journal article·2010-11

Improved mechanical long-term reliability of hip resurfacing prostheses by using silicon nitride

Wen Zhang, M. Titze, B. Cappi, D. C. Wirtz, R. Telle, H. Fischer

Publisher page

Abstract

Although ceramic prostheses have been successfully used in conventional total hip arthroplasty (THA) for many decades, ceramic materials have not yet been applied for hip resurfacing (HR) surgeries. The objective of this study is to investigate the mechanical reliability of silicon nitride as a new ceramic material in HR prostheses. A finite element analysis (FEA) was performed to study the effects of two different designs of prostheses on the stress distribution in the femur-neck area. A metallic (cobalt-chromium-alloy) Birmingham hip resurfacing (BHR) prosthesis and our newly designed ceramic (silicon nitride) HR prosthesis were hereby compared. The stresses induced by physiologically loading the femur bone with an implant were calculated and compared with the corresponding stresses for the healthy, intact femur bone. Here, we found stress distributions in the femur bone with the implanted silicon nitride HR prosthesis which were similar to those of healthy, intact femur bone. The lifetime predictions showed that silicon nitride is indeed mechanically reliable and, thus, is ideal for HR prostheses. Moreover, we conclude that the FEA and corresponded post-processing can help us to evaluate a new ceramic material and a specific new implant design with respect to the mechanical reliability before clinical application.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Advanced materials and composites · Dental materials and restorations

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (56)

