Quantifying physical degradation alongside recording and stimulation performance of 980 intracortical microelectrodes chronically implanted in three humans for 956-2130 days
Source abstract
Abstract
The clinical success of brain computer interfaces (BCI) depends on overcoming both biological and material challenges to ensure a long-term stable connection for neural recording and stimulation. This study systematically quantified damage that microelectrodes sustained during chronical implantation in three people with tetraplegia for 956-2130 days. Using scanning electron microscopy (SEM), we imaged 980 microelectrodes from eleven Neuroport arrays tipped with platinum (Pt, n = 8) and sputtered iridium oxide film (SIROF, n = 3). Arrays were implanted/explanted from posterior parietal, motor and somatosensory cortices across three clinical sites (Caltech/UCLA, Caltech/USC, APL/Johns Hopkins). From the electron micrographs, we quantified and correlated physical damage with functional outcomes measured in vivo, prior to explant (recording quality, noise, impedance and stimulation ability). Despite greater physical degradation, SIROF electrodes were twice as likely to record neural activity than Pt (measured by SNR). For SIROF, 1 kHz impedance significantly correlated with all physical damage metrics, recording metrics, and stimulation performance, suggesting a reliable measurement of in vivo degradation. We observed a new degradation type, primarily on stimulated electrodes ("pockmarked" vs "cracked") electrodes; however, no significant degradation due to stimulation or amount of charge delivered. We hypothesize erosion of the silicon shank accelerates damage to the electrode / tissue interface, following damage to the tip metal. These findings link quantitative measurements to the microelectrodes' physical condition and their capacity to record/stimulate. These data could lead to improved manufacturing processes or novel electrode designs to improve long-term performance of BCIs, making them vitally important as multi-year clinical trials of BCIs are becoming more common. STATEMENT OF SIGNIFICANCE: Long-term performance stability of the electrode-tissue interface is essential for clinical viability of brain computer interface (BCI) devices; currently, materials degradation is a critical component for performance loss. Across three human participants, ten micro-electrode arrays (plus one control) were implanted for 956-2130 days. Using scanning electron microscopy (SEM), we analyzed degradation of 980 electrodes, comparing two types of commonly implanted electrode tip metals: Platinum (Pt) and Sputtered Iridium Oxide Film (SIROF). We correlated observed degradation with in vivo electrode performance: recording (signal-to-noise ratio, noise, impedance) and stimulation (evoked somatosensory percepts). We hypothesize penetration of the electrode tip by biotic processes leads to erosion of the supporting silicon core, which then accelerates further tip metal damage. These data could lead to improved manufacturing processes or novel electrode designs towards the goal of a stable BCI electrical interface, spanning a multi-decade participant lifetime.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Neuroscience and Neural Engineering · EEG and Brain-Computer Interfaces · Muscle activation and electromyography studies
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (82)
Show source-provided references
1. The Utah Intracortical Electrode Array: a recording structure for potential brain-computer interfaces10.1016/S0013-4694(96)95176-0 ↗
2. Restoration of neural output from a paralyzed patient by a direct brain connection10.1097/00001756-199806010-00007 ↗
3. Real-time control of a robotic arm by neuronal ensembles10.1038/10131 ↗
4. Real-time prediction of hand tranjectory by ensembles of cortical neurons in primate10.1038/35042582 ↗
5. Instant neural control of a movement signal10.1038/416141a ↗
6. Direct cortical control of 3D neuroprosthetic devices10.1126/science.1070291 ↗
7. Cortical control of a prosthetic arm for self-feeding10.1038/nature06996 ↗
