The pathology of failed McKee-Farrar implants: correlation with modern metal-on-metal-implant failure
Source abstract
Abstract
The McKee-Farrar (MF) prosthesis was the first widely used total hip replacement (THR) to employ a metal-on-metal (MoM) articulation. These implants had a high rate of early aseptic loosening but a number achieved good long-term implant survival, stimulating the reintroduction of second and third generation implants of this type. In this study we analysed archival histopathology of periprosthetic tissues in twenty cases of MF aseptic implant failure to determine if there was evidence of an innate and adaptive immune response similar to that seen in modern MoM implants. The presence of macrophages, the extent of necrosis and the ALVAL response were graded semi-quantitatively. Variable but in most cases extensive tissue necrosis was associated with a heavy macrophage response to Cobalt-Chrome (Co-Cr) wear particles in periprosthetic tissues; most cases also contained evidence of a predominantly lymphocyte response which in eight cases was moderate or heavy (Oxford Grade 2/3). Our findings show that inflammatory and necrotic changes to deposition of Co-Cr wear particles are found in periprosthetic tissues of failed MF implants, indicating that there is an innate and adaptive response similar to that noted in second/third generation MoM implants; they also suggest that the pathobiological response to metal wear particles is likely to have contributed to MF implant failure in these cases.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Orthopedic Infections and Treatments · Total Knee Arthroplasty Outcomes
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (40)
Show source-provided references
1. Amstutz HC, Grigoris P. Metal on metal bearings in hip arthroplasty. Clin Orthop Relat Res. 1996;329 Suppl:S11–34.10.1097/00003086-199608001-00003 ↗
2. Cuckler JM. The rationale for metal-on-metal total hip arthroplasty. Clin Orthop Relat Res. 2005;441:132–6.10.1097/01.blo.0000193809.85587.f8 ↗
3. Knight SR, Aujla R, Biswas SP. Total hip arthroplasty - over 100 years of operative history. Orthop Rev (Pavia). 2011;3–2:e1610.4081/or.2011.e16 ↗
4. McKee GK, Watson-Farrar J. Replacement of arthritic hips by the McKee-Farrar prosthesis. J Bone Joint Surg Br. 1966;48–2:245–59.10.1302/0301-620X.48B2.245 ↗
5. Schmalzried TP, Szuszczewicz ES, Akizuki KH, Petersen TD, Amstutz HC. Factors correlating with long term survival of McKee-Farrar total hip prostheses. Clin Orthop Relat Res. 1996;329 Suppl:S48–59.10.1097/00003086-199608001-00005 ↗
6. Walker PS, Salvati E, Hotzler RK. The wear on removed McKee-Farrar total hip prostheses. J Bone Joint Surg Am. 1974;56–1:92–100.10.2106/00004623-197456010-00011 ↗
7. Clarke MT, Darrah C, Stewart T, Ingham E, Fisher J, Nolan JF. Long-term clinical, radiological and histopathological follow-up of a well-fixed Mckee-Farrar metal-on-metal total hip arthroplasty. J Arthroplasty. 2005;20–4:542–6.10.1016/j.arth.2004.09.063 ↗
8. Jacobsson SA, Djerf K, Wahlstrom O. A comparative study between McKee-Farrar and Charnley arthroplasty with long-term follow-up periods. J Arthroplasty. 1990;5–1:9–14.10.1016/S0883-5403(06)80003-3 ↗
9. Jacobsson SA, Djerf K, Wahlstrom O. Twenty-year results of McKee-Farrar versus Charnley prosthesis. Clin Orthop Relat Res. 1996;329 Suppl:S60–8.10.1097/00003086-199608001-00006 ↗
