Cathodic activation and inflammatory species are critical to simulating in vivo Ti-6Al-4V selective dissolution
Source abstract
Abstract
In vivo retrievals of metallic orthopedic implants have shown selective dissolution of Ti-6Al-4V, where the vanadium-rich β phase preferentially corrodes from the surface. This damage, typically observed in crevices, is not directly caused by wear mechanics and the underlying electrochemical mechanism remains poorly understood. Previous studies show that fretting corrosion can cause negative potential drops, resulting in a decrease in surface oxide passivation resistance and the electrochemical generation of reactive oxygen species (ROS) at metallic surfaces. In this study, we combine cathodic activation and hydrogen peroxide to induce selective dissolution in vitro. After a 600 s -1 V hold and 4 h recovery in 20 °C 1 M H2O2 solution, the Ti-6Al-4V β phase was preferentially dissolved. An initial activation threshold of -0.5 V induced a significant increase in β dissolution (p = 0.000). Above this threshold, little selective dissolution occurred. In an Arrhenius-like fashion, decreasing solution concentration to 0.1 M required 72 h to generate β dissolution instead of 4 h at 1 M. Heating 0.1 M solution to body temperature (37 °C) resulted in a decrease in the time needed to replicate a similar level of β dissolution (>90%). Electrochemical impedance shows that both cathodic activation and inflammatory species are necessary to induce selective dissolution, where the combinatorial effect causes a significant drop in oxide passivation resistance from 106 to 102 (p = 0.000). STATEMENT OF SIGNIFICANCE: Though hip arthroplasties are considered a successful procedure, revision rates of 2-4% result in tens of thousands of additional surgeries within the United States, subjecting patients to increased risk of complications. Corrosion is associated with implant failure and retrieval studies show that titanium and its alloys can severely corrode in vivo in ways not yet duplicated in vitro. Here, we reproduce selective dissolution of Ti-6Al-4V β phase simulating key characteristics of in vivo degradation observed in orthopedic retrievals. We establish both cathodically activated corrosion, a relatively unexplored concept, and the presence of inflammatory species as prerequisites, furthering our understanding of this clinically relevant damage mode. We introduce an Arrhenius-based approach to assess the concentration-temperature-time interactions present.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Titanium Alloys Microstructure and Properties · Hydrogen embrittlement and corrosion behaviors in metals
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (50)
Show source-provided references
1. 1.2 electrochemical behavior of metals in the biological milieu
2. Gilbert · 2012 · 1
3. Biomaterials in orthopaedics10.1098/rsif.2008.0151 ↗
4. Boyer · 1994
5. Corrosion and wear at the modular interface of uncemented femoral stems10.1302/0301-620X.76B1.8300685 ↗
6. Femoral stem fracture and in vivo corrosion of retrieved modular femoral hips10.1016/j.arth.2011.11.007 ↗
7. Corrosion between the components of modular femoral hip prostheses10.1302/0301-620X.74B4.1624507 ↗
8. Corrosion at the stem-sleeve interface of a modular titanium alloy femoral component as a reason for impaired disengagement10.1016/j.arth.2009.10.018 ↗
9. Proximal component modularity in THA—at what cost? An implant retrieval study10.1007/s11999-011-2155-9 ↗
10. Corrosion of modular titanium alloy stems in cementless hip replacement10.1520/JAI12810 ↗
11. Titanium corrosion mechanisms in the oral environment: a retrieval study10.3390/ma6115258 ↗
12. A multicenter retrieval study of the taper interfaces of modular hip prostheses10.1097/00003086-200208000-00018 ↗
13. In vivo severe corrosion and hydrogen embrittlement of retrieved modular body titanium alloy hip-implants10.1002/jbm.b.31171 ↗
14. In vivo oxide-induced stress corrosion cracking of Ti-6Al-4V in a neck–stem modular taper: emergent behavior in a new mechanism of in vivo corrosion10.1002/jbm.b.31943 ↗
15. What do we know about taper corrosion in total hip arthroplasty?10.1016/j.arth.2014.02.014 ↗
16. In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations. The effect of crevice, stress, motion, and alloy coupling10.1002/jbm.820271210 ↗
17. Gilbert · 1997 · 45
