journal article·2009-01Open access

Corrosion resistance evaluation of a Ca- and P-based bioceramic thin coating in Ti-6Al-4V

Paulo G. Coelho, Sérgio Luiz de Assis, Isolda Costa, Van P. Thompson

Abstract

Unlabelled The objective of this study was to physico/chemically characterize and determine the corrosion resistance of a Calcium-Phosphate (Ca-P) based bioceramic thin coating processed by a sputtering process on titanium alloy (Ti-6Al-4V). The samples utilized in this study were uncoated and coated disks of 10 mm diameter by 3 mm thickness. The coating was characterized by SEM, XPS + ion beam milling (IBM), thin-film mode XRD, and atomic force microscope (AFM) (n = 3). Coated and uncoated Ti-6Al-4V disk surfaces were tested in Phosphate Buffered Saline (PBS) at 25 degrees C through an area of 0.79 cm(2). A three-electrode cell set-up was used with a saturated calomel electrode (SCE) and a platinum wire as reference and counter electrodes. After 3, 17, and 25 days of immersion, electrochemical impedance spectroscopy (EIS) experiments were performed (n = 3). The EIS tests were carried out in potentiostatic mode at the open circuit potential (OCP). The frequency range considered was from 100 kHz to 10 mHz, using 10 mV root mean square as the amplitude of the perturbation signal. A potentiodynamic polarization scan using a frequency response analyzer potentiostat, was acquired following 3 days of immersion in PBS. The potentiodynamic polarization scans (n = 3) were carried out with a scan rate of 1 mV/s ranging from -0.8V(SCE) to 3.0V(SCE). Results The physico/chemical characterization showed an amorphous Ca- and P-based coating of approximately 400-700 nm thickness with Ca-P nanometer size particles embedded in a Ca-P matrix. The Bode phase angle diagrams showed highly capacitive results at low and medium frequencies for both surfaces tested. The polarization curves showed low current densities at the corrosion potential (E (corr)), in the order of 10(-8)A/cm(2), typical of passive materials with protective surface films. Coated sample current densities were comparable to the uncoated samples. Conclusion Coated and uncoated samples were stable in the test solution with a protective film maintained throughout the 25 day immersion test period.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Titanium Alloys Microstructure and Properties · Bone Tissue Engineering Materials · Hydrogen embrittlement and corrosion behaviors in metals

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (41)

