Mechanical failure of metal–polyethylene sandwich liner in metal-on-metal total hip replacement
Source abstract
Abstract
Metal-on-metal had been proposed as an optimal articulation in THRs, however, many monoblock prostheses have been recalled in the USA because of significant high rates of early failure. Metal-on-metal prostheses had been implanted in our institution, and this is a case history of a single patient, in whom metal-on-metal THRs with different femoral sizes of heads were implanted. A 57-year-old female patient underwent bilateral total hip replacements with metal-on-metal prostheses using metal-polyethylene "sandwich" liners 9 years ago on the right side and 7 years ago on the left side respectively. The only difference in both sides was the femoral head diameter of 28 mm in right and 34 mm in left. Seven years after the left surgery, the acetabular liner was dissociated, however, metallosis was not detected. Although the larger femoral head was thought to increase hip joint stability, it dictated a reduction in polyethylene thickness in this prosthesis design, and it was 4 mm in the left hip. Recently, metal-on-metal articulations are thought not to be optimal for hip joint bearing surface, however, this clinical failure was due to the polyethylene thickness and quality.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Advanced materials and composites
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (18)
Show source-provided references
1. Berry DJ, von Knoch M, Schleck CD, Harmsen WS. Effect of femoral head diameter and operative approach on risk of dislocation after primary total hip arthroplasty. JBJS Am. 2005;87:2456–63.10.2106/JBJS.D.02860 ↗
2. Kabo JM, Gebhard JS, Loren G, Amstutz HC. In vivo wear of polyethylene acetabular components. JBJS Br. 1993;75:254–8.
3. Pelt CE, Erickson J, Clarke I, Donaldson T, Layfield L, Peters CL. Histologic, serologic, and tribologic findings in failed metal-on-metal total hip arthroplasty. JBJS Am. 2013;95(21):e163.10.2106/JBJS.L.01446 ↗
4. Haddad FS, Thakrar RR, Hart AJ, Skinner JA, Nargol AV, Nolan JF, Gill HS, Murray DW, Blom AW, Case CP. Metal-on-metal bearings: the evidence so far. JBJS Br. 2011;93:572–9.10.2106/JBJS.J.00269 ↗
5. Cuckler JM. The rationale for metal-on-metal total hip arthroplasty. CORR. 2005;441:132–6.10.1097/01.blo.0000193809.85587.f8 ↗
6. Hug KT, Watters TS, Vail TP, Bolognesi MP. The withdrawn ASRTM THA and hip resurfacing systems how have our patients fared over 1 to 6 years? CORR. 2013;471:430–8.10.1007/s11999-012-2547-5 ↗
7. Ng VY, Arnott L, McShane MA. Perspectives in managing an implant recall: revision of 94 Durom Metasul acetabular components. JBJS Am. 2011;93(17):e100(1-5).
8. Müller ME. The benefits of metal-on-metal total hip replacements. CORR. 1995;311:54–9.
9. Milošev I, Trebse R, Kovac S, Cör A, Campbell P. Dissociation of the metal inlay from the polyethylene liner in an uncemented threaded cup. Arch Orthop Trauma Surg. 2005;125:134–41.10.1007/s00402-004-0781-3 ↗
10. Melvin JS, Karthikeyan T, Cope R, Fehring TK. Early failures in total hip arthroplasty—a changing paradigm. J Arthroplast. 2013; (in press).10.1016/j.arth.2013.12.024 ↗
11. Jameson SS, Lees D, James P, Serrano-Pedraza I, Partington PF, Muller SD, Meek RM, Reed MR. Lower rates of dislocation with increased femoral head size after primary total hip replacement: a five-year analysis of NHS patients in England. JBJS Br. 2011;93:876–80.
12. Cuckler JM, Moore KD, Lombardi AV Jr, McPherson E, Emerson R. Large versus small femoral heads in metal-on-metal total hip arthroplasty. J Arthroplast. 2004;19(8 Suppl 3):41–4.10.1016/j.arth.2004.09.006 ↗
13. Amlie E, Hovik O, Reikeras O. Dislocation after total hip arthroplasty with 28 and 32-mm femoral head. J Orthop Traumatol. 2010;11:111–5.10.1007/s10195-010-0097-8 ↗
14. Clarke MT, Lee PT, Villar RN. Dislocation after total hip replacement in relation to metal-on-metal bearing surfaces. JBJS Br. 2003;85:650–4.
15. Ohnsorge JAK, Davis J, Maus U, Saklak M, Weisskopf M, Wirtz DC. Early polyethylene wear and excessive acetabular granuloma in an uncemented HA-coated total hip arthroplasty—midterm results of a prospective study. HSSJ. 2006;2:114–20.10.1007/s11420-006-9001-8 ↗
16. Leali A, Fetto JF. Preservation of femoral bone mass after total hip replacements with a lateral flare stem. Int Orthop. 2004;28:151–4.10.1007/s00264-004-0554-1 ↗
17. Girarda J, Soenena M, Monninb C, Migauda H. Bilateral simultaneous metal inlay dissociation from the polyethylene liner of a metal-on-metal hip replacement. Orthop Traumatol. 2009;95:443–6.
18. Digas G, Kärrholm J, Thanner J, Herberts P. 5-year experience of highly cross-linked polyethylene in cemented and uncemented sockets: two randomized studies using radiostereometric analysis. Acta Orthop. 2007;78:746–54.10.1080/17453670710014518 ↗