8. Reach and grasp by people with tetraplegia using a neurally controlled robotic arm10.1038/nature11076 ↗
9. High-performance neuroprosthetic control by an individual with tetraplegia10.1016/S0140-6736(12)61816-9 ↗
10. Ten-dimensional anthropomorphic arm control in a human brain−machine interface: difficulties, solutions, and limitations10.1088/1741-2560/12/1/016011 ↗
11. R.A. Andersen, T. Aflalo, L. Bashford, D. Bjånes, S. Kellis, Exploring cognition with brain–Machine interfaces, doi:10.1146/Annurev-Psych-030221-030214 73 (2022) 131–158. doi:10.1146/ANNUREV-PSYCH-030221-030214.10.1146/annurev-psych-030221-030214 ↗
12. Cognitive control signals for neural prosthetics10.1126/science.1097938 ↗
13. High performance communication by people with paralysis using an intracortical brain-computer interface10.7554/eLife.18554 ↗
14. S.K. Wandelt, D.A. Bjånes, K. Pejsa, B. Lee, C. Liu, R.A. Andersen, Online internal speech decoding from single neurons in a human participant, MedRxiv (2022) 2022.11.02.22281775. doi:10.1101/2022.11.02.22281775.10.1101/2022.11.02.22281775 ↗
15. Clinical applications of penetrating neural interfaces and Utah Electrode Array technologies10.1088/1741-2560/13/6/061003 ↗
16. Proprioceptive and cutaneous sensations in humans elicited by intracortical microstimulation
17. Intracortical microstimulation of human somatosensory cortex10.1126/scitranslmed.aaf8083 ↗
18. Intracortical somatosensory stimulation to elicit fingertip sensations in an individual with spinal cord injury
19. Visual percepts evoked with an intracortical 96-channel microelectrode array inserted in human occipital cortex10.1172/JCI151331 ↗
20. Shape perception via a high-channel-count neuroprosthesis in monkey visual cortex10.1126/science.abd7435 ↗
21. Neural stimulation and recording performance in human sensorimotor cortex over 1500 days10.1088/1741-2552/ac18ad ↗
22. Neural control of cursor trajectory and click by a human with tetraplegia 1000 days after implant of an intracortical microelectrode array10.1088/1741-2560/8/2/025027 ↗
23. Intracortical recording stability in human brain-computer interface users10.1088/1741-2552/aab7a0 ↗
24. Chronic stability of a neuroprosthesis comprising multiple adjacent Utah arrays in monkeys10.1088/1741-2552/ace07e ↗
25. Reliability of signals from a chronically implanted, silicon-based electrode array in non-human primate primary motor cortex10.1109/TNSRE.2005.857687 ↗
26. Long-term stability of neural prosthetic control signals from silicon cortical arrays in rhesus macaque motor cortex10.1088/1741-2560/8/4/045005 ↗
27. Neural control of cursor trajectory and click by a human with tetraplegia 1000 days after implant of an intracortical microelectrode array10.1088/1741-2560/8/2/025027 ↗
28. Failure mode analysis of silicon-based intracortical microelectrode arrays in non-human primates10.1088/1741-2560/10/6/066014 ↗
29. Long-term performance of Utah slanted electrode arrays and intramuscular electromyographic leads implanted chronically in human arm nerves and muscles10.1088/1741-2552/abc025 ↗
30. Characterizing longitudinal changes in the impedance spectra of In-vivo peripheral nerve electrodes10.3390/mi9110587 ↗
31. Abiotic-biotic characterization of Pt/Ir microelectrode arrays in chronic implants10.3389/fneng.2014.00002 ↗
32. Mechanical failure modes of chronically implanted planar silicon-based neural probes for laminar recording10.1016/j.biomaterials.2014.10.040 ↗
33. Recent advances in neural electrode–tissue interfaces10.1016/j.cobme.2017.09.003 ↗
34. Response of brain tissue to chronically implanted neural electrodes10.1016/j.jneumeth.2005.08.015 ↗
35. Glial responses to implanted electrodes in the brain10.1038/s41551-017-0154-1 ↗
36. Neuropathological effects of chronically implanted, intracortical microelectrodes in a tetraplegic patient10.1088/1741-2552/ac127e ↗
37. A. Woolley, H. Desai, K.O.-J. of neural engineering, undefined 2013, chronic intracortical microelectrode arrays induce non-uniform, depth-related tissue responses, iopscience.Iop.OrgAJ Woolley, HA Desai, KJ OttoJournal Neural Eng. 2013•iopscience.Iop.Org 10 (2013). doi:10.1088/1741-2560/10/2/026007.10.1088/1741-2560/10/2/026007 ↗