Show source-provided references
1. J.E. Lemons, Biomaterials, biomechanics, tissue healing, and immediate-function dental implants. J. Oral. Implantol. 30, 318 (2004). doi: 10.1563/0712.110.1563/0712.1 ↗
2. J.E. Davies, Bone bonding at natural and biomaterial surfaces. Biomaterials 28, 5058 (2007). doi: 10.1016/j.biomaterials.2007.07.04910.1016/j.biomaterials.2007.07.049 ↗
3. T. Albrektsson, A. Wennerberg, Oral implant surfaces: Part 1—review focusing on topographic and chemical properties of different surfaces and in vivo responses to them. Int. J. Prosthodont. 17, 536 (2004)
4. T. Albrektsson, A. Wennerberg, Oral implant surfaces: Part 2—review focusing on clinical knowledge of different surfaces. Int. J. Prosthodont. 17, 544 (2004)
5. E. Eisenbarth, D. Velten, M. Muller, R. Thull, J. Breme, Biocompatibility of beta-stabilizing elements of titanium alloys. Biomaterials 25, 5705 (2004). doi: 10.1016/j.biomaterials.2004.01.02110.1016/j.biomaterials.2004.01.021 ↗
6. E. Eisenbarth, D. Velten, K. Schenk-Meuser, P. Linez, V. Biehl, H. Duschner et al., Interactions between cells and titanium surfaces. Biomol. Eng. 19, 243 (2002). doi: 10.1016/S1389-0344(02)00032-110.1016/S1389-0344(02)00032-1 ↗
7. T. Hanawa, M. Kaga, Y. Itoh, T. Echizenya, H. Oguchi, M. Ota, Cytotoxicities of oxides, phosphates and sulphides of metals. Biomaterials 13, 20 (1992). doi: 10.1016/0142-9612(92)90089-710.1016/0142-9612(92)90089-7 ↗
8. W.R. Lacefield, Hydroxyapatite coatings. Ann. N. Y. Acad. Sci. 523, 72 (1988). doi: 10.1111/j.1749-6632.1988.tb38501.x10.1111/j.1749-6632.1988.tb38501.x ↗
9. W.R. Lacefield, Current status of ceramic coatings for dental implants. Implant. Dent. 7, 315 (1998). doi: 10.1097/00008505-199807040-0001010.1097/00008505-199807040-00010 ↗
10. Y. Yang, K.H. Kim, J.L. Ong, A review on calcium phosphate coatings produced using a sputtering process—an alternative to plasma spraying. Biomaterials 26, 327 (2005). doi: 10.1016/j.biomaterials.2004.02.02910.1016/j.biomaterials.2004.02.029 ↗
11. J.L. Ong, D.L. Carnes, K. Bessho, Evaluation of titanium plasma-sprayed and plasma-sprayed hydroxyapatite implants in vivo. Biomaterials 25, 4601 (2004). doi: 10.1016/j.biomaterials.2003.11.05310.1016/j.biomaterials.2003.11.053 ↗
12. J. Lemons, F. Dietch-Misch, in Biomaterials for Dental Implants, ed. by C.E. Misch. Contemporary Implant Dentistry (Mosby Inc., Saint Louis, 1999), p. 271
13. J. Kay, Calcium phosphate coatings for dental implants. Dent. Clin. North Am. 36, 1 (1992)10.1016/S0011-8532(22)02451-X ↗
14. K.K.C. deGroot, J.G.C. Wolke, J.M. deBieck-Hogervorst, Plasma-sprayed coating of calcium phosphate, in Handbook of Bioactive Ceramics, Calcium Phosphate and Hydroxyapatite Ceramics, vol. II, ed. by T.H.L. Yamamuro, J. Wilson (CRC Press, Boca Raton, 1990), p. 17
15. Y.S. Park, K.Y. Yi, I.S. Lee, C.H. Han, Y.C. Jung, The effects of ion beam-assisted deposition of hydroxyapatite on the grit-blasted surface of endosseous implants in rabbit tibiae. Int. J. Oral Maxillofac. Implants 20, 31 (2005)
16. P.G. Coelho, J.E. Lemons, IBAD nanothick bioceramic incorporation on metallic implants for bone healing enhancement. From physico/chemical characterization to in-vivo performance evaluation. Nanotech 2005, 1, 316 (2005) ISBN, editor, Anaheim, CA
17. R.Z. LeGeros, S. Lin, R. Rohanizadeh, D. Mijares, J.P. LeGeros, Biphasic calcium phosphate bioceramics: preparation, properties and applications. J. Mater. Sci. Mater. Med. 14, 201 (2003). doi: 10.1023/A:102287242133310.1023/A:1022872421333 ↗
18. J. Pan, C. Leygraf, D. Thierry, A.M. Ektessabi, Corrosion resistance for biomaterial applications of TiO2 films deposited on titanium and stainless steel by ion-beam-assisted sputtering. J. Biomed. Mater. Res. 35, 309 (1997). doi:10.1002/(SICI)1097-4636(19970605)35:3<309::AID-JBM5>3.0.CO;2-L10.1002/(SICI)1097-4636(19970605)35:3<309::AID-JBM5>3.0.CO;2-L ↗
19. J. Pan, D. Thierry, C. Leygraf, Electrochemical and XPS studies of titanium for biomaterial applications with respect to the effect of hydrogen peroxide. J. Biomed. Mater. Res. 28, 113 (1994). doi: 10.1002/jbm.82028011510.1002/jbm.820280115 ↗
20. J. Pan, D. Thierry, C. Leygraf, Hydrogen peroxide toward enhanced oxide growth on titanium in PBS solution: blue coloration and clinical relevance. J. Biomed. Mater. Res. 30, 393 (1996). doi:10.1002/(SICI)1097-4636(199603)30:3<393::AID-JBM14>3.0.CO;2-L10.1002/(SICI)1097-4636(199603)30:3<393::AID-JBM14>3.0.CO;2-L ↗
21. P.G. Coelho, J.E. Lemons, Physico/chemical characterization and in vivo evaluation of nanothickness bioceramic depositions on alumina-blasted/acid-etched Ti-6Al-4 V implant surfaces. J. Biomed. Mater. Res. A. (2008)10.1002/jbm.a.32097 ↗
