journal article·2014-03

Geometric adaption of biodegradable magnesium alloy scaffolds to stabilise biological myocardial grafts. Part I

M. Bauer, T. Schilling, M. Weidling, D. Hartung, Ch. Biskup, P. Wriggers, F. Wacker, Fr. -W. Bach, A. Haverich, T. Hassel

Publisher page

Abstract

Synthetic patch materials currently in use have major limitations, such as high susceptibility to infections and lack of contractility. Biological grafts are a novel approach to overcome these limitations, but do not always offer sufficient mechanical durability in early stages after implantation. Therefore, a stabilising structure based on resorbable magnesium alloys could support the biological graft until its physiologic remodelling. To prevent early breakage in vivo due to stress of non-determined forming, these scaffolds should be preformed according to the geometry of the targeted myocardial region. Thus, the left ventricular geometry of 28 patients was assessed via standard cardiac magnetic resonance imaging (MRI). The resulting data served as a basis for a finite element simulation (FEM). Calculated stresses and strains of flat and preformed scaffolds were evaluated. Afterwards, the structures were manufactured by abrasive waterjet cutting and preformed according to the MRI data. Finally, the mechanical durability of the preformed and flat structures was compared in an in vitro test rig. The FEM predicted higher durability of the preformed scaffolds, which was proven in the in vitro test. In conclusion, preformed scaffolds provide extended durability and will facilitate more widespread use of regenerative biological grafts for surgical left ventricular reconstruction.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Magnesium Alloys: Properties and Applications · Cardiac Structural Anomalies and Repair · Aluminum Alloys Composites Properties

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (26)

