Microscale wear behavior and crosslinking of PEG-like coatings for total hip replacements
Source abstract
Abstract
The predominant cause of late-state failure of total hip replacements is wear-mediated osteolysis caused by wear particles that originate from the ultrahigh molecular weight polyethylene (UHMWPE) acetabular cup surface. One strategy for reducing wear particle formation from UHMWPE is to modify the surface with a hydrophilic coating to increase lubrication from synovial fluid. This study focuses on the wear behavior of hydrophilic coatings similar to poly(ethylene glycol) (PEG). The coatings were produced by plasma-polymerizing tetraglyme on UHMWPE in a chamber heated to 40 degrees C or 50 degrees C. Both temperatures yielded coatings with PEG-like chemistry and increased hydrophilicity relative to uncoated UHMWPE; however, the 40 degrees C coatings were significantly more resistant to damage induced by atomic force microscopy nanoscratching. The 40 degrees C coatings exhibited only one damage mode (delamination) and often showed no signs of damage after repeated scratching. In contrast, the 50 degrees C coatings exhibited three damage modes (roughening, thinning, and delamination), and always showed visible signs of damage after no more than two scratches. The greater wear resistance of the 40 degrees C coatings could not be explained by coating chemistry or hydrophilicity, but it corresponded to an approximately 26-32% greater degree of crosslinking relative to the 50 degrees C surfaces, suggesting that crosslinking should be a significant design consideration for hydrophilic coatings used for total hip replacements and other wear-dependent applications.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Advanced materials and composites · Total Knee Arthroplasty Outcomes
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (20)
Show source-provided references
1. Wang ML, Sharkey PF, Tuan RS. Particle bioreactivity and wear-mediated osteolysis. J Arthroplasty. 2004;19:1028–38.10.1016/j.arth.2004.03.024 ↗
2. Ingham E, Fisher J. Biological reactions to wear debris in total joint replacement. Proc Inst Mech Eng H. 2000;214:21–37.10.1243/0954411001535219 ↗
3. Archibeck MJ, Jacobs JJ, Roebuck KA, Glant TT. The basic science of periprosthetic osteolysis. J Bone Joint Surg Am. 2000;82A:1478–89.10.2106/00004623-200010000-00014 ↗
4. Kurtz SM, Muratoglu OK, Evans M, Edidin AA. Advances in the processing, sterilization, and crosslinking of ultra-high molecular weight polyethylene for total joint arthroplasty. Biomaterials. 1999;20:1659–88.10.1016/S0142-9612(99)00053-8 ↗
5. Muratoglu OK, Bragdon CR, O’Connor DO, et al. Unified wear model for highly crosslinked ultra-high molecular weight polyethylenes. Biomaterials. 1999;20:1463–70.10.1016/S0142-9612(99)00039-3 ↗
6. Baker DA, Bellare A, Pruitt L. The effects of degree of crosslinking on the fatigue crack initiation and propagation resistance of orthopedic-grade polyethylene. J Biomed Mater Res A. 2003;66A:146–54.10.1002/jbm.a.10606 ↗
7. Collier JP, Currier BH, Kennedy FE, et al. Comparison of cross-linked polyethylene materials for orthopedic applications. Clin Orthop Relat Res. 2003;414:289–304.10.1097/01.blo.0000073343.50837.03 ↗
8. Moro T, Takatori Y, Ishihara K, et al. Surface grafting of artificial joints with a biocompatible polymer for preventing periprosthetic osteolysis. Nat Mater. 2004;3:829–36.10.1038/nmat1233 ↗
9. Pavoor PV, Gearing BP, Muratoglu O, Cohen RE, Bellare A. Wear reduction of orthopaedic bearing surfaces using polyelectrolyte multilayer nanocoatings. Biomaterials. 2006;27:1527–33.10.1016/j.biomaterials.2005.08.022 ↗
10. Zhang M, Pare P, King R, James SP. A novel ultra high molecular weight polyethylene-hyaluronan microcomposite for use in total joint replacements. II. Mechanical and tribological property evaluation. J Biomed Mater Res A. 2007;82A:18–26.10.1002/jbm.a.31141 ↗
11. Kane SR, Ashby PD, Pruitt LA. Characterization and tribology of PEG-like coatings on UHMWPE for total hip replacements. J Biomed Mater Res A. 2009. doi: 10.1002/jbm.a.32484 .10.1002/jbm.a.32484 ↗
12. Kane SR, Ashby PD, Pruitt LA. ATR-FTIR as a thickness measurement technique for hydrated polymer-on-polymer coatings. J Biomed Mater Res B. 2009;91(2):613–20.10.1002/jbm.b.31436 ↗
13. Johnston EE, Bryers JD, Ratner BD. Plasma deposition and surface characterization of oligoglyme, dioxane, and crown ether nonfouling films. Langmuir. 2005;21:870–81.10.1021/la036274s ↗
14. Zhou J, Chakravartula A, Pruitt L, Komvopoulos K. Tribological and nanomechanical properties of unmodifiedand crosslinked ultra-high molecular weight polyethylene for total joint replacements. J Tribol-Trans ASME. 2004;126:386–94.10.1115/1.1611511 ↗
15. McKellop HA, Campbell P, Park SH, et al. The origin of submicron polyethylene wear debris in total hip arthroplasty. Clin Orthop Relat Res. 1995;311:3–20.
16. Wang A, Stark C, Dumbleton JH (1996) Mechanistic and morphological origins of ultra-high molecular weight polyethylene wear debris in total joint replacement prostheses [comment]. Proc Inst Mech Eng H. 210:141–55.10.1243/PIME_PROC_1996_210_407_02 ↗
17. Freeman ME, Furey MJ, Love BJ, Hampton JM. Friction, wear, and lubrication of hydrogels as synthetic articular cartilage. Wear. 2000;241:129–35.10.1016/S0043-1648(00)00387-2 ↗
18. Bavaresco VP, Zavaglia CAC, Reis MC, Gomes JR. Study on the tribological properties of pHEMA hydrogels for use in artificial articular cartilage. Wear. 2008;265:269–77.10.1016/j.wear.2007.10.009 ↗
19. Flory PJ. Principles of polymer chemistry. London: Cornell University Press; 1953.
20. Flory PJ, Rehner J. Statistical mechanics of cross-linked polymer networks. II. Swelling. J. Chem. Phys. 1943;11:521–6.10.1063/1.1723792 ↗