22. J.P. LeGeros, R.Z. LeGeros, A. Burgess, B. Edwards, J. Zitelli, in X-Ray Diffraction Method for the Quantitative Characterization of Calcium Phosphate Coatings. ed. by E. Horowitz, J.A. Parr. Characterization and Performance of Calcium Phosphate Coatings for Implants, vol 1196 (STP: ASTM, Philadelphia, 1994), p. 3310.1520/STP25181S ↗
23. Y.C. Tsui, C. Doyle, T.W. Clyne, Plasma sprayed hydroxyapatite coatings on titanium substrates. Part 2: optimisation of coating properties. Biomaterials 19, 2031 (1998). doi: 10.1016/S0142-9612(98)00104-510.1016/S0142-9612(98)00104-5 ↗
24. R.J. Furlong, J.F. Osborn, Fixation of hip prostheses by hydroxyapatite ceramic coatings. J. Bone Joint Surg. Br. 5, 741 (1991)10.1302/0301-620X.73B5.1654336 ↗
25. R.Z. LeGeros, Properties of osteoconductive biomaterials: calcium phosphates. Clin. Orthop. Relat. Res. 81 (2002). doi: 10.1097/00003086-200202000-0000910.1097/00003086-200202000-00009 ↗
26. J.L. Ong, K. Bessho, D.L. Carnes, Bone response to plasma-sprayed hydroxyapatite and radiofrequency-sputtered calcium phosphate implants in vivo. Int. J. Oral Maxillofac. Implants 17, 581 (2002)
27. T.W. Bauer, J. Schils, The pathology of total joint arthroplastyI. Mechanisms of implant fixation. Skeletal Radiol. 28, 423 (1999). doi: 10.1007/s00256005054110.1007/s002560050541 ↗
28. J.L. Ong, L.C. Lucas, Post-deposition heat treatments for ion beam sputter deposited calcium phosphate coatings. Biomaterials 15, 337 (1994). doi: 10.1016/0142-9612(94)90245-310.1016/0142-9612(94)90245-3 ↗
29. J.L. Ong, C.W. Prince, L.C. Lucas, Cellular response to well-characterized calcium phosphate coatings and titanium surfaces in vitro. J. Biomed. Mater. Res. 29, 165 (1995). doi: 10.1002/jbm.82029020510.1002/jbm.820290205 ↗
30. A.E. Porter, L.W. Hobbs, V.B. Rosen, M. Spector, The ultrastructure of the plasma-sprayed hydroxyapatite-bone interface predisposing to bone bonding. Biomaterials 23, 725 (2002). doi: 10.1016/S0142-9612(01)00177-610.1016/S0142-9612(01)00177-6 ↗
31. K. Soballe, S. Overgaard, E.S. Hansen, H. Brokstedt-Rasmussen, M. Lind, C. Bunger, A review of ceramic coatings for implant fixation. J. Long. Term. Eff. Med. Implants. 9, 131 (1999)
32. L. Sun, C.C. Berndt, K.A. Gross, A. Kucuk, Material fundamentals and clinical performance of plasma-sprayed hydroxyapatite coatings: a review. J. Biomed. Mater. Res. 58, 570 (2001). doi: 10.1002/jbm.105610.1002/jbm.1056 ↗
33. J.G. Wolke, K. de Groot, J.A. Jansen, In vivo dissolution behavior of various RF magnetron sputtered Ca–P coatings. J. Biomed. Mater. Res. 39, 524 (1998). doi:10.1002/(SICI)1097-4636(19980315)39:4<524::AID-JBM3>3.0.CO;2-L10.1002/(SICI)1097-4636(19980315)39:4<524::AID-JBM3>3.0.CO;2-L ↗
34. J.G. Wolke, J.P. van der Waerden, H.G. Schaeken, J.A. Jansen, In vivo dissolution behavior of various RF magnetron-sputtered Ca–P coatings on roughened titanium implants. Biomaterials 24, 2623 (2003). doi: 10.1016/S0142-9612(03)00067-X10.1016/S0142-9612(03)00067-X ↗
35. J.E.G. Gonzalez, J.C. Mirza-Rosca, Study of the corrosion behavior of titanium and some of its alloys for biomedical and dental applications. J. Electroanal. Chem. 471, 109 (1999). doi: 10.1016/S0022-0728(99)00260-010.1016/S0022-0728(99)00260-0 ↗
36. R.W. Hsu, C. Yang, C. Huang, Y. Chen, Electrochemical corrosion properties of Ti-6Al-4 V implant alloy in the biological environment. Mater. Sci. Eng. A 380, 100 (2004). doi: 10.1016/j.msea.2004.03.06910.1016/j.msea.2004.03.069 ↗
37. J. Pan, D. Thierry, C. Leygraf, Electrochemical impedance spectroscopy study of the passive oxide film on titanium for implant application. Electrochim. Acta 41, 1143 (1996). doi: 10.1016/0013-4686(95)00465-310.1016/0013-4686(95)00465-3 ↗
38. M.C. Advincula, D. Petersen, F. Rahemtulla, R. Advincula, J.E. Lemons, Surface analysis and biocorrosion properties of nanostructured surface sol-gel coatings on Ti6Al4 V titanium alloy implants. J. Biomed. Mater. Res. B Appl. Biomater. 80, 107 (2007). doi: 10.1002/jbm.b.3057510.1002/jbm.b.30575 ↗
39. I.C. Lavos-Valereto, I. Costa, S. Wolynec, The electrochemical behavior of Ti-6Al-7Nb alloy with and without plasma-sprayed hydroxyapatite coating in Hank’s solution. J. Biomed. Mater. Res. 63, 664 (2002). doi: 10.1002/jbm.1035110.1002/jbm.10351 ↗
40. P. Tengvall, A. Askendal, I. Lundstrom, H. Elwing, Studies of surface activated coagulation: antisera binding onto methyl gradients on silicon incubated in human plasma in vitro. Biomaterials 13, 367 (1992). doi: 10.1016/0142-9612(92)90042-M10.1016/0142-9612(92)90042-M ↗
41. P. Tengvall, I. Lundstrom, L. Sjoqvist, H. Elwing, L.M. Bjursten, Titanium-hydrogen peroxide interaction: model studies of the influence of the inflammatory response on titanium implants. Biomaterials 10, 166 (1989). doi: 10.1016/0142-9612(89)90019-710.1016/0142-9612(89)90019-7 ↗