Show source-provided references
1. Grigoris P, Roberts P, Panousis K, Jin Z. Hip resurfacing arthroplasty: the evolution of contemporary designs. Proc Inst Mech Eng H. 2006;220:95–105.10.1243/095441105X69042 ↗
2. McMinn D, Treacy R, Lin K, Pynsent P. Metal on metal surface replacement of the hip: experience of the McMinn prosthesis. Clin Orthop Relat Res. 1996;329(Suppl):89–98.10.1097/00003086-199608001-00009 ↗
3. Schmalzried TP, Fowble VA, Ure KJ, Amstutz HC. Metal on metal surface replacement of the hip: technique, fixation, and early results. Clin Orthop Relat Res. 1996;329(Suppl):106–14.10.1097/00003086-199608001-00011 ↗
4. De Smet KA, Pattyn C, Verdonk R. Early results of primary Birmingham hip resurfacing using a hybrid metal-on-metal couple. Hip Int. 2002;12:158–62.10.1177/112070000201200225 ↗
5. Pollard TCB, Basu C, Ainsworth R, Lai W, Bannister GC. Is the Birmingham hip resurfacing worthwhile? Hip Int. 2003;13:25–8.10.1177/112070000301300104 ↗
6. Kishida Y, Sugano N, Nishii T, Miki H, Yamaguchi K, Yoshikawa H. Preservation of the bone mineral density of the femur after surface replacement of the hip. J Bone Joint Surg Br. 2004;86:185–9.10.1302/0301-620X.86B2.14338 ↗
7. Cossey AJ, Back DL, Shimmin A, Young D, Spriggins AJ. The nonoperative management of periprosthetic fractures associated with the Birmingham hip resurfacing procedure. J Arthroplast. 2005;20:358–61.10.1016/j.arth.2004.08.007 ↗
8. Treacy RB, McBryde CW, Pynsent PB. Birmingham hip resurfacing arthroplasty: a minimum follow-up of five years. J Bone Joint Surg Br. 2005;87:167–70.10.1302/0301-620X.87B2.15030 ↗
9. Loughead JM, Chesney D, Holland JP, McCaskie AW. Comparison of offset in Birmingham hip resurfacing and hybrid total hip arthroplasty. J Bone Joint Surg Br. 2005;87:163–6.10.1302/0301-620X.87B2.15151 ↗
10. Back DL, Dalziel R, Young D, Shimmin A. Early results of primary Birmingham hip resurfacings: an independent prospective study of the first 230 hips. J Bone Joint Surg Br. 2005;87:324–9.10.1302/0301-620X.87B3.15556 ↗
11. Little CP, Ruiz AL, Harding IJ, Mclardy-Smith P, Gundle R, Murray DW, Athanasou NA. Osteonecrosis in retrieved femoral heads after failed resurfacing arthroplasty of the hip. J Bone Joint Surg Br. 2005;87:320–3.10.1302/0301-620X.87B3.15330 ↗
12. De Smet KA. Belgium experience with metal-on-metal surface arthroplasty. Orthop Clin North Am. 2005;36:203–13.10.1016/j.ocl.2005.01.004 ↗
13. Amstutz HC, Campbell PA, Le Duff MJ. Fracture of the neck of the femur after surface arthroplasty of the hip. J Bone Joint Surg Am. 2004;86:1874–7.10.2106/00004623-200409000-00003 ↗
14. Shimmin AJ, Back D. Femoral neck fractures following Birmingham hip resurfacing: a national review of 50 cases. J Bone Joint Surg Br. 2005;87:463–4.10.1302/0301-620X.87B4.15498 ↗
15. Shimmin AJ, Bare J, Back DL. Complications associated with hip resurfacing arthroplasty. Orthop Clin North Am. 2005;36:187–93.10.1016/j.ocl.2005.01.002 ↗
16. Taylor M. Finite element analysis of the resurfaced femoral head. Proc Inst Mech Eng H. 2006;220:289–97.10.1243/095441105X93631 ↗
17. Goebel S, Blanke M, Hendrich C. Surface replacement of the hip: contra. In: Lazennec JY, Dietrich M, editors. Ceramics in orthopaedics, 9th Biolox® symposium proceedings. Darmstadt: Steinkopff Verlag; 2004. p. 119–26.
18. Clarke MT, Lee PTH, Arora A, Villar RN. Levels of metal ions after small- and large-diameter metal-on-metal hip arthroplasty. J Bone Joint Surg Br. 2003;85:913–7.10.2106/00004623-200305000-00023 ↗
19. Jacobs JJ, Skipor AK, Doorn PF, Campbell PA, Schmalzried TP, Black J, Amstutz HC. Cobalt and chromium concentrations in patients with metal on metal total hip replacement. Clin Orthop Relat Res. 1996;329S:256–63.10.1097/00003086-199608001-00022 ↗
20. Jacobs JJ, Skipor AK, Patterson LM, Hallab NJ, Paprosky WG, Black J, Galante JO. Metal release in patients who have had a primary total hip arthroplasty. A prospective, controlled, longitudinal study. J Bone Joint Surg Am. 1998;80:1447–58.10.2106/00004623-199810000-00006 ↗
21. Savarino L, Granchi D, Ciapetti G, Cenni E, Greco M, Rotini R, Veronesi CA, Baldini N, Giunti A. Ion release in stable hip arthroplasties using metal-on-metal articulating surfaces: a comparison between short- and medium-term results. J Biomed Mater Res. 2003;66A:450–6.10.1002/jbm.a.10595 ↗
22. Savarino L, Greco M, Cenni E, Cavasinni L, Rotini R, Baldini N, Giunti A. Differences in ion release after ceramic-on-ceramic and metal-on-metal total hip replacement, medium-term follow-up. J Bone Joint Surg Br. 2006;88:472–6.10.1302/0301-620X.88B4.17333 ↗
23. Daniel J, Ziaee H, Pradhan C, Pynsent PB, McMinn DJ. Blood and urine metal ion levels in young and active patients after Birmingham hip resurfacing arthroplasty: four-year results of a prospective longitudinal study. J Bone Joint Surg Br. 2007;89:169–73.10.1302/0301-620X.89B2.18519 ↗
24. Mont MA, Schmalzried TP. Modern metal-on-metal hip resurfacing: important observations from the first ten years. J Bone Joint Surg Am. 2008;90:3–11.10.2106/JBJS.H.00750 ↗
25. Hart AJ, Skinner JA, Winship P, Faria N, Kulinskaya E, Webster D, Muirhead-Allwood S, Aldam CH, Anwar H, Powell JJ. Circulating levels of cobalt and chromium from metal-on-metal hip replacement are associated with CD8+ T-cell lymphopenia. J Bone Joint Surg Br. 2009;91:835–42.10.1302/0301-620X.91B6.21844 ↗
