In vitro and in vivo evaluation of the safety and stability of the TAXUS® Paclitaxel-Eluting Coronary Stent
Source abstract
Abstract
Functional aspects of the styrene-b-isobutylene-b-styrene triblock copolymer (SIBS) which is incorporated into a drug-eluting stent (DES) coating are described. The SIBS copolymer is employed on the TAXUS Paclitaxel-Eluting Coronary Stent as a carrier for paclitaxel (PTx). Optical and scanning electron microscopic analysis of stents explanted from rabbit and porcine models after 2 years and 6 months, respectively, showed that the SIBS coating maintained physical integrity. Gel permeation chromatography (GPC) of the copolymer extracted from the coating verified that no polymer degradation occurred over the same period of time. The coating on TAXUS Stents was shown to maintain physical integrity after 400 million cycles of pulsatile or mechanical (tensile) fatigue, simulating 10 years real time use. Inspection of the samples compared to untested controls showed no change in the coating under these cyclic simulated conditions. Films prepared with the same formulation found on TAXUS Stents maintained mechanical strength and resistance throughout the time of testing. Intentional defects introduced into the stent coating were shown to have only a minimal impact on PTx release. These data support the suitability of the SIBS copolymer as a drug carrier for DES applications.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Coronary Interventions and Diagnostics · Integrated Circuits and Semiconductor Failure Analysis · Cardiac Valve Diseases and Treatments
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (16)
Show source-provided references
1. G.W. Stone, J.W. Moses, S.G. Ellis, J. Schofer, K.D. Dawkins, M.C. Morice, A. Colombo, E. Schampaert, E. Grube, A.J. Kirtane, D.E. Cutlip, M. Fahy, S.J. Pocock, R. Mehran, M.B. Leon, Safety and efficacy if Sirolimus- and Paclitaxel-Eluting coronary stents. N. Engl. J. Med. 356(10), 998–1008 (2007). doi: 10.1056/NEJMoa06719310.1056/NEJMoa067193 ↗
2. J.E. Sousa, P.W. Serruys, M.A. Costa, New frontiers in cardiology: drug-eluting stents: part I. Circulation 107, 2274–2279 (2003). doi: 10.1161/01.CIR.0000069330.41022.9010.1161/01.CIR.0000069330.41022.90 ↗
3. J.E. Sousa, P.W. Serruys, M.A. Costa, New frontiers in cardiology: drug-eluting stents: part ii. Circulation 107, 2383–2389 (2003). doi: 10.1161/01.CIR.0000069331.67148.2F10.1161/01.CIR.0000069331.67148.2F ↗
4. E. Grube, S. Silber, K.E. Hauptmann, R. Mueller, L. Buellesfeld, U. Gerckens, M.E. Russell, I. Taxus, Six- and twelve-month results from a randomized, double-blind trial on a slow-release Paclitaxel-Eluting Stent for de novo coronary lesions. Circulation 107(1), 38–42 (2003). doi: 10.1161/01.CIR.0000047700.58683.A110.1161/01.CIR.0000047700.58683.A1 ↗
5. G.W. Stone et al., for the Taxus-IV investigators; A polymer-based, paclitaxel-eluting stent in patients with coronary artery disease. N. Engl. J. Med. 350(3):221–231 (2004). doi: 10.1056/NEJMoa03244110.1056/NEJMoa032441 ↗
6. K.D. Dawkins, et al., on behalf of the Taxus-VI investigators; Clinical efficacy of polymer-based paclitaxel-eluting stents in the treatment of complex, long coronary artery lesions from a multicenter, randomized trial. Circulation 112:3306–3313 (2005). doi: 10.1161/CIRCULATIONAHA.105.55219010.1161/CIRCULATIONAHA.105.552190 ↗
7. K.R. Kamath, J.J. Barry, K.M. Miller, The Taxus™ drug-eluting stent: a new paradigm in controlled drug delivery. Adv. Drug Deliv. Rev. 58, 412–436 (2006). doi: 10.1016/j.addr.2006.01.02310.1016/j.addr.2006.01.023 ↗
8. F. Strickler, R. Faust, M. Boden, R. Richard, S. McFadden, M. Bravo, J. Lindquest, P. Straumann, Characterization of biostability and vascular compatibility and safety of different compositions of poly(styrene-b-isobutylene-b-styrene copolymers. Polymer Prepr. 48(1), 360 (2007)
9. Z. Fodor, M. Gyor, H.C. Wang, R. Faust, Living carbocationic polymerization of styrene in the presence of proton trap. J. Macromol. Sci. Pure Appl. Chem. A30(5), 349–363 (1993)10.1080/10601329308009412 ↗
10. M. Gyor, Z. Fodor, H.C. Wang, R. Faust, Polyisobutylene based thermoplastic elastomer I synthesis and characterization of SIBS triblock copolymers. J. Macromol. Sci. Pure Appl. Chem. A31(12), 2055–2065 (1994)10.1080/10601329409350119 ↗
11. S.V. Ranade, K.M. Miller, R. Richard, A.K. Chan, M.J. Allen, M.N. Helmus, Physical characterization of controlled release of paclitaxel from the Taxus™ Express2™ drug eluting stent. J. Biomed. Mater. Res. 71A, 625–634 (2004). doi: 10.1002/jbm.a.3018810.1002/jbm.a.30188 ↗
12. L. Slusarski, G. Janowska, Thermal decomposition of homo- and copolymers of isobutylene. J. Therm. Anal. 19, 435–447 (1980). doi: 10.1007/BF0190397510.1007/BF01903975 ↗
13. R. Pazur, M. Troquet, J. GArdette, Thermo- and photooxidation of polyisobutylene Part I evolutions at temperatures above 50°C. J. Polym. Sci. Polym. Chem. 35(9), 1689–1701 (1997). doi: 10.1002/(SICI)1099-0518(19970715)35:9<1689::AID-POLA10>3.0.CO;2-110.1002/(SICI)1099-0518(19970715)35:9<1689::AID-POLA10>3.0.CO;2-1 ↗
14. I. Mita, T. Hisano, K. Horie, A. Okamoto, Photoinitiated thermal degradation of polymers. Part I. Elementary process of degradation of polystyrene. Macromolecules 21, 3003–3010 (1988). doi: 10.1021/ma00188a01910.1021/ma00188a019 ↗
15. A. Colombo, J. Drzewiecki, A. Banning, E. Grube, K. Hauptmann, S. Silber, D. Dudek, S. Fort, F. Schiele, K. Zmudka, G. Guagliumi, M.E. Russell, Randomized study to assess the effectiveness of slow- and moderate-release polymer-based Paclitaxel-Eluting Stents for coronary artery lesions. Circulation 108, 788–794 (2003). doi: 10.1161/01.CIR.0000086926.62288.A610.1161/01.CIR.0000086926.62288.A6 ↗
16. R.S. Schwartz, E.R. Edelman for the consensus committee, A. Carter, N. Chronos, C. Rogers, K.A. Robinson, R. Waksman, J. Weinberg, R.L. Wilensky, D.N. Jensen, B.D. Zuckerman, R. Virmani, Drug-eluting stents in preclinical studies: recommended evaluation from a consensus group. Circulation 106:1867–1873 (2002). doi: 10.1161/01.CIR.0000033485.20594.6F10.1161/01.CIR.0000033485.20594.6F ↗