Show source-provided references
1. Adhyapak SM, Parachuri VR. Architecture of the left ventricle: insights for optimal surgical ventricular restoration. Heart Fail Rev. 2010;15(1):73–83.10.1007/s10741-009-9151-0 ↗
2. Athanasuleas CL, Stanley AW, Buckberg GD, Dor V, Di DM, Siler W. Surgical anterior ventricular endocardial restoration (SAVER) for dilated ischemic cardiomyopathy. Semin Thorac Cardiovasc Surg. 2001;13(4):448–58.10.1053/stcs.2001.30372 ↗
3. Buckberg GD. Form versus disease: optimizing geometry during ventricular restoration. Eur J Cardiothorac Surg. 2006;29(Suppl 1):S238–44.10.1016/j.ejcts.2006.02.015 ↗
4. Dor V, Saab M, Coste P, Kornaszewska M, Montiglio F. Left ventricular aneurysm: a new surgical approach. Thorac Cardiovasc Surg. 1989;37(1):11–9.10.1055/s-2007-1013899 ↗
5. Dor V. Surgery for left ventricular aneurysm. Curr Opin Cardiol. 1990;5(6):773–80.10.1097/00001573-199012000-00009 ↗
6. Calafiore AM, Iaco AL, Abukoudair W, Penco M, Di MM. Left ventricular surgical remodeling after the STICH trial. Thorac Cardiovasc Surg. 2011;59(4):195–200.10.1055/s-0030-1270738 ↗
7. Badylak SF, Kochupura PV, Cohen IS, Doronin SV, Saltman AE, Gilbert TW, Kelly DJ, Ignotz RA, Gaudette GR. The use of extracellular matrix as an inductive scaffold for the partial replacement of functional myocardium. Cell Transplant. 2006;15:S29–40.10.3727/000000006783982368 ↗
8. Taheri SA, Ashraf H, Merhige M, Miletich RS, Satchidanand S, Malik C, Naughton J, Zhao Q. Myoangiogenesis after cell patch cardiomyoplasty and omentopexy in a patient with ischemic cardiomyopathy. Tex Hear Inst J. 2005;32(4):598–601.
9. Tudorache I, Kostin S, Meyer T, Teebken O, Bara C, Hilfiker A, Haverich A, Cebotari S. Viable vascularized autologous patch for transmural myocardial reconstruction. Eur J Cardiothorac Surg. 2009;36(2):306–11.10.1016/j.ejcts.2009.02.038 ↗
10. Ruel MA, Sellke FW, Bianchi C, Khan TA, Faro R, Zhang JP, Cohn WE. Endogenous myocardial angiogenesis and revascularization using a gastric submucosal patch. Ann Thorac Surg. 2003;75(5):1443–9.10.1016/S0003-4975(02)04708-2 ↗
11. Schilling T, Cebotari S, Tudorache I, Haverich A. Tissue engineering of vascularized myocardial prosthetic tissue. Biological and solid matrices. Chirurg. 2011;82(4):319–24.10.1007/s00104-010-2032-1 ↗
12. Bach F, Haverich A, Cebotari S, Biskup C, Schuster B. Supporting element for tissue implants. Patent WO 2011/101142 A1; 2011.
13. Hoffmeister BK, Handley SM, Wickline SA, Miller JG. Ultrasonic determination of the anisotropy of Young’s modulus of fixed tendon and fixed myocardium. J Acoust Soc Am. 1996;100(6):3933–40.10.1121/1.417246 ↗
14. Biskup C, Hepke M, Grittner N, Plorin T, Hassel T, Bormann D, Bach F-W, Schilling T, Hilfiker A, Cebotari S, Tudorache I, Meyer T, Haverich A AWIJ cutting of structures made of magnesium alloys for the cardiovascular surgery. American WJTA Conference and Expo; 2009.
15. Schilling T, Brandes G, Tudorache I, Cebotari S, Hilfiker A, Meyer T, Biskup C, Bauer M, Waldmann KH, Bach FW, Haverich A, Hassel T. In vivo degradation of magnesium alloy LA63 scaffolds for temporary stabilization of biological myocardial grafts in a swine model. Biomed Tech (Berl). 2013;58:407–516.
16. Menicanti L, Castelvecchio S, Ranucci M, Frigiola A, Santambrogio C, de Vincentiis C, Brankovic J, Di DM. Surgical therapy for ischemic heart failure: single-center experience with surgical anterior ventricular restoration. J Thorac Cardiovasc Surg. 2007;134(2):433–41.10.1016/j.jtcvs.2006.12.027 ↗
17. Daniels AU, Chang MK, Andriano KP. Mechanical properties of biodegradable polymers and composites proposed for internal fixation of bone. J Appl Biomater. 1990;1(1):57–78.10.1002/jab.770010109 ↗
18. Wu W, Gastaldi D, Yang K, Tan L, Petrini L, Migliavacca F. Finite element analyses for design evaluation of biodegradable magnesium alloy stents in arterial vessels. Mat Sci Eng B. 2011;176:1733–40.10.1016/j.mseb.2011.03.013 ↗
19. Grogan J, Leen S, McHugh P. Comparing coronary stent material performance on a common geometric platform through simulated bench testing. J Mech Behav Biomed Mater. 2012;12:129–38.10.1016/j.jmbbm.2012.02.013 ↗
20. Feng B, Veress A, Sitek A, Gullberg G, Roy D. Estimation of mechanical properties from gated SPECT and cine MRI data using a finite-element mechanical model of the left ventricle. IEEE Trans Nucl Sci. 2001;48:725–33.10.1109/23.940154 ↗
21. Feng L, Weixue L, Ling X, Guohua W. The construction of three-dimensional composite finite element mechanical model of human left ventricle. JSME Int J Ser C. 2001;44:125–33.10.1299/jsmec.44.125 ↗
22. Watanabe H, Sugiura S, Kafuku H, Hisada T. Multiphysics simulation of left ventricular filling dynamics using fluid-structure interaction finite element method. Biophys J. 2004;87:2074–85.10.1529/biophysj.103.035840 ↗
23. Wong J, Kuhl E. Generating fibre orientation maps in human heart models using Poisson interpolation. Comput Methods Biomech Biomed Eng 2012;1–10.10.1080/10255842.2012.739167 ↗
24. Kirkland NT, Waterman J, Birbilis N, Dias G, Woodfield TB, Hartshorn RM, Staiger MP. Buffer-regulated biocorrosion of pure magnesium. J Mater Sci Mater Med. 2012;23(2):283–91.10.1007/s10856-011-4517-y ↗
25. Kirkland NT, Birbilis N, Staiger MP. Assessing the corrosion of biodegradable magnesium implants: a critical review of current methodologies and their limitations. Acta Biomater. 2012;8(3):925–36.10.1016/j.actbio.2011.11.014 ↗
26. Witte F, Fischer J, Nellesen J, Crostack HA, Kaese V, Pisch A, Beckmann F, Windhagen H. In vitro and in vivo corrosion measurements of magnesium alloys. Biomaterials. 2006;27(7):1013–8.10.1016/j.biomaterials.2005.07.037 ↗