Cytotoxicity study of homopolymers and copolymers of 2-hydroxyethyl methacrylate and some alkyl acrylates for potential use as temporary skin substitutes
Source abstract
Abstract
The cytotoxicities of both homopolymers and copolymers of 2-hydroxyethyl methacrylate (HEMA) and various alkyl acrylates (AA) were investigated using cell culture techniques. The particular alkyl acrylates used were methyl acrylate (MA), ethyl acrylate (EA) and butyl acrylate (BA). The AA contents of the polymers studied were in the range of 0-25% by weight. They were synthesized via bulk polymerization in the form of thin sheets of 0.5+/-0.1 mm thickness. After sterilization using ethylene oxide gas, cytotoxicity tests, including both direct and indirect contact tests, were performed using cell lines of L929 and normal human dermal fibroblasts. The results showed that P(HEMA-co-EA) with EA contents of 15-25% and P(HEMA-co-BA) with BA contents of 10-25% appeared to be non-cytotoxic and attractive for more specific cytocompatibility tests and in vivo studies with a view to their use as temporary skin substitutes.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Advanced Polymer Synthesis and Characterization · Photopolymerization techniques and applications · Antimicrobial agents and applications
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (14)
Show source-provided references
1. D. F. WilliamsBiocompatibility of Clinical Implant Materials, Vol. 2 (CRC Press, Inc., Florida, USA, 1981).
2. J. P. Montheard, M. Chatzopoulos and D. Chappard, Macromol. Chem. Phys. C. 32 (1992) 1.10.1080/15321799208018377 ↗
3. P. A. Davies, S. J. Huang, L. Nicolais and L. AmbrosioHigh Performance Biomaterials, edited by M. Szycher (Technomic, Basel, 1991), p. 343.
4. M. D. Blanco, R. M. Trigo, C. Teijon, C. Gomez and J. M. Teijon, Biomaterials 19 (1998) 861.10.1016/S0142-9612(97)00247-0 ↗
5. M. Santin, S. J. Huang, S. Iannace, L. Ambrosio, L. Nicolais and G. Peluso, Biomaterials 17 (1996) 1459.10.1016/0142-9612(96)89769-9 ↗
6. B. W. Ziegelaar, J. H. Fitton, A. B. Clayton, S. T. Platten, M. A. L. Maley and T. V. Chirila, Biomaterials 20 (1999) 1979.10.1016/S0142-9612(99)00019-8 ↗
7. H. Uludag and M. V. Sefton, J. Biomed. Mater. Res. 27 (1993) 1213.10.1002/jbm.820271002 ↗
8. X. Lou, S. Vijayasekaran, T. V. Chirila, M. A. L. Maley, C. R. Hicks and I. J. Constable, J. Biomed. Mater. Res. 47 (1999) 404.10.1002/(SICI)1097-4636(19991205)47:3<404::AID-JBM16>3.0.CO;2-F ↗
9. S. Vijayasekaran, J. H. Fitton, C. R. Hicks, T. V. Chirila, G. J. Crawford and I. J. Constable, Biomaterials 19 (1998) 2255.10.1016/S0142-9612(98)00128-8 ↗
10. ISO 10993-5: 1992(E) Biological evaluation of medical devices, Part 5.
11. USP 24 (87) Biological reactivity tests, in vitro (2000), p. 1813.
12. J. A. Plumb, R. Milroy and S. B. Kaye, Cancer Res. 49 (1989) 4435.
13. S. I. Ertel, B. D. Ratner, A. Kaul, M. B. Schway and T. A. Hortbett, J. Biomed. Mater. Res. 28 (1994) 667.10.1002/jbm.820280603 ↗
14. D. Hor'ak, M. Cervinka and V. Pouza, Biomaterials 18 (1997) 1355.10.1016/S0142-9612(97)00059-8 ↗