Load-bearing capacity of an experimental dental implant made of Nb-1Zr
Source abstract
Abstract
Abstract Although implants have undergone a remarkable development over the past decades, modern implants still show complications that make the improvement of materials necessary. The presented study investigates the load-bearing capacity of an experimental dental implant made of a niobium alloy (Nb1Zr) compared to identical implants made of Ti6Al4V using chewing simulation for artificial aging. Eight implants each with an experimental design were manufactured from Nb1Zr and Ti6Al4V. A standardized abutment made of Ti6Al4V was fabricated for each implant and screwed into the implant with a screw made of the same material. A shape-identical crown for an upper first molar was fabricated for all implants using the CAD/CAM technique. All specimens were artificially aged using chewing simulation for 1 × 10 6 cycles and thermocycling between 5 °C and 55 °C for 4 × 10 3 cycles. After that, all specimens were loaded until failure. This was followed by a 3D analysis of the deformation of the samples. 100% of the samples survived the artificial aging. The Nb1Zr samples failed at 2595 ± 1069 N. In the Ti6Al4V group, failure occurred at 2958 ± 1058 N. The first deformations occurred in both groups from a load of at least 1114 N. The 3D analysis revealed deformations of 0.08 mm in the implant shoulder area of the Nb1Zr implants and of 0.04 mm in the Ti6Al4V implants. This difference was significant ( p = 0.016). The investigated Nb1Zr alloy appears to be suitable for implants. The measured load-bearing capacity is significantly higher than the expected intraoral masticatory forces. Graphical Abstract
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Titanium Alloys Microstructure and Properties · Advanced materials and composites
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (21)
Show source-provided references
1. Poli PP, de Miranda FV, Polo TOB, Santiago Junior JF, Lima Neto TJ, Rios BR, et al. Titanium allergy caused by dental implants: a systematic literature review and case report. Materials. 2021;14:5239. https://doi.org/10.3390/ma1418523910.3390/ma14185239 ↗
2. Thiem D, Behrens E, Beuer F, Bonsmann M, Cosgarea R, Dick K, et al. Keramikimplantate - S3 Leitlinie. AWMF online - Das Portal der wissenschaftlichen Medizin 2022:AWMF-Registriernmmer: 083-039
3. O’Brien B Niobium biomaterials. In: Niinomi M, Narushima T, Nakai M, editors. Advances in metallic biomaterials. Berlin, Heidelberg: Springer; 2015. p. 245-7210.1007/978-3-662-46836-4_11 ↗
4. Olivares-Navarrete R, Olaya J, Ramirez C, Rodil S. Biocompatibility of niobium coatings. Coatings. 2011;1:72–87. https://doi.org/10.3390/coatings101007210.3390/coatings1010072 ↗
5. Rubitschek F, Niendorf T, Karaman I, Maier H. Corrosion fatigue behavior of a biocompatible ultrafine-grained niobium alloy in simulated body fluid. J Mech Behav Biomed Mater. 2012;5:181–92. https://doi.org/10.1016/j.jmbbm.2011.08.02310.1016/j.jmbbm.2011.08.023 ↗
6. Bruze M, Netterlid E, Siemund I. Aluminium – allergin of the year 2022. Dermatitis. 2022;33:10–15. https://doi.org/10.1097/DER.000000000000083610.1097/DER.0000000000000836 ↗
7. Sharkeev YP, Eroshenko AY, Kulyashova KS, Fortuna SV, Suvorov KA, Epple M, et al. Microstructure, mechanical and biological properties of zirconium alloyed with niobium after severe plastic deformation. Mater. Sci. Eng. Technol. 2013;44:198–204. https://doi.org/10.1002/mawe.20130011310.1002/mawe.201300113 ↗
8. Dreyer H, Grischke J, Tiede C, Eberhard J, Schweitzer A, Toikkanen SE, et al. Epidemiology and risk factors of peri-implantitis: a systematic review. J Periodontal Res. 2018;53:657–81. https://doi.org/10.1111/jre.1256210.1111/jre.12562 ↗
9. Christiansen C. Simulationsgestützte entwicklung eines neuartigen kieferimplantats und experimentelle bewertung der mechanischen belastbarkeit [Masterthesis]. Hannover: Leibniz Universität Hannover; 2022.
