Effect of iodixanol particle size on the mechanical properties of a PMMA based bone cement
Source abstract
Abstract
Iodixanol (IDX) is a water soluble opacifier widely used in radiographical examinations of blood vessels and neural tissue, and it has been suggested as a potential contrast media in acrylic bone cement. The effect of the iodixanol particle size on the polymerisation process of the bone cement, the molecular weight, and the quasi-static mechanical properties have been investigated in this article. The investigation was performed using radiolucent Palacos powder mixed with 8 wt% of iodixanol with particle sizes ranging from 3 to 20 microm MMD, compared with commercial Palacos R (15 wt% ZrO2) as control. Tensile, compressive and flexural tests showed that smaller particles (groups with 3, 4, and 5 microm particles) resulted in significantly lower mechanical properties than the larger particles (groups with 15, 16, and 20 microm particles). There was no difference in molecular weight between the groups. The thermographical investigation showed that the IDX cements exhibit substantially lower maximum temperatures than Palacos R, with the 4 microm IDX group having the lowest maximum temperature. The isothermal and the constant rate differential scanning calorimetry (DSC) did not show any difference in polymerisation heat (DeltaH) or glass transition temperature (Tg) between radiolucent cement, or cement containing either IDX, or ZrO2. The findings show that the particle size for a bone cement containing iodixanol should be above 8 microm MMD.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Bone Tissue Engineering Materials · Dental materials and restorations · Orthopaedic implants and arthroplasty
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (27)
Show source-provided references
1. L. CARAVIA, D. DOWSON, J. FISHER and B. JOBBINS, Proc. Inst. Mech. Eng [H.] 204 (1990) 6510.1243/PIME_PROC_1990_204_230_02 ↗
2. J. R. COOPER, D. DOWSON, J. FISHER and B. JOBBINS, J. Med. Eng Technol. 15 (1991) 6310.3109/03091909109009970 ↗
3. G. H. ISAAC, B. M. WROBLEWSKI, J. R. ATKINSON and D. DOWSON, Clin. Orthop. Relat. R. 276 (1992) 11510.1097/00003086-199203000-00015 ↗
4. P. ASPENBERG and H. VAN DER VIS, Clin. Orthop. Relat. R. 352 (1998) 75
5. J. A. WIMHURST, R. A. BROOKS and N. RUSHTON, J. Bone. Joint. Surg. Br. 83B (2001) 27810.1302/0301-620X.83B2.10392 ↗
6. J. A. WIMHURST, R. A. BROOKS, N. RUSHTON, J. Bone. Joint. Surg. Br. 83B (2001) 58810.1302/0301-620X.83B4.10513 ↗
7. F. KJELLSON, T. ALMEN, K. E. TANNER, I. D. MCCARTHY and L. LIDGREN, J. Biomed. Mater. Res. 70B (2004) 35410.1002/jbm.b.30060 ↗
8. T. ALMÉN, Acta Radiol. Suppl. 399 (1995) 210.1177/0284185195036S39902 ↗
9. J. S. WANG, J. DIAZ, A. SABOKBAR, N. A. ATHANASOU, F. KJELLSON, K. E. TANNER, I. D. MCCARTHY and L. LIDGREN, J Roy Soc Interface 2 (2005) 7110.1098/rsif.2004.0009 ↗
10. F. KJELLSON, B. BRUDELI, I. D. MCCARTHY, L. LIDGREN, J. Biomed. Mater. Res. 71A (2004) 29210.1002/jbm.a.30162 ↗
11. F. KJELLSON, J. S. WANG, T. ALMEN, A. MATTSSON, J. KLAVENESS, K. E. TANNER and L. LIDGREN, J. Mater. Sci Mater. Med. 12 (2001) 88910.1023/A:1012867824140 ↗
12. A. BORZACCHIELLO, L. AMBROSIO, L. NICOLAIS, E. J. HARPER, K. E. TANNER and W. BONFIELD, J. Mater. Sci Mater. Med. 9 (1998) 83510.1023/A:1008996112042 ↗
13. A. BORZACCHIELLO, L. AMBROSIO, L. NICOLAIS, E. J. HARPER, K. E. TANNER and W. BONFIELD, J. Mater. Sci Mater. Med. 9 (1998) 31710.1023/A:1008898712929 ↗
14. S. N. ABDULGHANI, S. N. NAZHAT, J. C. BEHIRI and S. DEB, J. Biomater. Sci Polym. Ed. 14 (2003) 122910.1163/156856203322553455 ↗
15. K. FRIEDRICH and U. A. KARSCH, J. Mater. Sci. 16 (1981) 216710.1007/BF00542377 ↗
16. C. LIU, S. M. GREEN, N. D. WATKINS, P. J. GREGG and A. W. MCCASKIE, Proc. Inst. Mech. Eng [H.] 215 (2001) 35910.1243/0954411011535948 ↗
17. S. DEB, S. ABDULGHANI and J. C. BEHIRI, Biomaterials 23 (2002) 338710.1016/S0142-9612(02)00039-X ↗
18. S. ABDULGHANI, Thesis title:“An investigation into the mechanical and biological properties of acrylic bone cement containing triphenyl bismuth (TPB) as an alternative radiopacifier” Department of Materials, Queen Mary, University of London (2003)
19. M. P. GINEBRA, L. ALBUIXECH, E. FERNANDEZ-BARRAGAN, C. APARICIO, F. J. GIL, R. J. SAN, B. VAZQUEZ and J. A. PLANELL, Biomaterials 23 (2002) 187310.1016/S0142-9612(01)00314-3 ↗
20. J. GRAHAM, L. PRUITT, M. RIES and N. GUNDIAH, J. Arthroplasty 15 (2000) 102810.1054/arth.2000.8188 ↗
21. K. F. HUGHES, M. D. RIES and L. A. PRUITT, J. Biomed. Mater. Res. A 65 (2003) 12610.1002/jbm.a.10373 ↗
22. N. J. DUNNE and J. F. ORR, J. Mater. Sci Mater. Med. 13 (2002) 1710.1023/A:1013670132001 ↗
23. S. TOKSVIG-LARSEN, H. FRANZEN and L. RYD, Acta Orthop. Scand. 62 (1991) 10210.3109/17453679108999232 ↗
24. K.-D. KÜHN, in “Bone Cements” (Springer-Verlag, Berlin Heidelberg, 2000)10.1007/978-3-642-59762-6 ↗
25. G. BAROUD, M. SAMARA, T. STEFFEN, J. Biomed. Mater. Res. 68B (2004) 11210.1002/jbm.b.20009 ↗
26. G. G. ODIEN, in “Principles of Polymerization” (John Wiley & Sons, Inc, New York, 1981)
27. J. M. YANG, Biomaterials 18 (1997) 129310.1016/S0142-9612(97)00060-4 ↗