7 ↗
7. Scholes SC, Unsworth A. The wear properties of CFR-PEEK-OPTIMA articulating against ceramic assessed on a multidirectional pin-on-plate machine. Proc Inst Mech Eng H. 2007;221(3):281–9.10.1243/09544119JEIM224 ↗
8. Scholes SC, Inman IA, Unsworth A, Jones E. Tribological assessment of a flexible carbon-fibre-reinforced poly(ether–ether-ketone) acetabular cup articulating against an alumina femoral head. Proc Inst Mech Eng H. 2008;222(3):273–83.10.1243/09544119JEIM334 ↗
9. Chevalier J, De Aza AH, Fantozzi G, Schehl M, Torrecillas R. Crack growth resistance of alumina, zirconia and zirconia toughened alumina ceramics for joint prostheses. Biomaterials. 2002;23(3):937–45.10.1016/S0142-9612(01)00206-X ↗
10. Chevalier J, Gremillard L. Ceramics for medical applications: a picture for the next 20 years. J Eur Ceram Soc. 2009;29(7):1245–55. doi: 10.1016/j.jeurceramsoc.2008.08.025 .10.1016/j.jeurceramsoc.2008.08.025 ↗
11. Pria PD. Evolution and new application of the alumina ceramics in joint replacement. Eur J Orthop Surg Traumatol. 2007;17(3):253–6. doi: 10.1007/s00590-006-0181-1 .10.1007/s00590-006-0181-1 ↗
12. Al-Hajjar M, Leslie IJ, Tipper J, Williams S, Fisher J, Jennings LM. Effect of cup inclination angle during microseparation and rim loading on the wear of BIOLOX (R) delta ceramic-on-ceramic total hip replacement. J Biomed Mater Res Part B Appl Biomater. 2010;95B(2):263–8. doi: 10.1002/Jbm.B.31708 .10.1002/jbm.b.31708 ↗
13. Smith SL, Unsworth A. Simplified motion and loading compared to physiological motion and loading in a hip joint simulator. Proc Inst Mech Eng Part H J Eng Med. 2000;214(H3):233–8.10.1243/0954411001535723 ↗
14. ISO Standard 14242-1:2002(E). Implants for surgery—wear of total hip-joint prostheses. Part 1: loading and displacement parameters for wear-testing machines and corresponding environmental conditions for test. 1st ed. Switzerland: International Standardization Organization; 2002.
15. ISO Standard 14242-2:2000(E). Implants for surgery—wear of total hip-joint prostheses. Part 2: methods of measurement.1st ed. Switzerland: International Standardization Organization; 2000.
16. Scholes SC, Unsworth A. Comparison of friction and lubrication of different hip prostheses. Proc Inst Mech Eng Part H J Eng Med. 2000;214(H1):49–57.10.1243/0954411001535237 ↗
17. Hart AJ, Buddhdev P, Winship P, Faria N, Powell JJ, Skinner JA. Cup inclination angle of greater than 50 degrees increases whole blood concentrations of cobalt and chromium ions after metal-on-metal hip resurfacing. Hip Int. 2008;18(3):212–9.10.5301/HIP.2008.948 ↗
18. Nevelos JE, Ingham E, Doyle C, Nevelos AB, Fisher J. The influence of acetabular cup angle on the wear of “BIOLOX Forte” alumina ceramic bearing couples in a hip joint simulator. J Mater Sci Mater Med. 2001;12(2):141–4.10.1023/A:1008970027306 ↗
19. Smith SL, Unsworth A. A comparison between gravimetric and volumetric techniques of wear measurement of UHMWPE acetabular cups against zirconia and cobalt-chromium-molybdenum femoral heads in a hip simulator. Proc Inst Mech Eng Part H J Eng Med. 1999;213(6):475–83.10.1243/0954411991535086 ↗
20. Barbour PSM, Stone MH, Fisher J. A hip joint simulator study using simplified loading and motion cycles generating physiological wear paths and rates. Proc Inst Mech Eng Part H J Eng Med. 1999;213(H6):455–67.10.1243/0954411991535068 ↗
21. Williams SR, Wu JJ, Unsworth A, Khan I. Tribological and surface analysis of 38 mm alumina-as-cast Co–Cr–Mo total hip arthroplasties. Proc Inst Mech Eng Part H J Eng Med. 2009;223(H8):941–54. doi: 10.1243/09544119jeim590 .10.1243/09544119JEIM590 ↗
22. Williams SR, Wu JJ, Unsworth A, Khan I. Wear and surface analysis of 38 mm ceramic-on-metal total hip replacements under standard and severe wear testing conditions. Proc Inst Mech Eng Part H J Eng Med. 2011;225(H8):783–96. doi: 10.1177/0954411911404773 .10.1177/0954411911404773 ↗
23. Pritchett JW. Heat generated by hip resurfacing prostheses: an in vivo pilot study. J Long Term Effects of Med Implants. 2011;21(1):55–62.10.1615/JLongTermEffMedImplants.v21.i1.40 ↗
24. Bergmann G, Graichen F, Rohlmann A, Verdonschot N, van Lenthe GH. Frictional heating of total hip implants. Part 1: measurements in patients. J Biomech. 2001;34(4):421–8.10.1016/S0021-9290(00)00188-3 ↗