How do gait frequency and serum-replacement interval affect polyethylene wear in knee-wear simulator tests?
Source abstract
Abstract
Polyethylene wear (PE) is known to be a limiting factor in total joint replacements. However, a standardized wear test (e.g. ISO standard) can only replicate the complex in vivo loading condition in a simplified form. In this study, two different parameters were analyzed: (a) Bovine serum, as a substitute for synovial fluid, is typically replaced every 500,000 cycles. However, a continuous regeneration takes place in vivo. How does serum-replacement interval affect the wear rate of total knee replacements? (b) Patients with an artificial joint show reduced gait frequencies compared to standardized testing. What is the influence of a reduced frequency? Three knee wear tests were run: (a) reference test (ISO), (b) testing with a shortened lubricant replacement interval, (c) testing with reduced frequency. The wear behavior was determined based on gravimetric measurements and wear particle analysis. The results showed that the reduced test frequency only had a small effect on wear behavior. Testing with 1 Hz frequency is therefore a valid method for wear testing. However, testing with a shortened replacement interval nearly doubled the wear rate. Wear particle analysis revealed only small differences in wear particle size between the different tests. Wear particles were not linearly released within one replacement interval. The ISO standard should be revised to address the marked effects of lubricant replacement interval on wear rate.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Orthopedic Infections and Treatments
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (37)
Show source-provided references
1. Purdue P, Koulouvaris P, Nestor B, Sculco T. The central role of wear debris in periprosthetic osteolysis. HSS J. 2006;2(2):102–13.10.1007/s11420-006-9003-6 ↗
2. Willert HG, Semlitsch M, Peltier L. Tissue reactions to plastic and metallic wear products of joint endoprostheses. Clin Orthop Relat R. 1996;333:4–14.
3. Levick JR. Synovial fluid: determinants of volume and turnover and material concentration. In: Kettner KE, et al., editors. Articular cartilage and osteoarthritis. New York: Raven Press; 1992. p. 529–41.
4. Saari T, Tranberg R, Zügner R, Uvehammer J, Kärrholm J. Changed gait pattern in patients with total knee arthroplasty but minimal influence of tibial insert design. Acta Orthop. 2005;76(2):253–60.10.1080/00016470510030661 ↗
5. Kretzer JP, Jakubowitz E, Sonntag R, Hofmann K, Heisel C, Thomsen M. Effect of joint laxity on polyethylene wear in total knee replacement. J Biomech. 2010;43(6):1092–6.10.1016/j.jbiomech.2009.12.016 ↗
6. Niedzwiecki S, Klapperich C, Short J, Jani S, Ries M, Pruitt L. Comparison of three joint simulator wear debris isolation techniques: acid digestion, base digestion, and enzyme cleavage. J Biomed Mater Res. 2001;56(2):245–9.10.1002/1097-4636(200108)56:2<245::AID-JBM1091>3.0.CO;2-T ↗
7. Kretzer JP, Jakubowitz E, Reinders J, Lietz E, Moradi B, Hofmann K, et al. Wear analysis of unicondylar mobile bearing and fixed bearing knee systems: a knee simulator study. Acta Biomater. 2011;7(2):710–5.10.1016/j.actbio.2010.09.031 ↗
8. Fisher J, Dowson D. Tribology of total artificial joints. Proc Inst Mech Eng H. 1991;205(2):73–9.10.1243/PIME_PROC_1991_205_271_02 ↗
9. Jin ZM, Medley JB, Dowson D. Fluid film lubrication in artificial hip joints. In: Dowson D, Priest M, Dalmaz G, Lubrecht AA, editors. Tribology series. Amsterdam: Elsevier; 2003. p. 237–56.
10. Sonntag R, Reinders J, Rieger JS, Heitzmann DWW, Kretzer JP. Hard-on-hard lubrication in the artificial hip under dynamic loading conditions. PLoS ONE. 2013;8(8):71622.10.1371/journal.pone.0071622 ↗
11. Wright KWJ, Dobbs HS, Scales JT. Wear studies on prosthetic materials using the pin-on-disc machine. Biomaterials. 1982;3(1):41–8.10.1016/0142-9612(82)90060-6 ↗
12. Fisher J, Dowson D, Hamdzah H, Lee HL. The effect of sliding velocity on the friction and wear of UHMWPE for use in total artificial joints. Wear. 1994;175(1–2):219–25.10.1016/0043-1648(94)90185-6 ↗
13. Schwenke T, Borgstede LL, Schneider E, Andriacchi TP, Wimmer MA. The influence of slip velocity on wear of total knee arthroplasty. Wear. 2005;259(7–12):926–32.10.1016/j.wear.2005.01.049 ↗
