journal article·2001-04Open access

Pin-on-plate studies on the effect of rotation on the wear of metal-on-metal samples

S. C. Scholes, A. Unsworth

Abstract

An important cause of osteolysis and subsequent loosening of replacement hip joints is the body's biological response to polyethylene wear debris. Interest has thus been renewed in hard bearing surfaces such as metal-on-metal implants. Tests were performed on a pin-on-plate machine to determine the effects of pin rotation on the wear of two different compositions of cobalt chrome molybdenum (CoCrMo) against itself (high carbon and low carbon). With reciprocating motion only, the low carbon material gave an order of magnitude higher wear than the high carbon material. The overall wear (that for both the pin and the plate) was significantly reduced with added rotation for the low carbon material but remained approximately the same for the high carbon material. However, the wear of the low carbon material was not reduced below that of the high carbon material which remained the best material in terms of wear.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Advanced materials and composites · Total Knee Arthroplasty Outcomes

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (20)

Show source-provided references
1. B. M. WROBLEWSKI and P. D. SINEY, Clin. Orthop. Rel. Res. 292 (1993) 191.10.1097/00003086-199307000-00024 ↗
2. J. J. CALLAGHAN, E. E. FOREST, J. P. OLEJNICZAK, D. D. GOETZ and R. C. JOHNSTON, J. Bone Jt. Surg. 80-A(5) (1998) 704.10.2106/00004623-199805000-00011 ↗
3. D. W. HOWIE, B. VERNON-ROBERTS, R. OAKESHOTT, B. MANTHEY, J. Bone Jt. Surg. 70-A(2) (1988) 257.10.2106/00004623-198870020-00015 ↗
4. T. P. SCHMALZRIED, M. JASTY and W. H. HARRIS, J. Bone Jt. Surg. 74-A(6) (1992) 849.10.2106/00004623-199274060-00006 ↗
5. D. W. HOWIE, B. MANTHEY, S. HAY and B. VERNONROBERTS, Clin. Orthop. Rel. Res. (1993) 352.10.1097/00003086-199307000-00046 ↗
6. P. F. DOORN, P. A. CAMPBELL and H. C. AMSTUTZ, Clin. Orthop. Rel. Res. 329S (1996) S206.10.1097/00003086-199608001-00018 ↗
7. M. SCHMIDT, H. WEBER and R. SCHÖN, Clin. Orthop. Rel. Res. 329S (1996) S35.10.1097/00003086-199608001-00004 ↗
8. M. KOTHARI, D. L. BARTEL and J. F. BOOKER, Clin. Orthop. Rel. Res. 329S (1996) S141.10.1097/00003086-199608001-00014 ↗
9. M. E. MÜLLER, Clin. Orthop. Rel. Res. 311 (1995) 54.
10. H. C. AMSTUTZ and P. GRIGORIS, Clin. Orthop. Rel. Res. 329S (1996) S11.10.1097/00003086-199608001-00003 ↗
11. S-A. JACOBSSON, K. DJERF and O. WAHLSTRÖM, Clin. Orthop. Rel. Res. 329S (1996) S60.10.1097/00003086-199608001-00006 ↗
12. T. P. SCHMALZRIED, P. C. PETERS, B. T. MAURER, C. R. BRAGDON and W. H. HARRIS, J. Arthroplasty 113 (1996) 322.10.1016/S0883-5403(96)80085-4 ↗
13. M. SEMLITSCH, SERC/IMechE Annual Expert Meeting on ``Failure of Joint Prostheses'', 28–30 November (1993) 40.
14. F. W. CHAN, J. D. BOBYN, J. B. MEDLEY, J. J. KRYGIER, S. YUE and M. TANZER, Clin. Orthop. Rel. Res. 333 (1996) 96.
15. R. FARRAR and M. B. SCHMIDT, 43rd Annual Meeting, ORS, San Francisco, California, February (1997), 71–12.
16. A. WANG, C. STARK and J. H. DUMBLETON, Proc. Instn. Mech. Engrs 210 (1996) 141.10.1243/PIME_PROC_1996_210_407_02 ↗
17. V. SAIKKO, J. Biomed. Mater. Res. 41 (1998) 58.10.1002/(SICI)1097-4636(199807)41:1<58::AID-JBM7>3.0.CO;2-P ↗
18. J. L. TIPPER, P. J. FIRKINS, E. INGHAM, J. FISHER, M. H. STONE and R. FARRAR, J. Mater. Sci.: Mater. Med. 10 (1999) 353.
19. ASTM F732-82. Annual Book of ASTM Standards 13.01 (1982) 263.
20. S. C. SCHOLES, The tribology of hard bearing surfaces for use in hip prostheses. Ph.D. thesis, University of Durham (1999) 207.