Interface micromechanics of transverse sections from retrieved cemented hip reconstructions: an experimental and finite element comparison

Daan Waanders, Dennis Janssen, Sanaz Berahmani, Mark A. Miller, Kenneth A. Mann, Nico Verdonschot

Abstract

In finite element analysis (FEA) models of cemented hip reconstructions, it is crucial to include the cement-bone interface mechanics. Recently, a micromechanical cohesive model was generated which reproduces the behavior of the cement-bone interface. The goal was to investigate whether this cohesive model was directly applicable on a macro level. From transverse sections of retrieved cemented hip reconstructions, two FEA-models were generated. The cement-bone interface was modeled with cohesive elements. A torque was applied and the cement-bone interface micromotions, global stiffness and stem translation were monitored. A sensitivity analysis was performed to investigate whether the cohesive model could be improved. All results were compared with experimental findings. That the original cohesive model resulted in a too compliant macromechanical response; the motions were too large and the global stiffness too small. When the cohesive model was modified, the match with the experimental response improved considerably.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Composite Material Mechanics · Elasticity and Material Modeling

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (38)

Show source-provided references
1. Hailer NP, Garellick G, Karrholm J. Uncemented and cemented primary total hip arthroplasty in the Swedish Hip Arthroplasty Register. Acta Orthop. 2010;81:34–41.10.3109/17453671003685400 ↗
2. Goto K, Kawanabe K, Kowalski R, Baker D, Nakamura T. Bonding ability evaluation of bone cement on the cortical surface of rabbit’s tibia. J Mater Sci Mater Med. 2010;21:139–46.10.1007/s10856-009-3861-7 ↗
3. Oonishi H, Ohashi H, Oonishi H Jr, Kim SC. THA with hydroxyapatite granules at cement–bone interface: 15- to 20-year results. Clin Orthop Relat Res. 2008;466:373–9.10.1007/s11999-007-0057-7 ↗
4. Lucksanasombool P, Higgs WA, Ignat M, Higgs RJ, Swain MV. Comparison of failure characteristics of a range of cancellous bone–bone cement composites. J Biomed Mater Res A. 2003;64:93–104.10.1002/jbm.a.10361 ↗
5. Waanders D, Janssen D, Mann KA, Verdonschot N. The mechanical effects of different levels of cement penetration at the cement–bone interface. J Biomech. 2010;43:1167–75.10.1016/j.jbiomech.2009.11.033 ↗
6. Mann KA, Miller MA, Verdonschot N, Izant TH, Race A. Functional interface micromechanics of 11 en-bloc retrieved cemented femoral hip replacements. Acta Orthop. 2010;81:308–17.10.3109/17453674.2010.480938 ↗
7. Bishop NE, Schoenwald M, Schultz P, Puschel K, Morlock MM. The condition of the cement mantle in femoral hip prosthesis implantations–a post mortem retrieval study. Hip Int. 2009;19:87–95.10.1177/112070000901900202 ↗
8. Venesmaa PK, Kroger HP, Jurvelin JS, Miettinen HJ, Suomalainen OT, Alhava EM. Periprosthetic bone loss after cemented total hip arthroplasty: a prospective 5-year dual energy radiographic absorptiometry study of 15 patients. Acta Orthop Scand. 2003;74:31–6.10.1080/00016470310013617 ↗
9. Miller MA, Eberhardt AW, Cleary RJ, Verdonschot N, Mann KA. Micromechanics of postmortem-retrieved cement–bone interfaces. J Orthop Res. 2010;28:170–7.10.1002/jor.20893 ↗
10. Stolk J, Janssen D, Huiskes R, Verdonschot N. Finite element-based preclinical testing of cemented total hip implants. Clin Orthop Relat Res. 2007;456:138–47.10.1097/BLO.0b013e31802ba491 ↗
11. Hung JP, Chen JH, Chiang HL, Wu JS. Computer simulation on fatigue behavior of cemented hip prostheses: a physiological model. Comput Methods Programs Biomed. 2004;76:103–13.10.1016/j.cmpb.2004.05.004 ↗
12. Katoozian H, Davy DT. Effects of loading conditions and objective function on three-dimensional shape optimization of femoral components of hip endoprostheses. Med Eng Phys. 2000;22:243–51.10.1016/S1350-4533(00)00030-8 ↗
13. Afsharpoya B, Barton DC, Fisher J, Purbach B, Wroblewski M, Stewart TD. Cement mantle stress under retroversion torque at heel-strike. Med Eng Phys. 2009;31:1323–30.10.1016/j.medengphy.2009.09.006 ↗
14. Lewis G, Duggineni R. Finite element analysis of a three-dimensional model of a proximal femur-cemented femoral THJR component construct: influence of assigned interface conditions on strain energy density. Bio-Med Mater Eng. 2006;16:319–27.