10. Jantsch S, Schwagerl W, Zenz P, Semlitsch M, Fertschak W. Long-term results after implantation of McKee-Farrar total hip prostheses. Arch Orthop Trauma Surg. 1991;110–5:230–7.10.1007/BF00572878 ↗
11. Amstutz HC, Campbell P, McKellop H, Schmalzreid TP, Gillespie WJ, Howie D, Jacobs J, Medley J, Merritt K. Metal on metal total hip replacement workshop consensus document. Clin Orthop Relat Res. 1996;329 Suppl:S297–303.10.1097/00003086-199608001-00027 ↗
12. de Steiger RN, Hang JR, Miller LN, Graves SE, Davidson DC. Five-year results of the ASR XL acetabular system and the ASR hip resurfacing system: an analysis from the Australian orthopaedic association national joint replacement registry. J Bone Joint Surg Am. 2011;93–24:2287–93.10.2106/JBJS.J.01727 ↗
13. Langton DJ, Jameson SS, Joyce TJ, Gandhi JN, Sidaginamale R, Mereddy P, Lord J, Nargol AV. Accelerating failure rate of the ASR total hip replacement. J Bone Joint Surg Br. 2011;93–8:1011–6.10.1302/0301-620X.93B8.26040 ↗
14. Langton DJ, Jameson SS, Joyce TJ, Hallab NJ, Natu S, Nargol AV. Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Joint Surg Br. 2010;92–1:38–46.10.1302/0301-620X.92B1.22770 ↗
15. Bauer T. What does the histology tell us? In: Jones LC, Greenwold AS, Haggard WO, editors. Metal-on-metal bearing. A clinical practicum. New York, Springer; 2013. P. 153–62.10.1007/978-1-4614-8999-3_9 ↗
16. Grammatopoulos G, Pandit H, Kamali A, Maggiani F, Glyn-Jones S, Gill HS, Murray DW, Athanasou N. The correlation of wear with histological features after failed hip resurfacing arthroplasty. J Bone Joint Surg Am. 2013;95–12:e8110.2106/JBJS.L.00775 ↗
17. Mahendra G, Pandit H, Kliskey K, Murray D, Gill HS, Athanasou N. Necrotic and inflammatory changes in metal-on-metal resurfacing hip arthroplasties. Acta Orthop. 2009;80–6:653–9.10.3109/17453670903473016 ↗
18. Natu S, Sidaginamale RP, Gandhi J, Langton DJ, Nargol AV. Adverse reactions to metal debris: histopathological features of periprosthetic soft tissue reactions seen in association with failed metal on metal hip arthroplasties. J Clin Pathol. 2012;65–5:409–18.10.1136/jclinpath-2011-200398 ↗
19. Pandit H, Glyn-Jones S, McLardy-Smith P, Gundle R, Whitwell D, Gibbons CL, Ostlere S, Athanasou N, Gill HS, Murray DW. Pseudotumours associated with metal-on-metal hip resurfacings. J Bone Joint Surg Br. 2008;90–7:847–51.10.1302/0301-620X.90B7.20213 ↗
20. Campbell P, Ebramzadeh E, Nelson S, Takamura K, De Smet K, Amstutz HC. Histological features of pseudotumour-like tissues from metal-on-metal hips. Clin Orthop Relat Res. 2010;468:2321–7.10.1007/s11999-010-1372-y ↗
21. Watters TS, Cardona DM, Menon KS, Vinson EN, Bolognesi MP, Dodd LG. Aseptic lymphocyte-dominated vasculitis-associated lesion: a clinicopathologic review of an underrecognized cause of prosthetic failure. Am J Clin Pathol. 2010;134–6:886–93.10.1309/AJCPLTNEUAH8XI4W ↗
22. Brown GC, Lockshin MD, Salvati EA, Bullough PG. Sensitivity to metal as a possible cause of sterile loosening after cobalt-chromium total hip-replacement arthroplasty. J Bone Joint Surg Am. 1977;59–2:164–8.10.2106/00004623-197759020-00004 ↗
23. Doorn PF, Mirra JM, Campbell PA, Amstutz HC. Tissue reaction to metal on metal total hip prostheses. Clin Orthop Relat Res. 1996;329 Suppl:S187–205.10.1097/00003086-199608001-00017 ↗