10. Soriano-Valero S, Roman-Rodriguez JL, Agustin-Panadero R, Bellot-Arcis C, Fons-Font A, Fernandez-Estevan L. Systematic review of chewing simulators: reality and reproducibility of in vitro studies. J. Clin. Exp. Dent. 2020;12:e1189–95. https://doi.org/10.4317/jced.5727910.4317/jced.57279 ↗
11. Demenko V, Linetskiy I, Nesvit K, Shevchenko A. Ultimate masticatory force as a criterion in implant selection. J Dent Res. 2011;90:1211–5. https://doi.org/10.1177/002203451141744210.1177/0022034511417442 ↗
12. Rosentritt M, Siavikis G, Behr M, Kolbeck C, Handel G. Approach for valuating the significance of laboratory simulation. J Dent. 2008;36:1048–53. https://doi.org/10.1016/j.jdent.2008.09.00110.1016/j.jdent.2008.09.001 ↗
13. Schroeder G, Rosch P, Kunzelmann KH. Influence of the preparation design on the survival probability of occlusal veneers. Dent Mater. 2022;38:646–54. https://doi.org/10.1016/j.dental.2022.02.00310.1016/j.dental.2022.02.003 ↗
14. Monje A, Roccuzzo A, Buser D, Wang HL. Influence of buccal bone wall thickness on the peri-implant hard and soft tissue dimensional changes: a systematic review. Clin Oral Implant Res. 2023;34:8–27. https://doi.org/10.1111/clr.1417710.1111/clr.14177 ↗
15. Steiner M, Mitsias ME, Ludwig K, Kern M. In vitro evaluation of a mechanical testing chewing simulator. Dent. Mater. 2009;25:494–9. https://doi.org/10.1016/j.dental.2008.09.01010.1016/j.dental.2008.09.010 ↗
16. Dittmer S, Dittmer MP, Kohorst P, Jendras M, Borchers L, Stiesch M. Effect of implant-abutment connection design on load bearing capacity and failure mode of implants. J. Prosthodont 2011;20:510–6. https://doi.org/10.1111/j.1532-849X.2011.00758.x10.1111/j.1532-849X.2011.00758.x ↗
17. Zhai Z, Nakano T, Chen Y, Watanabe S, Matsuoka T, Ishigaki S. Implant deformation and implant-abutment fracture resistance after standardized artificial aging: an in vitro study. Clin. Implant Dent Relat Res. 2023;25:107–17. https://doi.org/10.1111/cid.1315710.1111/cid.13157 ↗
18. Amelya A, Kim JE, Woo CW, Otgonbold J, Lee KW. Load-bearing capacity of posterior CAD/CAM implant-supported fixed partial dentures fabricated with different esthetic materials. Int J Prosthodont 2019;32:201–4. https://doi.org/10.11607/ijp.612810.11607/ijp.6128 ↗
19. Molinero-Mourelle P, Abou-Ayash S, Bragger U, Schimmel M, Ozcan M, Yilmaz B, et al. Load bearing capacity of 3-unit screw-retained implant-supported fixed dental prostheses with a mesial and distal cantilever on a single implant: a comparative in vitro study. J Mech Behav Biomed Mater. 2024;151:106395. https://doi.org/10.1016/j.jmbbm.2024.10639510.1016/j.jmbbm.2024.106395 ↗
20. Sarafidou K, Stiesch M, Dittmer MP, Borchers L, Kohorst P. Load-bearing capacity of implant-supported, tooth-supported, and combined zirconia-fixed dental prostheses. Implant Dent. 2011;20:311–7. https://doi.org/10.1097/ID.0b013e31822635d610.1097/ID.0b013e31822635d6 ↗
21. Hendi A, Mirzaee S, Falahchai M. The effect of different implant-abutment types and heights on screw loosening in cases with increased crown height space. Clin Exp Dent Res. 2024;10:e894. https://doi.org/10.1002/cre2.89410.1002/cre2.894 ↗