journal article·2022-02Open access

In vivo biocompatibility evaluation of 3D-printed nickel–titanium fabricated by selective laser melting

Hendrik Naujokat, Ali Ihsan Gökkaya, Yahya Açil, Klaas Loger, Tim Klüter, Sabine Fuchs, Jörg Wiltfang

Abstract

Abstract Nickel–titanium (NiTi) belongs to the group of shape-memory alloys (SMAs), which are characterized by flexibility and reversible deformability. Advanced techniques in 3D printing by selective laser-melting (SLM) process allow the manufacturing of complex patient-specific implants from SMAs. Osteosynthesis materials made of NiTi could be used for minimally invasive surgical approaches in oral- and maxillofacial surgery. However, the in vivo biocompatibility has not yet been fully investigated, especially in load-sharing and load-bearing implants. The aim of this study was to evaluate the in vivo biocompatibility of SLM-produced NiTi for intraosseous and subperiosteal applications. Test specimens were implanted into the frontonasal bone of ten miniature pigs. To assess peri-implant bone metabolism, fluorescent dye was administered after 2, 4, 6, 10, 12, and 14 weeks intraperitoneally. Specimens and the surrounding tissues were harvested after 8 and 16 weeks for histological analysis. While the NiTi implants presented a higher bone-to-implant contact ratio (BIC) after 8 than after 16 weeks (43.3 vs. 40.3%), the titanium implants had a significantly higher BIC after 16 weeks (33.6 vs. 67.7%). Histologically, no signs of peri-implant inflammation or foreign-body reaction were detectable. With respect to this preliminary study design, 3D-printed NiTi shows sufficient biocompatibility for intraosseous and subperiosteal implant placement.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Bone Tissue Engineering Materials · Dental materials and restorations · Anatomy and Medical Technology

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (20)

Show source-provided references
1. Kanno T, Sukegawa S, Furuki Y, Nariai Y, Sekine J. Overview of innovative advances in bioresorbable plate systems for oral and maxillofacial surgery. Jpn Dent Sci Rev. 2018;54:127–38. https://doi.org/10.1016/j.jdsr.2018.03.00310.1016/j.jdsr.2018.03.003 ↗
2. Pilling E, Eckelt U, Loukota R, Schneider K, Stadlinger B. Comparative evaluation of ten different condylar base fracture osteosynthesis techniques. Br J Oral Maxillofac Surg. 2010;48:527–31.10.1016/j.bjoms.2009.09.010 ↗
3. Gabrielli MF, Monnazzi MS, Passeri LA, Carvalho WR, Gabrielli M, Hochuli-Vieira E. Orbital wall reconstruction with titanium mesh: retrospective study of 24 patients. Craniomaxillofacial trauma reconstruction. 2011;4:151–6.10.1055/s-0031-1286120 ↗
4. San Ju G, Son KM, Choi WY, Cheon JS. Clinical usefulness of fixation of absorbable implants with cyanoacrylate in comminuted fractures of the maxilla. Arch Craniofacial Surg. 2019;20:233.10.7181/acfs.2019.00325 ↗
5. Joy-anne NO, Su Y, Lu X, Kuo P-H, Du J, Zhu D. Bioactive glass coatings on metallic implants for biomedical applications. Bioact Mater. 2019;4:261–70.10.1016/j.bioactmat.2019.09.002 ↗
6. Rana M. Three-dimensional planning and computer-assisted surgery in craniofacial reconstruction. Int J Oral Maxillofac Surg. 2017;46:44.
7. Zupanc J, Vahdat‐Pajouh N, Schäfer E. New thermomechanically treated NiTi alloys–a review. Int Endod J. 2018;51:1088–103.10.1111/iej.12924 ↗
8. Aihara H, Zider J, Fanton G, Duerig T. Combustion synthesis porous nitinol for biomedical applications. Int J Biomater. 2019;2019:4307461 https://doi.org/10.1155/2019/430746110.1155/2019/4307461 ↗
9. Chekotu JC, Groarke R, O’Toole K, Brabazon D. Advances in selective laser melting of nitinol shape memory alloy part production. Materials 2019;12:809.10.3390/ma12050809 ↗
10. Nwashindi A, Dim E. Adverse effects of nickel in transosseous wires and surgical implants: literature review. Niger J Med. 2014;23:335–43.
11. Richmann A, Kurzen S, Carron B, Richerzhagen B, editors. Cutting diamond tools using the Laser MicroJet® technology on a 5-axis machine. In Proceedings of lasers in manufacturing conference 2015, Munich, Germany, June 22–25; LIM; 2015.
12. Ayers RA, Simske SJ, Bateman T, Petkus A, Sachdeva R, Gyunter V. Effect of nitinol implant porosity on cranial bone ingrowth and apposition after 6 weeks. J Biomed Mater Res: Off J Soc Biomater, Jpn Soc Biomater, Aust Soc Biomater. 1999;45:42–7.10.1002/(SICI)1097-4636(199904)45:1<42::AID-JBM6>3.0.CO;2-Q ↗
13. Simske S, Sachdeva R. Cranial bone apposition and ingrowth in a porous nickel–titanium implant. J Biomed Mater Res. 1995;29:527–33.10.1002/jbm.820290413 ↗
14. Pan Z, Patil P. Titanium osteosynthesis hardware in maxillofacial trauma surgery: to remove or remain? A retrospective study. Eur J Trauma Emerg Surg. 2014;40:587–91.10.1007/s00068-013-0348-5 ↗
15. Rosa JH, Villanueva NL, Sanati-Mehrizy P, Factor SH, Taub PJ. Review of maxillofacial hardware complications and indications for salvage. Craniomaxillofac Trauma Reconstr. 2016;9:134–40.10.1055/s-0035-1570074 ↗
16. Assad M, Jarzem P, Leroux MA, Coillard C, Chernyshov AV, Charette S, et al. Porous titanium‐nickel for intervertebral fusion in a sheep model: Part 1. Histomorphometric and radiological analysis1. J Biomed Mater Res B Appl Biomater. 2003;64:107–20.10.1002/jbm.b.10530 ↗
17. Kujala S, Ryhänen J, Danilov A, Tuukkanen J. Effect of porosity on the osteointegration and bone ingrowth of a weight-bearing nickel–titanium bone graft substitute. Biomaterials 2003;24:4691–7.10.1016/S0142-9612(03)00359-4 ↗
18. Klopfleisch R, Jung F. The pathology of the foreign body reaction against biomaterials. J Biomed Mater Res A. 2017;105:927–40.10.1002/jbm.a.35958 ↗
19. Trindade R, Albrektsson T, Tengvall P, Wennerberg A. Foreign body reaction to biomaterials: on mechanisms for buildup and breakdown of osseointegration. Clin implant Dent Relat Res. 2016;18:192–203.10.1111/cid.12274 ↗
20. Anderson JM, Rodriguez A, Chang DT. Foreign body reaction to biomaterials. Semin Immunol. 2008;20:86–100. https://doi.org/10.1016/j.smim.2007.11.00410.1016/j.smim.2007.11.004 ↗