24. Evans EM, Freeman MA, Miller AJ, Vernon-Roberts B. Metal sensitivity as a cause of bone necrosis and loosening of the prosthesis in total joint replacement. J Bone Joint Surg Br. 1974;56-B-4:626–42.10.1302/0301-620X.56B4.626 ↗
25. Howie DW. Tissue response in relation to type of wear particles around failed hip arthroplasties. J Arthroplasty. 1990;5-4:337–48.10.1016/S0883-5403(08)80093-9 ↗
26. Jones DA, Lucas HK, O’Driscoll M, Price CH, Wibberley B. Cobalt toxicity after McKee hip arthroplasty. J Bone Joint Surg Br. 1975;57–3:289–96.10.1302/0301-620X.57B3.289 ↗
27. Winter G. Tissue recation to metallic wear and corrosion products in human patients. J Biomed Materials Res. 1974;8:11–26.10.1002/jbm.820080304 ↗
28. Phillips EA, Klein GR, Cates HE, Kurtz SM, Steinbeck M. Histological characterization of periprosthetic tissue responses for metal-on-metal hip replacement. J Long Term Eff Med Implants. 2014;24:13–23.10.1615/JLongTermEffMedImplants.2014010275 ↗
29. Brown SR, Davies WA, DeHeer DH, Swanson AB. Long-term survival of McKee-Farrar total hip prostheses. Clin Orthop Relat Res. 2002;402:157–63.10.1097/00003086-200209000-00013 ↗
30. Somerville EW, Bentley GJ. McKee-Farrar total hip replacement arthroplasty. a review of the first 100 cases at the nuffield orthopaedic centre. Orthopaedics Oxford 1970;3:73–7.
31. Bentley G, Duthie RB. A comparative review of the McKee-Farrar and Charnley total hip prostheses. Clin Orthop Relat Res. 1973;95:127–42.
32. Willert HG, Buchhorn GH, Gobel D, Koster G, Schaffner S, Schenk R, Semlitsch M. Wear behavior and histopathology of classic cemented metal on metal hip endoprostheses. Clin Orthop Relat Res. 1996;329 Suppl:S160–86.10.1097/00003086-199608001-00016 ↗
33. Higuchi F, Inoue A, Semlitsch M. Metal-on-metal CoCrMo McKee-Farrar total hip arthroplasty: characteristics from a long-term follow-up study. Arch Orthop Trauma Surg. 1997;116–3:121–4.10.1007/BF00426058 ↗
34. Schmalzried TP, Peters PC, Maurer BT, Bragdon CR, Harris WH. Long-duration metal-on-metal total hip arthroplasties with low wear of the articulating surfaces. J Arthroplasty. 1996;11–3:322–31.10.1016/S0883-5403(96)80085-4 ↗
35. Zahiri CA, Schmalzried TP, Ebramzadeh E, Szuszczewicz ES, Salib D, Kim C, Amstutz HC. Lessons learned from loosening of the McKee-Farrar metal-on-metal total hip replacement. J Arthroplasty. 1999;14–3:326–32.10.1016/S0883-5403(99)90059-1 ↗
36. Davies AP, Willert HG, Campbell PA, Learmonth ID, Case CP. An unusual lymphocytic perivascular infiltration in tissues around contemporary metal-on-metal joint replacements. J Bone Joint Surg Am. 2005;87–1:18–27.10.2106/JBJS.C.00949 ↗
37. Willert HG, Buchhorn GH, Fayyazi A, Flury R, Windler M, Koster G, Lohmann CH. Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints. A clinical and histomorphological study. J Bone Joint Surg Am. 2005;87–1:28–36.10.2106/JBJS.A.02039pp ↗
38. Hallab NJ, Jacobs JJ. Biological effects of implant debris. Bull NYU Hospit Jt Dis. 2009;67:182–8.
39. Athanasou NA. The pathobiology and pathology of aseptic implant failure. Bone Joint Res. 2016;162:162–8.10.1302/2046-3758.55.BJR-2016-0086 ↗
40. Madan S, Jowett RL, Goodwin MI. Recurrent intrapelvic cyst complicating metal-on-metal cemented total hip arthroplasty. Arch Orthop Trauma Surg. 2000;120–9:508–10.10.1007/s004020000171 ↗