journal article·2018-01

Three cases of metallosis associated with spine instrumentation

Reed Ayers, Mackenzie Miller, Jeffery Schowinsky, Evalina Burger, Vikas Patel, Christopher Kleck

Publisher page

Abstract

The characteristic trait of metallosis is the presence of tissue staining. Analyzing explanted revision hardware from spine surgeries and performing ICPMS/AES analysis on removed tissue samples, a clinically relevant definition for metallosis may be developed. Results of the analysis identified hardware fretting wear and corrosion, and ICP-MS/AES analysis identified elevated metal ion concentrations in all cases. This supports the hypothesis that corrosion may be linked to poor health outcomes and potential need for revision surgery. Using failure analysis methods, corrosion products from orthopedic spine implants can be identified to begin to characterize metallosis in a clinically relevant manner. Failure analysis for patients undergoing revision spine surgery for other causes. Using failure analysis methods we conducted a retrieval analysis of explanted hardware and tissue. Implant failure with surrounding tissue metal staining is a current issue in the field of orthopedics. Specifically in spine, this issue is under reported and over looked as a clinically significant finding. Metallosis is most commonly used to describe the presence of tissue staining however this is not a clinically relevant definition. There is a need for further research to provide a better understanding of metallosis leading to better patient outcomes. Patients were screened for this study during a radiological review prior to surgery. All explanted hardware was documented and inspected for signs of wear and corrosion using non-destructive testing. A tissue sample that is normally removed and discarded was collected for ICP-MS/AES analysis. The presence of fretting corrosion, galling and corrosion fatigue was found on all explanted hardware. Cr levels are significantly higher than what was previously published as normal in muscle 0.03 μg/g. One case was a 4 order of magnitude elevation with the other 2 approximately 2 orders of magnitude increase. Titanium and Co concentrations are also significantly increased. The Ca to P ratio for all samples is approximately 1.7:1 suggesting some form of apatitic crystal present due to drying of the tissues. In all cases, the Al, Mo, V, Co, content in surrounding tissue is significantly elevated >10× above "normal," 8.4 ± 4.8; 1.61 ± 1.41; 0.06 ± 0.03; 1.35 ± 1.97 μg/g respectively. A "normal" reference titanium level could only be found for whole blood, 0.00072 ± 0.1412 μ/g. Iron and Ni measurements are within typical values presented in previous studies. No single mechanism for the release of metal ion in a patient is clear. Evidence suggests a tribocorrosive process due to both wear and environmental factors. Specific biologic mechanisms, such as the possible presence of bacteria may affect the fretting and corrosion of spinal instrumentation must be explored in conjunction with thorough review of patient outcomes. Such an effort can potentially reduce patient readmission and increase successful patient outcomes.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Titanium Alloys Microstructure and Properties · Orthopedic Infections and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (35)

