Evaluation of an osteostimulative putty in the sheep spine
Source abstract
Abstract
The objective of this study is to evaluate local response to a bioactive glass based composite putty (NovaBone Putty) in a vertebral body defect model in sheep, as compared to NovaBone, a bioactive glass particulate. Two time periods were used for the study, 6 and 12 weeks. Empty defects were also used as a control. In comparing the three test groups, the relative amount of new bone for both grafted defects was substantially greater than for the empty controls (P < 0.05). At 6 weeks, the bone formation was 42% for NovaBone Putty, 27% for NovaBone and 1.2% for the ungrafted empty defect. At 12 weeks, the bone formation was 51.4% for NovaBone Putty, 47.3% for NovaBone and 5.1% for the empty defect. NovaBone Putty, the test material, had greater bone content than the NovaBone, both of which were significantly greater than the empty control.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Bone Tissue Engineering Materials · Spine and Intervertebral Disc Pathology · Orthopaedic implants and arthroplasty
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (21)
Show source-provided references
1. LeGeros RZ. Calcium phosphate materials in restorative dentistry: a review. Adv Dent Res. 1988;2:164–83.10.1177/08959374880020011101 ↗
2. Hench LL, Splinter RJ, Greelee TK, Allen WC. Bonding mechanisms at the interface of ceramic prosthetic materials. J Biomed Mater Res. 1971;2:117–41.10.1002/jbm.820050611 ↗
3. Hench LL, Paschall HA. Direct bonding of bioactive glass-ceramic materials to bone and muscle. J Biomed Mater Res Symp. 1973;4:25–42.10.1002/jbm.820070304 ↗
4. Hench LL, West JK. Biological applications of bioactive glasses. Life Chem Rep. 1996;13:187–241.
5. Filgueiras MR, LaTorre GP, Hench LL. Solution effects on the surface reactions of a bioactive glass. J Biomed Mater Res. 1993;27:445–53.10.1002/jbm.820270405 ↗
6. Zhong JP, Greenspan DC. Bioglass surface reactivity: from in vitro to in vivo. Bioceramics. 1998;11:415–8.
7. Xynos ID, Edgar AJ, Buttery LDK, Hench LL, Polak JM. Gene-expression profiling of human osteoblasts following treatment with the ionic products of bioglass® 45S5 dissolution. J Biomed Mater Res. 2001;55:151–7.10.1002/1097-4636(200105)55:2<151::AID-JBM1001>3.0.CO;2-D ↗
8. Xynos ID, Hukkanen MVJ, Batten JJ, Buttery LD, Hench LL, Polak JM. Bioglass® 45S5 stimulates osteoblast turnover and enhance bone formation in vitro: implications and applications for bone tissue engineering. Calcif Tissue Int. 2000;67:321–9.10.1007/s002230001134 ↗
9. Matsuda T, Davies JE. The in vitro response of osteoblasts to bioactive glass. Biomaterials. 1987;8:275–84.10.1016/0142-9612(87)90115-3 ↗
10. Vrouwenvelder WCA, Groot CG, de Groot K. Behavior of fetal rat osteoblasts cultured in vitro on bioactive glass and nonreactive glasses. Biomaterials. 1992;13:382–92.10.1016/0142-9612(92)90044-O ↗
11. Vrouwenvelder WCA, Groot CG, de Groot K. Histological and biochemical evaluation of osteoblasts cultured on bioactive glass, hydroxylapatite, titanium alloy, and stainless steel. J Biomed Mater Res. 1993;27:465–75.10.1002/jbm.820270407 ↗
12. Price N, Bendall SP, Frondoza C, Jinnah RH, Hungerford DS. Human osteoblast-like cells (MG63) proliferate on a bioactive glass surface. J Biomed Mater Res. 1997;37:394–400.10.1002/(SICI)1097-4636(19971205)37:3<394::AID-JBM10>3.0.CO;2-C ↗
13. Loty C, Sautier JM, Tan MT, Oboeuf M, Jallot E, Boulekbache H, Greenspan D, Forest N. Bioactive glass stimulates in vitro osteoblast differentiation and creates a favorable template for bone tissue formation. J Bone Miner Res. 2001;16:231–9.10.1359/jbmr.2001.16.2.231 ↗
14. Xynos ID, Edgar AJ, Buttery LDK, Hench LL, Polak JM. Ionic products of bioactive glass dissolution increase proliferation of human osteoblasts and induce insulin-like growth factor II mRNA expression and protein synthesis. Biochem Biophys Res Commun. 2000;276:461–5.10.1006/bbrc.2000.3503 ↗
15. Sun JY, Yang YS, Zhong JP, Greenspan DC. The effect of the ionic products of bioglass® dissolution on human osteoblasts growth cycle in vitro. J Tissue Eng Regen Med. 2007;1:281–6.10.1002/term.34 ↗
16. Oonishi H, Kushitani S, Yasukawa E, Iwaki H, Hench LL, Wilson J, Tsuji E, Sugihara T. Particulate bioglass compared with hydroxyapatite as a bone graft substitute. Clin Orthop Relat Res. 1997;334:316–25.10.1097/00003086-199701000-00041 ↗
17. Chou L, Al-Bazie S, Cottrell D, Giordano R, Nathason D. Atomic and molecular mechanisms underlying the osteogenic effects of bioglass materials. Bioceramics. 1998;11:265–8.
18. Wheeler DL, Eschbach EJ, Hoellrich RG, Montfort MJ, Chamberland DL. Assessment of resorbable bioactive material for grafting of critical-size cancellous defects. J Orthop Res. 2000;18:140–8.10.1002/jor.1100180120 ↗
19. Moreira-Gonzalez A, Lobocki C, Barakat K, Andrus L, Bradford M, Gilsdorf M, Jackson IT. Evaluation of 45S5 bioactive glass combined as a bone substitute in the reconstruction of critical size calvarial defects in rabbits. J Craniofac Surg. 2005;16:63–70.10.1097/00001665-200501000-00013 ↗
20. Amato MM, Blaydon SM, Scribbick FW Jr, Belden CJ, Shore JW, Neuhaus RW, Kelley PS, Holck DE. Use of bioglass for orbital volume augmentation in enophthalmos: a rabbit model (oryctolagus cuniculus). Ophthal Plast Reconstr Surg. 2003;19:455–65.10.1097/01.IOP.0000092795.83665.FD ↗
21. Kobayashi H, Turner AS, Seim HB 3rd, Kawamoto T, Bauer TW. Evaluation of a silica-containing bone graft substitute in a vertebral defect model. J Biomed Mater Res A. 2010;92(2):596–603.10.1002/jbm.a.32397 ↗