The objective of this study was to compare the benefits of titanium 2448 (Ti2448) half-pin and titanium-6 aluminium-4 vanadium (TAV) half-pin on reducing pin loosening during external fracture fixation. Although having similar strength, Ti2448 half-pin had even lower elastic modules (33 GPa) when compared with TAV half-pin (110 GPa), which was similar to that of cortical bone (20 GPa). In the external fixation of tibial model fractures and canine cadaveric tibia fractures, Ti2448 half-pin had greater recoverable deformation and less stress concentration at the pin-bone interface in compression, torsion, and four-points bending test. Then, tibial fractures were created in 24 dogs and stabilized with four half-pins of either Ti2448 or TAV in each animal. At 4 and 8 weeks postoperatively, fracture healing and pin loosening was assessed by radiographic grading scale. The scores of Ti2448 group were significantly higher than those of TAV group. Micro-CT analysis also indicated larger quantity and higher quality of newly formed bone at pin-bone interface in Ti2448 group. Histology observation showed the newly formed bone integrated well into the threads of Ti2448 half-pins. In contrast, there was a layer of necrotic tissue between the bone tissue and TAV half-pin at pin-bone interface in TAV group. The extraction torque values of Ti2448 half-pins near the fracture line were significantly higher than those TAV pins. In conclusion, the Ti2448 half-pin with low elastic modulus could enhance osseointegration and reduce pin loosening when compared with TAV half-pin. It is a promising biomaterial for constructing external fixation system in clinical application.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopedic Surgery and Rehabilitation · Orthopaedic implants and arthroplasty · Titanium Alloys Microstructure and Properties
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (31)
Show source-provided references
1. Egger EL. Complications of external fixation. A problem-oriented approach. Vet Clin North Am Small Anim Pract. 1991;21:705–33.10.1016/S0195-5616(91)50080-2 ↗
2. Green SA. Complications of external skeletal fixation. Clin Orthop Relat Res. 1983;180:109–16.
3. Harari J. Complications of external skeletal fixation. Vet Clin North Am Small Anim Pract. 1992;22:99–107.10.1016/S0195-5616(92)50006-7 ↗
4. Pilliar RM, Lee JM, Maniatopoulos C. Observations on the effect of movement on bone ingrowth into porous-surfaced implants. Clin Orthop Relat Res. 1986;208:108–13.10.1097/00003086-198607000-00023 ↗
5. Oberli H, Frigg R, Schenk R. External fixator: surgical technique, pinless fixator, change in procedure. Helv Chir Acta. 1994;60:1073–80.
6. Hipp JA, Edgerton BC, An KN, Hayes WC. Structural consequences of transcortical holes in long bones loaded in torsion. J Biomech. 1990;23:1261–8.10.1016/0021-9290(90)90383-E ↗
7. Edgerton BC, An KN, Morrey BF. Torsional strength reduction due to cortical defects in bone. J Orthop Res. 1990;8:851–5.10.1002/jor.1100080610 ↗
8. Giotakis N, Narayan B. Stability with unilateral external fixation in the tibia. Strategies Trauma Limb Reconstr. 2007;2:13–20.10.1007/s11751-007-0011-y ↗
9. Moroni A, Faldini C, Marchetti S, Manca M, Consoli V, Giannini S. Improvement of the bone-pin interface strength in osteoporotic bone with use of hydroxyapatite-coated tapered external-fixation pins: a prospective, randomized clinical study of wrist fractures. J Bone Joint Surg Am. 2001;83:717–21.10.2106/00004623-200105000-00010 ↗
10. Schmidmaier G, Lucke M, Wildemann B, Haas NP, Raschke M. Prophylaxis and treatment of implant-related infections by antibiotic-coated implants: a review. Injury. 2006;37(suppl 2):S105–12.10.1016/j.injury.2006.04.016 ↗
11. Moroni A, Heikkila J, Magyar G, Toksvig-Larsen S, Giannini S. Fixation strength and pin tract infection of hydroxyapatite-coated tapered pins. Clin Orthop Relat Res. 2001;388:209–17.10.1097/00003086-200107000-00029 ↗
12. Safinia L, Datan N, Hohse M, Mantalaris A, Bismarck A. Towards a methodology for the effective surface modification of porous polymer scaffolds. Biomaterials. 2005;26:7537–47.10.1016/j.biomaterials.2005.05.078 ↗
13. Campbell AA, Song L, Li XS, Nelson BJ, Bottoni C, Brooks DE, DeJong ES. Development, characterization, and anti-microbial efficacy of hydroxyapatite-chlorhexidine coatings produced by surface-induced mineralization. J Biomed Mater Res. 2000;53:400–7.10.1002/1097-4636(2000)53:4<400::AID-JBM14>3.0.CO;2-Z ↗
14. Oni OO, Capper M, Soutis C. Factors which may increase stresses at the pin–bone interface in external fixation: a finite element analysis study. Afr J Med Med Sci. 1999;28:13–5.