Show source-provided references
1. Wynn-Jones H, Macnair R, Wimhurst J, Chirodian N, Derbyshire B, Toms A, Cahir J. Silent soft tissue pathology is common with a modern metal-on-metal hip arthroplasty. Acta Orthop. 2011;82:301–7.10.3109/17453674.2011.579518 ↗
2. Heffernan EJ, Alkubaidan FO, Nielsen TO, Munk PL. The imaging appearances of metallosis. Skelet Radiol. 2008;37:59–62.10.1007/s00256-007-0362-y ↗
3. Morteza H, Wollert T, Langford GM, Gilbert JL. Electrochemical control of cell death by reduction-induced intrinsic apoptosis and oxidation-induced necrosis on CoCrMo alloy in vitro. Biomaterials. 2012;33:6295–304.10.1016/j.biomaterials.2012.05.054 ↗
4. Morteza H, Gilbert JL. Study of cellular dynamics on polarized CoCrMo alloy using time-lapse live-cell imaging. Acta Biomater. 2013;9:220–8.
5. Nadim H. Metal sensitivity in patients with orthopedic implants. J Clin Rheumatol. 2001;7:215–18.10.1097/00124743-200108000-00004 ↗
6. Elena F, Pigatto PD, Guzzi G, Altomare G. Metal sensitivity in patients with orthopaedic implants: a prospective study. Contact Dermat. 2011;64:273–79.10.1111/j.1600-0536.2011.01886.x ↗
7. Elves MW, Wilson JN, Scales JT, Kemp HB. Incidence of metal sensitivity in patients with total joint replacements. BMJ. 1975;4:376–78.10.1136/bmj.4.5993.376 ↗
8. Cousen PJ, Gawkrodger DJ. Metal allergy and second-generation metal-on-metal arthroplasties. Contact Dermat. 2012;66:55–62.10.1111/j.1600-0536.2011.01970.x ↗
9. Watters TS, Cardona DM, Menon KS, Vinson EN, Bolognesi MP, Dodd LG. Aseptic lymphocyte-dominated vasculitis-associated lesion. Am J Clin Pathol. 2010;134:886–93.10.1309/AJCPLTNEUAH8XI4W ↗
10. del Rio J, Beguiristain J, Duart J. Metal levels in corrosion of spinal implants. Eur Spine J. 2007;16:1055–61.10.1007/s00586-007-0311-4 ↗
11. Goldenberg Y, Tee JW, Salinas-La Rosa CM, Murphy M. Spinal metallosis: a systematic review. Eur Spine J. 2016;25:1467–73.10.1007/s00586-015-4347-6 ↗
12. Takahashi S, Delécrin J, Passuti N. Intraspinal metallosis causing delayed neurologic symptoms after spinal instrumentation surgery. Spine. 2001;26:1495–9.10.1097/00007632-200107010-00024 ↗
13. Tezer M, Kuzgun U, Hamzaoglu A, Ozturk C, Kabukcuoglu F, Sirvanci M. Intraspinal metalloma resulting in late paraparesis. Arch Orthop Trauma Surg. 2005;125:417–21.10.1007/s00402-005-0802-x ↗
14. Kmieć K, Synder M, Kozłowski P, Drobniewski M, Sibiński M. Metal debris concentrations in soft tissues adjacent to loosened femoral stems is higher in uncemented than cemented implants. BMC Musculoskelet Disord. 2014;15:267–71.10.1186/1471-2474-15-267 ↗
15. Schroeder HA. The role of chromium in mammalian nutrition. Am J Clin Nutr. 1968;21:230–44.10.1093/ajcn/21.3.230 ↗
16. Vincent JB. The biochemistry of chromium. J Nutr. 2000;130:715–8.10.1093/jn/130.4.715 ↗
17. Pasha Q, Malik SA, Iqbal J, Shaheen N, Shah MH. Comparative evaluation of trace metal distribution and correlation in human malignant and benign breast tissues. Biol Trace Elem Res. 2008;125:30–40.10.1007/s12011-008-8158-z ↗
18. Farah IO, Nguyen PX, Arslan Z, Ayensu W, Cameron JA. Significance of differential metal loads in normal versus cancerous cadaver tissues. Biomed Sci Instrum. 2010;46:404–9.
19. Harrington JM, Young DJ, Essader AS, Sumner SJ, Levine KE. Analysis of human serum and whole blood for mineral content by ICP-MS and ICP-OES: development of a mineralomics method. Biol Trace Elem Res. 2014;160:132–42.10.1007/s12011-014-0033-5 ↗
20. Ayers RA, Burger EL, Kleck CJ, Patel VV. Metallurgy of spinal instrumentation. In: Niinomi M, Narushima T, Nakai M, editors. Advances in metallic biomaterials technology. Springer series in biomaterials science and engineering. Heidelberg: Springer; 2015. p. 53–70.
21. Gilbert JL. Mechanically assisted corrosion of metallic biomaterials, corrosion: environments and industries, Vol 13C, ASM Handbook, ASM International; 2006. p. 826–36.10.31399/asm.hb.v13c.a0004206 ↗
22. Bundy KJ, Zardiakas LD. Corrosion fatigue and stress corrosion cracking in metallic biomaterials, corrosion: environments and industries, Vol 13C, ASM Handbook, ASM International; 2006. p. 853–90.10.31399/asm.hb.v13c.a0004208 ↗
23. Serhan H, Slivka M, Albert T, Kwak SD. Is galvanic corrosion between titanium alloy and stainless steel spinal implants a clinical concern? Spine J. 2004;4:379–87.10.1016/j.spinee.2003.12.004 ↗
24. Oh K-T, Kim K-N. Electrochemical properties of suprastructures galvanically coupled to a titanium implant. Biomed Mater Res Part B: Appl Biomater. 2004;70B:318–31.10.1002/jbm.b.30046 ↗
25. Antunes RA, Lopes de Oliveira MC. Corrosion fatigue of biomedical metallic alloys: mechanisms and mitigation. Acta Biomater. 2012;8:937–62.10.1016/j.actbio.2011.09.012 ↗
26. Waterhouse RB, Taylor DE. Fretting debris and the delamination theory of wear. Wear. 1974;29:337–44.10.1016/0043-1648(74)90019-2 ↗
27. Waterhouse B. The role of adheshion and delamination in the fretting wear of metallic materials. Wear. 1977;45:355–64.10.1016/0043-1648(77)90026-6 ↗
28. Jacobs JJ, Gilbert JL, Urban RM. Current concepts review—corrosion of metal orthopaedic implants. J Bone Joint Surg Am. 1998;80:268–82.10.2106/00004623-199802000-00015 ↗
29. Gilbert JL, Mali SA. Medical implant corrosion: electrochemistry at metallic biomaterial surfaces. In: Elaz N, editor. Degradation of Implant Materials. Lausanne: Springer; 2012. p. 1–28.
30. Morita M, Sasada T, Hayashi H, Tsukamoto Y. The corrosion fatigue properties of surgical implants in a living body. J Biomed Mater Res. 1988;22:529–40.10.1002/jbm.820220608 ↗
31. Bhola R, Bhola SM, Mishra B, Olson DL. Corrosion in titanium dental implants/prostheses—a review. Trends Biomater Artif Organs. 2011;25:34–46.
32. Ganesh BKC, Ramaniah N, Chandrasekhar Rao PV. Effect of heat treatment on dry sliding wear of titanium aluminum-vanadium (Ti-6Al-4V) implant alloy. J Mech Eng Res. 2012;4:67–74.
33. Ribeiro M, Monteiro FJ, Ferraz MP. Infection of orthopedic implants with emphasis on bacterial adhesion process and techniques used in studying bacterial-material interactions. Biomatter. 2012;2:176–94.10.4161/biom.22905 ↗
34. Tunney MM, Patrick S, Curran MD, Ramage G, Hanna D, Nixon JR, Gorman SP, Davis RI, Anderson N. Detection of prosthetic hip infection at revision arthroplasty by immunofluorescence microscopy and PCR amplification of the bacterial 16S rRNA gene. J Clin Microbiol. 1999;37:3281–90.10.1128/JCM.37.10.3281-3290.1999 ↗
35. Shiono Y, Ishii K, Nagai S, Kakinuma H, Sasaki A, Funao H, Kuramoto T, Yoshioka, Ishihama H, Isogai N, Kenichiro K, Tsuji T, Okada Y, Koyasu S, Nakamura M, Toyama Y, Aizawa M, Matsumoto. Delayed propionibacterium acnes surgical site infections occur only in the presence of an implant. Sci Rep. 2016;6:1–10.10.1038/srep32758 ↗