15. Colombi P. Fatigue analysis of cemented hip prosthesis: model definition and damage evolution algorithms. Int J Fatigue. 2002;24:895–901.10.1016/S0142-1123(01)00203-1 ↗
16. Verdonschot N, Huiskes R. The effects of cement-stem debonding in THA on the long-term failure probability of cement. J Biomech. 1997;30:795–802.10.1016/S0021-9290(97)00038-9 ↗
17. Mann KA, Miller MA, Cleary RJ, Janssen D, Verdonschot N. Experimental micromechanics of the cement–bone interface. J Orthop Res. 2008;26:872–9.10.1002/jor.20575 ↗
18. Waanders D, Janssen D, Mann KA, Verdonschot N. The behavior of the micro-mechanical cement–bone interface affects the cement failure in total hip replacement. J Biomech. 2011;44:228–34.10.1016/j.jbiomech.2010.10.020 ↗
19. Perez MA, Palacios J. Comparative finite element analysis of the debonding process in different concepts of cemented hip implants. Ann Biomed Eng. 2010;38:2093–106.10.1007/s10439-010-9996-3 ↗
20. Moreo P, Perez MA, Garcia-Amar JM, Doblare M. Modelling the mixed-mode failure of cement–bone interfaces. Eng Fract Mech. 2006;73:1379–95.10.1016/j.engfracmech.2006.01.005 ↗
21. Mann KA, Damron LA. Predicting the failure response of cement–bone constructs using a non-linear fracture mechanics approach. J Biomech Eng. 2002;124:462–70.10.1115/1.1488167 ↗
22. Alfano G, Crisfield MA. Finite element interface models for the delamination analysis of laminated composites: mechanical and computational issues. Int J Numer Methods Eng. 2001;50:1701–36.10.1002/nme.93 ↗
23. Wang JY, Tozzi G, Chen J, Contal F, Lupton C, Tong J. Bone–cement interfacial behaviour under mixed mode loading conditions. J Mech Behav Biomed Mater. 2010;3:392–8.10.1016/j.jmbbm.2010.03.001 ↗
24. Mann KA, Mocarski R, Damron LA, Allen MJ, Ayers DC. Mixed-mode failure response of the cement–bone interface. J Orthop Res. 2001;19:1153–61.10.1016/S0736-0266(01)00036-5 ↗
25. Waanders D, Janssen D, Bertoldi K, Mann KA, Verdonschot N. Mixed-mode loading of the cement–bone interface: a finite element study. Comput Methods Biomech Biomed Eng. 2011;14:145–55.10.1080/10255842.2010.535814 ↗
26. Waanders D, Janssen D, Mann KA, Verdonschot N. Morphology based cohesive zone modeling of the cement–bone interface from postmortem retrievals. J Mech Behav Biomed Mater. 2011;4:1492–503.10.1016/j.jmbbm.2011.05.019 ↗
27. Bergmann G, Graichen F, Rohlmann A. Is staircase walking a risk for the fixation of hip implants? J Biomech. 1995;28:535–53.10.1016/0021-9290(94)00105-D ↗
28. Douglas D, Peucker T. Algorithms for the reduction of the number of points required to represent a digitized line or its caricature. Can Cartogr. 1973;10:112–22.10.3138/FM57-6770-U75U-7727 ↗
29. Janssen D, Mann KA, Verdonschot N. Micro-mechanical modeling of the cement–bone interface: the effect of friction, morphology and material properties on the micromechanical response. J Biomech. 2008;41:3158–63.10.1016/j.jbiomech.2008.08.020 ↗
30. Harper EJ, Bonfield W. Tensile characteristics of ten commercial acrylic bone cements. J Biomed Mater Res. 2000;53:605–16.10.1002/1097-4636(200009)53:5<605::AID-JBM22>3.0.CO;2-5 ↗
31. Lewis G. Properties of acrylic bone cement: state of the art review. J Biomed Mater Res. 1997;38:155–82.10.1002/(SICI)1097-4636(199722)38:2<155::AID-JBM10>3.0.CO;2-C ↗
32. Zar JH. Biostatistical analysis. 5th ed. Upper Saddle River: Pearson Prentice Hall; 2010.
33. Mann KA, Gupta S, Race A, Miller MA, Cleary RJ. Application of circular statistics in the study of crack distribution around cemented femoral components. J Biomech. 2003;36:1231–4.10.1016/S0021-9290(03)00091-5 ↗
34. Ural A. Prediction of Colles’ fracture load in human radius using cohesive finite element modeling. J Biomech. 2009;42:22–8.10.1016/j.jbiomech.2008.10.011 ↗
35. Salomonsson K, Andersson T. Modeling and parameter calibration of an adhesive layer at the meso level. Mech Mater. 2008;40:48–65.10.1016/j.mechmat.2007.06.004 ↗
36. Zelle J, Janssen D, Peeters S, Brouwer C, Verdonschot N. Mixed-mode failure strength of implant–cement interface specimens with varying surface roughness. J Biomech. 2011;44:780–3.10.1016/j.jbiomech.2010.10.037 ↗
37. Miller MA, Race A, Waanders D, Cleary RJ, Janssen D, Verdonschot N, Mann KA. Multi-axial loading micromechanics of the cement–bone interface in post-mortem retrievals and lab-prepared specimens. J Mech Behav Biomed Mater. 2011;4:366–74.10.1016/j.jmbbm.2010.11.004 ↗
38. Sundfeldt M, Carlsson LV, Johansson CB, Thomsen P, Gretzer C. Aseptic loosening, not only a question of wear: a review of different theories. Acta Orthop. 2006;77:177–97.10.1080/17453670610045902 ↗