14. Lancaster J. Basic mechanism of friction and wear of polymers. Plast Polym. 1973;41(156):297–305.
15. Harris W. Edge loading has a paradoxical effect on wear in metal-on-polyethylene total hip arthroplasties. Clin Orthop Relat R. 2012;470(11):3077–82.10.1007/s11999-012-2330-7 ↗
16. Kamali A, Hussain A, Li C, Pamu J, Daniel J, Ziaee H, et al. Tribological performance of various CoCr microstructures in metal-on-metal bearings. J Bone Joint Surg Br Vol. 2010;92-B(5):717–25.10.1302/0301-620X.92B5.23320 ↗
17. Brandt J-M, Charron KDJ, Zhao L, MacDonald SJ, Medley JB. Commissioning of a displacement-controlled knee wear simulator and exploration of some issues related to the lubricant. Proc Inst Mech Eng H. 2011;225(8):736–52.10.1177/0954411911406061 ↗
18. Brandt JM, Brière LK, Marr J, MacDonald SJ, Bourne RB, Medley JB. Biochemical comparisons of osteoarthritic human synovial fluid with calf sera used in knee simulator wear testing. J Biomed Mater Res A. 2010;94A(3):961–71.10.1002/jbm.a.32728 ↗
19. John KS. The effect of serum protein concentration on wear rates in a hip simulator. J Biomater Appl. 2010;25(2):145–59.10.1177/0885328209347962 ↗
20. Liao Y-S, Benya PD, McKellop HA. Effect of protein lubrication on the wear properties of materials for prosthetic joints. J Biomed Mater Res. 1999;48(4):465–73.10.1002/(SICI)1097-4636(1999)48:4<465::AID-JBM10>3.0.CO;2-Y ↗
21. Mazzucco DPHD, Spector MPHD. The John Charnley award paper: the role of joint fluid in the tribology of total joint arthroplasty. Clin Orthop Relat R. 2004;429:17–32.10.1097/01.blo.0000150315.47697.b9 ↗
22. Scholes SC, Unsworth A. The effects of proteins on the friction and lubrication of artificial joints. Proc Inst Mech Eng H. 2006;220(6):687–93.10.1243/09544119JEIM21 ↗
23. Schwenke T, Kaddick C, Schneider E, Wimmer MA. Fluid composition impacts standardized testing protocols in ultrahigh molecular weight polyethylene knee wear testing. Proc Inst Mech Eng H. 2005;219(6):457–64.10.1243/095441105X34392 ↗
24. Wang A, Essner A, Schmidig G. The effects of lubricant composition on in vitro wear testing of polymeric acetabular components. J Biomed Mater Res B. 2004;68B(1):45–52.10.1002/jbm.b.10077 ↗
25. Heuberger MP, Widmer MR, Zobeley E, Glockshuber R, Spencer ND. Protein-mediated boundary lubrication in arthroplasty. Biomaterials. 2005;26(10):1165–73.10.1016/j.biomaterials.2004.05.020 ↗
26. Liao YS, McKellop H, Lu Z, Campbell P, Benya P. The effect of frictional heating and forced cooling on the serum lubricant and wear of UHMW polyethylene cups against cobalt–chromium and zirconia balls. Biomaterials. 2003;24(18):3047–59.10.1016/S0142-9612(03)00148-0 ↗
27. Lu Z, McKellop H, Liao P, Benya P. Potential thermal artifacts in hip joint wear simulators. J Biomed Mater Res. 1999;48(4):458–64.10.1002/(SICI)1097-4636(1999)48:4<458::AID-JBM9>3.0.CO;2-B ↗
28. ISO 14243-1: 2009; Implants for surgery - wear of total knee-joints prostheses - part 1: loading and displacement parameters for wear-testing machines with load control and corresponding environmental conditions for test. International Organization for Standardization, Geneva.
29. ISO 14243-3: 2004; Implants for surgery—wear of total joint knee-prostheses—part 3: loading and displacement parameters for wear testing machines with displacement control and corresponding environmetal conditions for test. International Organization for Standardization, Geneva.
30. Barnett PI, Fisher J, Auger DD, Stone MH, Ingham E. Comparison of wear in a total knee replacement under different kinematic conditions. J Mater Sci Mater Med. 2001;12(10):1039–42.10.1023/A:1012894023657 ↗
31. Galvin AL, Kang L, Udofia I, Jennings LM, McEwen HMJ, Jin Z, et al. Effect of conformity and contact stress on wear in fixed-bearing total knee prostheses. J Biomech. 2009;42(12):1898–902.10.1016/j.jbiomech.2009.05.010 ↗
32. McEwen HMJ, Barnett PI, Bell CJ, Farrar R, Auger DD, Stone MH, et al. The influence of design, materials and kinematics on the in vitro wear of total knee replacements. J Biomech. 2005;38(2):357–65.10.1016/j.jbiomech.2004.02.015 ↗
33. McEwen HMJ, Fisher J, Goldsmith AAJ, Auger DD, Hardaker C, Stone MH. Wear of fixed bearing and rotating platform mobile bearing knees subjected to high levels of internal and external tibial rotation. J Mater Sci Mater Med. 2001;12(10):1049–52.10.1023/A:1012850224565 ↗
34. Jennings LM, Bell CJ, Ingham E, Komistek RD, Stone MH, Fisher J. The influence of femoral condylar lift-off on the wear of artificial knee joints. Proc Inst Mech Eng H. 2007;221(H3):305–14.10.1243/09544119JEIM215 ↗
35. Galvin AL, Jennings LM, Tipper JL, Ingham E, Fisher J. Wear and creep of highly crosslinked polyethylene against cobalt chrome and ceramic femoral heads. Proc Inst Mech Eng H. 2010;224(H10):1175–83.10.1243/09544119JEIM741 ↗
36. Malikian R, Maruthainar K, Stammers J, Cannon SR, Carrington R, Skinner JA, et al. In vivo roughening of retrieved total knee arthroplasty femoral components. Knee. 2014;21(1):278–82.10.1016/j.knee.2012.09.008 ↗
37. Elfick APD, Hall RM, Pinder IM, Unsworth A. The influence of femoral head surface roughness on the wear of ultrahigh molecular weight polyethylene sockets in cementless total hip replacement. J Biomed Mater Res. 1999;48(5):712–8.10.1002/(SICI)1097-4636(1999)48:5<712::AID-JBM17>3.0.CO;2-R ↗