Study of mechanical degradation of UHMWPE acetabular components due to clinical X-ray procedures
Source abstract
Abstract
The use of multi-component femoral implants to replace the femur head and re-establish bone motion has been widespread since the 70s. Frequently these implants have spherical metallic heads made of, for example, 316-L stainless steel or Cr-Co alloys, which allow rotational motion towards a polymeric component (UHMWPE). One of the major causes of implant rejection is the generation of UHMWPE debris on the surface between the implant head and the polymeric component. The gamma ray sterilization of implants and the periodical X-ray medical control could contribute to premature degradation of the polymeric surface, resulting in increased wear and shortened lifetime of the implant. In this work we study the degradation degree of the polymeric UHMWPE component as function of the X-ray dose. The elasto-plastic deformation and recovery were carried out by means of a nanohardness tester equipment and the polymer degradation was measured using a fast Fourier transform infra-red (FT-IR) equipment. The results show the compromise among the irradiation doses, the surface oxidation and the mechanical properties of the samples.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Advanced materials and composites · Titanium Alloys Microstructure and Properties
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (20)
Show source-provided references
1. R. L. CLOUGH, Nucl. Instrum. Methods Phys. Res. B: Beam Inter. Mater. Atoms 185 (2001) 810.1016/S0168-583X(01)00966-1 ↗
2. J. P. COLLIER, D. K. SPERLING, J. H. CURRIER, L. C. SUTULA, K. A. SAUM and M. B MAYOR, J. Arthroplasty, 11(4) (1996) 37710.1016/S0883-5403(96)80026-X ↗
3. M. JASTY and W. H. HARRIS, Adv. Oper. Orthop. 1 (1993) 1
4. H. MCKELLOP, F. W. SHEN, B. LU, P. CAMPBELL and R. SALOVEY, J. Orthop. Res. 17 (1999) 15710.1002/jor.1100170203 ↗
5. O. K. MURATOGLU, C. R. BRAGDON, D. O. O’CONNOR, M. JASTY and W. H. HARRIS, Proc. 45th Ann. Meeting Orthop. Res. Soc. 817 (1999)
6. L. C. SUTULA, J. P. COLLIER, K. A. SAUM, B. H. CURRIER, J. H. CURRIER, W. M. SANFORD, M. B. MAYOR, R. E. WOODING, D. K. SPERLING, I. R. WILLIAMS and D. J. KASPRZAK, Clinical Orthop. Related Res. 319 (1995) 28
7. K. S. TOOHEY, T. A. BLANCHET and D. D. HECKELMAN, Wear 255 (2003) 107610.1016/S0043-1648(03)00270-9 ↗
8. K. WUNSCH and H. J. DALCOLMO, Inter. J. Radiat. Appl. Instrumen.. Part C. Radiat. Phys. Chem. 39(5) (1992) 443
9. S. M. KURTZ, O. K. MURATOGLU, R. F. GSELL, W. SHEN, C. F. COOPER, J. BUCHANAN, S. SPIEGELBERG, S. S. YAU and A. A. EDIDIN, Proc. 45th Ann. Meeting Orthop. Res. Soc., 98 (1999)
10. A. WANG, A. ESSNER, V. K. POLINENI, C. STARK and H. H. DUMBLETON, Tribol. Inter. 31 (1998) 1710.1016/S0301-679X(98)00005-X ↗
11. L. COSTA, M. P. LUDA, L. TROSSARELLI, E. M. B. PREVER, M. CROVA and P. GALLINARO, Biomaterials 19 (1998)10.1016/S0142-9612(98)00013-1 ↗
12. H. E. JOHNS, E. R. EPP, D. V. CORMACK and S. O. FEDORUK, J.Radiol. 25 (1952) 30210.1259/0007-1285-25-294-302 ↗
13. T. A. BLANCHET and B. R. BURROUGHS, J. Biomed. Mater. Res.: Appl. Biomater. 58(6) (2001) 68410.1002/jbm.1070 ↗
14. D. C. SUN, G. SCHMIDIG, C. STARK and J. H. DUMBLETON, Proc. 21st Ann. Meeting Trans. Soc. Biomater. 362 (1995)
15. V. PREMNATH, W. H. HARRIS, M. JASTY and E. W. MERRILL, Biomaterials 17 (1996) 174110.1016/0142-9612(95)00349-5 ↗
16. M. C. BUNCICK, D. E. THOMAS, K. S. MCKINNY and M. S. JAHAN, Appl. Surf. Sci. 156 (2000) 9710.1016/S0169-4332(99)00367-0 ↗
17. L. A. VASCONCELLOS, E. BLANDO and R. HÜBLER, Proc. III Congresso Latino Americano de Órgãos artificiais e Biomateriais (III COLAOB), CD, BR139, (2004)
18. A. W. LEE, J. P. SANTERRE and E. BOYNTON, Biomaterials 21 (2000) 85110.1016/S0142-9612(99)00257-4 ↗
19. B. R. BURROUGHS and T. A. BLANCHET, Tribol. Trans. 44(2) (2001) 21510.1080/10402000108982451 ↗
20. ASTM D 2765, Annual Book of ASTM Standards 8 (2001) 114