18. Mechanically assisted crevice corrosion in metallic biomaterials: a review10.1080/10667857.2016.1223909 ↗
19. The reduction half cell in biomaterials corrosion: oxygen diffusion profiles near and cell response to polarized titanium surfaces10.1002/(SICI)1097-4636(199811)42:2<321::AID-JBM18>3.0.CO;2-L ↗
20. Potential and frequency effects on fretting corrosion of Ti6Al4V and CoCrMo surfaces10.1002/jbm.a.34564 ↗
21. Gilbert · 2007
22. Gilbert · 2011
23. Microstructural aspects of Ti6Al4V degradation in H2O2-containing phosphate buffered saline10.1016/j.corsci.2021.109640 ↗
24. Titanium electrochemistry in the presence of the inflammatory species H2O2
25. In vivo severe corrosion and hydrogen embrittlement of retrieved modular body titanium alloy hip-implants10.1002/jbm.b.31171 ↗
26. Area-dependent impedance-based voltage shifts during tribocorrosion of Ti-6Al-4V biomaterials: theory and experiment10.1088/2051-672X/4/3/034002 ↗
27. A metallic biomaterial tribocorrosion model linking fretting mechanics, currents, and potentials: model development and experimental comparison10.1002/jbm.b.34643 ↗
28. The effect of static applied potential on the 24-hour impedance behavior of commercially pure titanium in simulated biological conditions10.1002/jbm.b.31564 ↗
29. A fluorescent approach for detecting and measuring reduction reaction byproducts near cathodically-biased metallic surfaces: Reactive oxygen species production and quantification10.1016/j.bioelechem.2019.05.020 ↗
30. In vitro evidence for cell-accelerated corrosion within modular junctions of total hip replacements10.1002/jor.24447 ↗
31. Mechanical, chemical and biological damage modes within head-neck tapers of CoCrMo and Ti6Al4V contemporary hip replacements10.1002/jbm.b.33972 ↗
32. Nonlinear optimization using the generalized reduced gradient method, revue française d'automatique, informatique, recherche opérationnelle10.1051/ro/197408V300731 ↗
33. Analysis of electrochemical impedance spectra using phase angle symmetry across log frequency10.1149/1945-7111/ab69f6 ↗
34. Temperature-dependence corrosion behavior of Ti6Al4V in the presence of HCl10.3389/fmats.2022.880702 ↗
35. Determinants of corrosion resistance of Ti-6Al-4V alloy dental implants in an in vitro model of peri-implant inflammation10.1371/journal.pone.0210530 ↗
36. Corrosion kinetics and topography analysis of Ti–6Al–4V alloy subjected to different mouthwash solutions10.1016/j.msec.2014.06.033 ↗
37. Effect of inflammatory conditions and H2O2 on bare and coated Ti–6Al–4V surfaces: corrosion behavior, metal ion release and Ca-P formation under long-term immersion in DMEM10.1016/j.apsusc.2015.08.261 ↗
38. Hydrogen peroxide treatment on Ti–6Al–4V alloy: a promising surface modification technique for orthopaedic application10.1016/j.apsusc.2009.09.069 ↗
39. Influence of different concentration of hydrogen peroxide on the corrosion behavior of Ti-6Al-4V alloy immersed in physiological solution10.1088/1757-899X/572/1/012006 ↗
40. 2019
41. The effect of hydrogen peroxide on the electrochemical behavior of Ti and some of its alloys for dental applications10.1016/j.matchemphys.2005.10.032 ↗
42. Effect of hydrogen peroxide on titanium surfaces: in situ imaging and step-polarization impedance spectroscopy of commercially pure titanium and titanium, 6-aluminum, 4-vanadium10.1002/jbm.a.10116 ↗
43. Corrosion kinetics and topography analysis of Ti–6Al–4V alloy subjected to different mouthwash solutions10.1016/j.msec.2014.06.033 ↗
44. Corrosion behavior of Ti-6Al-4V with different thermomechanical treatments and microstructures10.5006/1.3452400 ↗
45. Elucidating the corrosion-related degradation mechanisms of a Ti-6Al-4V dental implant10.1016/j.dental.2020.01.008 ↗
46. A synergistic effect of albumin and H2O2 accelerates corrosion of Ti6Al4V10.1016/j.actbio.2015.07.046 ↗
47. Titanium is not “the most biocompatible metal” under cathodic potential: the relationship between voltage and MC3T3 preosteoblast behavior on electrically polarized cpTi surfaces10.1002/jbm.a.32622 ↗
48. The effect of cathodic electrochemical potential of Ti-6Al-4V on cell viability: voltage threshold and time dependence10.1002/jbm.b.32970 ↗
49. Oxidative stress, inflammation, and the corrosion of metallic biomaterials: corrosion causes biology and biology causes corrosion
50. Hydrogen peroxide in the human body10.1016/S0014-5793(00)02197-9 ↗