Mechanical failure of a Thompson’s hemiarthoplasty stem 28 years post-implantation: an investigation with electron microscopy
Source abstract
Abstract
Tauhe authors would like to report a mechanical failure of a Thompson's prosthesis, 28 years post-implantation. A detailed examination of the specimen revealed no defects in the prosthesis and a dominating 'brittle component' fracture of the stem. In this context the detailed fractographic study by scanning electron microscopy (SEM) revealed no detrimental manufactural defects that may have produced microcracks and consequently risked initiating the fracture propagation. In contrast, the fracture was mainly a fatigue one with a mixed mode of microscopic trans- and intergranular crack propagation. To the best of our knowledge, such a mechanism of implant failure in a cementless stem has never before reached 28 years neither in a Thompson's nor any other type of prosthesis, and in the already reported case, it exceeded 30 years [N. Wolson and J. P. Waadell, Can. J. Surg. 38(6) (1995) 542], however the stem's ultrastructure has never been investigated under electron microscopy, which arguably can provide a useful assessment of a fatigue fracture. Tauhe authors introduce the question of revising our standards when evaluating the newly designed and expensive implants and propose re-focusing on surgical technique, rather than purely on implant properties.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Orthopedic Infections and Treatments
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (8)
Show source-provided references
1. T. A. GRUEN, G. M. MCNEICE and H. C. AMSTUTZ, Clin. Orthop. 141 (1979) 17
2. N. WOLSON and J. P. WAADELL, Can. J. Surg. 38(6) (1995) 542
3. B. M. WROBLEWSKI, Acta Orthop. Scand. 53(2) (1982) 27910.3109/17453678208992216 ↗
4. D. C. KWOK and R. L. CRUESS, Clin. Orthop. 169 (1982) 179
5. L. MARMOR and T. A. GRUEN, Clin. Orthop. 190 (1984) 14810.1097/00003086-198411000-00023 ↗
6. L. N. GILBERTSON and R. D. ZIPP (eds) “Fractography and material science” (1981) ASTM-STP 73310.1520/STP733-EB ↗
7. L. N GILBERTSON and R. D. ZIPP, “Fractography in failure analysis” (American Society for Testing and Materials, 1978) ASTM-STP 645
8. J. E. MASTERS and A. U. JOSEPH (eds) “Fractography in modern engineering material: composites and metals” (1987) ASTM-STP 94810.1520/STP948-EB ↗