15. Pieske O, Geleng P, Zaspel J, Piltz S. Titanium alloy pins versus stainless steel pins in external fixation at the wrist: a randomized prospective study. J Trauma. 2008;64:1275–80.10.1097/TA.0b013e31815e40e0 ↗
16. Hao YL, Li SJ, Sun SY, Zheng CY, Yang R. Elastic deformation behaviour of Ti-24Nb-4Zr-7.9Sn for biomedical applications. Acta Biomater. 2007;3:277–86.10.1016/j.actbio.2006.11.002 ↗
17. Fricka KB, Mahar AT, Lee SS. Biomechanical analysis of antegrade and retrograde flexible intramedullary nail fixation of pediatric femoral fractures using a synthetic bone model. J Pediatr Orthop. 2004;24:167–71.10.1097/01241398-200403000-00006 ↗
18. Yang CY, Simmons DJ, Lozano R. The healing of grafts combining freeze-dried and demineralized allogeneic bone in rabbits. Clin Orthop Relat Res. 1994;298:286–95.
19. Li M, Liu X, Liu X, Ge B. Calcium phosphate cement with BMP-2-loaded gelatin microspheres enhances bone healing in osteoporosis: a pilot study. Clin Orthop Relat Res. 2010;468:1978–85.10.1007/s11999-010-1321-9 ↗
20. Ganser A, Thompson R, Tami I, Neuhoff D, Steiner A, Ito K. An in vivo experimental comparison of stainless steel and titanium schanz screws for external fixation. Eur J Trauma Emerg Surg. 2007;33:59–68.10.1007/s00068-007-6053-5 ↗
21. Pettine KA, Chao EY, Kelly PJ. Analysis of the external fixator pin–bone interface. Clin Orthop Relat Res. 1993;293:18–27.10.1097/00003086-199308000-00004 ↗
22. Yang CY, Simmons DJ, Lozano R. The healing of grafts combining freeze-dried and demineralized allogeneic bone in rabbits. Clin Orthop Relat Res. 1994;298:286e95.
23. Wang L, Fan H, Zhang ZY, Lou AJ, Pei GX, Jiang S, Mu TW, Qin JJ, Chen SY, Jin D. Osteogenesis and angiogenesis of tissue-engineered bone constructed by prevascularized β-tricalcium phosphate scaffold and mesenchymal stem cells. Biomaterials. 2010;31:9452–61.10.1016/j.biomaterials.2010.08.036 ↗
24. Moroni A, Vannini F, Mosca M, Giannini S. State of the art review: techniques to avoid pin loosening and infection in external fixation. J Orthop Trauma. 2002;16(3):189–95.10.1097/00005131-200203000-00009 ↗
25. Parameswaran AD, Roberts CS, Seligson D, Voor M. Pin tract infection with contemporary external fixation: how much of a problem? J Orthop Trauma. 2003;17(7):503–7.10.1097/00005131-200308000-00005 ↗
26. Schatzker J, Horne JG, Sumner-Smith G. The effect of movement on the holding power of screws in bone. Clin Orthop. 1975;111:257–62.10.1097/00003086-197509000-00032 ↗
27. Piza G, Caja VL, Gonzalez-Viejo MA, Navarro A. Hydroxyapatite-coated external-fixation pins. The effect on pin loosening and pin-track infection in leg lengthening for short stature. J Bone Joint Surg Br. 2004;86:892–7.10.1302/0301-620X.86B6.13875 ↗
28. Caja VL, Moroni A. Hydroxyapatite coated external fixation pins: an experimental study. Clin Orthop Relat Res. 1996;325:269–75.10.1097/00003086-199604000-00033 ↗
29. Neuhoff D, Thompson RE, Frauchiger VM, Ganser A, Steiner A, Ito K. Anodic plasma chemical treatment of titanium Schanz screws reduces pin loosening. J Orthop Trauma. 2005;19:543–50.10.1097/01.bot.0000164592.96724.a9 ↗
30. Partale K et al. Poly(d, l-lactide) coating is capable of enhancing osseous integration of Schanz screws in the absence of infection. J Biomed Mater Res B Appl Biomater. 2005;74:608–16.10.1002/jbm.b.30254 ↗
31. Hyldahl C, Pearson S, Tepic S, Perren SM. Induction and prevention of pin loosening in external fixation: an in vivo study on sheep tibiae. J Orthop Trauma. 1991;5:485–92.10.1097/00005131-199112000-00017 ↗