26. Willert HG, Buchhorn GH, Fayyazi A, Flury R, Windler M, Köster G, Lohmann CH. Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints. A clinical and histomorphological study. J Bone Joint Surg Am. 2005;87:28–36.10.2106/JBJS.A.02039pp ↗
27. Griss P, Werner E. Alumina ceramic, bioglass and silicon nitride. A comparative biocompatibility study. In: Hastings GW, Williams DF, editors. Mechanical properties of biomaterials. London: John Wiley & Sons Ltd; 1980. p. 217–25.
28. Willmann G. Survival rate and reliability of ceramic femoral heads for total hip arthroplasty. Mat-wiss u Werkstofftech. 1998;29:595–604.10.1002/mawe.19980291010 ↗
29. Christel PS. Biocompatibility of surgical-grade dense polycrystalline alumina. Clin Orthop Relat Res. 1992;379:10–8.
30. Dorlot JM. Long-term effects of alumina components in total hip prostheses. Clin Orthop. 1992;282:47–52.10.1097/00003086-199209000-00006 ↗
31. Jazrawi LM, Kummer FJ, DiCesare PE. Alternative bearing surfaces for total joint arthroplasty. J Am Acad Orthop Surg. 1998;6:198–203.10.5435/00124635-199807000-00001 ↗
32. Granchi D, Savarino L, Ciapetti G, Cenni E, Rotini R, Mieti M, Baldini N, Giunti A. Immunological changes in patients with total joint replacement following idiopathic osteoarthritis of the hip. J Bone Joint Surg Br. 2003;85:758–64.10.1302/0301-620X.85B5.13729 ↗
33. Bizot P, Nizard R, Hamadouche M, Hannouche D, Sedel L. Prevention of wear and osteolysis: alumina-on-alumina bearing. Clin Orthop. 2001;393:85–93.10.1097/00003086-200112000-00010 ↗
34. Shimmin A, Beaule PE, Campbell P. Metal-on-metal hip resurfacing arthroplasty. J Bone Joint Surg Am. 2008;90:637–54.10.2106/JBJS.G.01012 ↗
35. Ong KL, Manley MT, Kurtz SM. Have contemporary hip resurfacing designs reached maturity? A review. J Bone Joint Surg Am. 2008;90:81–8.10.2106/JBJS.H.00574 ↗
36. Cappi B, Neuss S, Salber J, Telle R, Knüchel R, Fischer H. Cytocompatibility of high strength non-oxide ceramics. J Biomed Mater Res A. 2009;93:67–76.10.1002/jbm.a.32527 ↗
37. Andrisano AO, Dragoni E, Strozzi A. Axisymmetric mechanical analysis of ceramic heads for total hip replacement. Proc Inst Mech Eng H. 1990;204:157–67.10.1243/PIME_PROC_1990_204_250_02 ↗
38. Prendergast PJ. Finite element models in tissue mechanics and orthopaedic implant design. Clin Biomech. 1997;12:343–66.10.1016/S0268-0033(97)00018-1 ↗
39. Ong KL, Kurtz SM, Manley MT, Rushton N, Mohammed NA, Field RE. Biomechanics of the Birmingham hip resurfacing arthroplasty. J Bone Joint Surg Br. 2006;88:1110–5.10.1302/0301-620X.88B8.17567 ↗
40. Kaltenborn N, Sax M, Müller FA, Müller L, Dieker H, Kaiser A, Telle R, Fischer H. Coupling of phosphate on alumina surfaces for bioactivation. J Am Ceram Soc. 2007;90:1644–6.10.1111/j.1551-2916.2007.01601.x ↗
41. Bergmann G, Deuretzbacher G, Heller M, Graichen F, Rohlmann A, Strauss J, Duda GN. Hip contact forces and gait patterns from routine activities. J Biomech. 2001;34:859–71.10.1016/S0021-9290(01)00040-9 ↗
42. Nemeth NN, Manderscheid JM, Gyekenyesi JP. Ceramics analysis and reliability evaluations of structures (CARES). NASA Technical Paper 2916. Cleveland: Glenn Research Centre; 1989.
43. Nemeth NN, Powers LM, Janosik LA, Gyekenyesi JP. CARES/LIFE. Users and programmers manual. Cleveland: NASA, Glenn Research Centre; 1993.
44. Munz D, Fett T. Ceramics: mechanical properties, failure behaviour, materials selection. 1st ed. Berlin: Springer; 1999.10.1007/978-3-642-58407-7 ↗
45. Weibull W. A statistical theory of the strength of materials. Handlingar 151: Ingeniörs Vetenskaps Akademien; 1939.
46. Griffith AA. The phenomena of rupture and flaw in solids. Philos Trans R Soc. 1924;221:163–89.10.1098/rsta.1921.0006 ↗
47. Fischer H, Weber M, Marx R. Lifetime prediction of all-ceramic bridges by computational methods. J Dent Res. 2003;82:238–42.10.1177/154405910308200317 ↗
48. Andrews MJ, Wereszczak AA, Breder K. Predictions of the inert strength distribution of Si3N4 diesel valves. Ceram Eng Sci Proc. 1999;20(3):555–63.10.1002/9780470294567.ch63 ↗
49. DIN EN 843-1. Advanced technical ceramics—mechanical properties of monolithic ceramics at room temperature—part 1: determination of flexural strength; 2006.
50. DIN EN 843-3. Advanced technical ceramics—mechanical properties of monolithic ceramics at room temperature—part 3: determination of subcritical crack growth parameters from constant stressing rate flexural strength tests; 2005.
51. Long JP, Santner TJ, Bartel DL. Hip resurfacing increases bone strains associated with short-term femoral neck fracture. J Orthop Res. 2009;27(10):1319–25.10.1002/jor.20884 ↗
52. Heijink A, Zobitz ME, Nuyts R, Morrey BF, An KN. Prosthesis design and stress profile after hip resurfacing: a finite element analysis. J Orthop Surg. 2008;16:326–32.10.1177/230949900801600312 ↗
53. Watanabe Y, Shiba N, Matsuo S, Higuchi F, Tagawa Y, Inoue A. Biomechanical study of the resurfacing hip arthroplasty. J Arthroplast. 2000;15:505–11.10.1054/arth.2000.1359 ↗
54. Brown TA, Kohan L, Ben-Nissan B. Assessment by finite element analysis of the impact of osteoporosis and osteoarthritis on hip resurfacing. 5th Australasian Congress on Applied Mechanics, ACAM 2007. Brisbane.
55. Südmeyer IJ. Dissertation: Zementfreie Hüftkopf-Oberflächenprothese aus Aluminiumoxid Orientierende FEM-Zuverlässigkeitsanalysen. Aachen: RWTH; 2008.
56. N N. Materials data silicon nitride. Germany: ESK Ceramics GmbH & Co. KG Kempten; 2006.