38. Scanning electron microscopy of chronically implanted intracortical microelectrode arrays in non-human primates10.1088/1741-2560/13/2/026003 ↗
39. Classifying intracortical brain-machine interface signal disruptions based on system performance and applicable compensatory strategies: a review10.3389/fnbot.2020.558987 ↗
40. Large-scale recording of neuronal ensembles10.1038/nn1233 ↗
41. Brain-controlled interfaces: movement restoration with neural prosthetics10.1016/j.neuron.2006.09.019 ↗
42. Chronic recording capability of the utah intracortical electrode array in cat sensory cortex10.1016/S0165-0270(98)00031-4 ↗
43. A. Degenhart, J. Eles, R. Dum, … J.M.-J. of neural, undefined 2016, histological evaluation of a chronically-implanted electrocorticographic electrode grid in a non-human primate, Iopscience.Iop.OrgAD Degenhart, J Eles, R Dum, JL Mischel, I Smalianchuk, B Endler, RC AshmoreJournal Neural Eng. 2016•iopscience.Iop.Org 13 (2016). doi:10.1088/1741-2560/13/4/046019.10.1088/1741-2560/13/4/046019 ↗
44. Long-term implants of Parylene-C coated microelectrodes10.1007/BF02441836 ↗
45. Long-term reliability of Al2O3 and Parylene C bilayer encapsulated Utah electrode array based neural interfaces for chronic implantation10.1088/1741-2560/11/2/026016 ↗
46. Characterization of parylene-C degradation mechanisms: in vitro reactive accelerated aging model compared to multiyear in vivo implantation10.1016/j.biomaterials.2019.119731 ↗
47. Neural electrode resilience against dielectric damage may be improved by use of highly doped silicon as a conductive material10.1016/j.jneumeth.2017.10.002 ↗
48. Unprotected sidewalls of implantable silicon-based neural probes and conformal coating as a solution10.1038/s41529-021-00154-9 ↗
49. In vitro comparison of sputtered iridium oxide and platinum-coated neural implantable microelectrode arrays10.1088/1748-6041/5/1/015007 ↗
50. Factors affecting degradation of sputtered iridium oxide used for neuroprosthetic applications
51. Ultramicro-sized sputtered iridium oxide electrodes in buffered saline: behavior, stability, and the effect of the perimeter to area ratio on their electrochemical properties10.1016/j.electacta.2022.140514 ↗
52. Flexible IrOx neural electrode for mouse vagus nerve stimulation10.1016/j.actbio.2023.01.026 ↗
53. Neural stimulation and recording electrodes10.1146/annurev.bioeng.10.061807.160518 ↗
54. Structural analysis of explanted microelectrode arrays
55. Explant analysis of Utah electrode arrays implanted in Human cortex for brain-computer-interfaces10.3389/fbioe.2021.759711 ↗
56. Rapid evaluation of the durability of cortical neural implants using accelerated aging with reactive oxygen species10.1088/1741-2560/12/2/026003 ↗
57. Automated reactive accelerated aging for rapid in vitro evaluation of neural implant performance10.1063/1.5024686 ↗
58. Comprehensive characterization and failure modes of tungsten microwire arrays in chronic neural implants10.1088/1741-2560/9/5/056015 ↗
59. Quantifying long-term microelectrode array functionality using chronic in vivo impedance testing10.1088/1741-2560/9/2/026028 ↗
60. Blood brain barrier (BBB)-disruption in intracortical silicon microelectrode implants10.1016/j.biomaterials.2018.02.036 ↗
61. Neuroinflammation, oxidative stress, and blood-brain barrier (BBB) disruption in acute Utah electrode array implants and the effect of deferoxamine as an iron chelator on acute foreign body response10.1016/j.biomaterials.2018.09.040 ↗
62. In vitro comparison of sputtered iridium oxide and platinum-coated neural implantable microelectrode arrays10.1088/1748-6041/5/1/015007 ↗
63. Decoding motor imagery from the posterior parietal cortex of a tetraplegic human10.1126/science.aaa5417 ↗
64. Implicit mechanisms of intention10.1016/j.cub.2022.03.047 ↗
65. Impedance and noise characterizations of Utah and microwire electrode arrays10.1109/JERM.2018.2862417 ↗
66. Root mean square error (RMSE) or mean absolute error (MAE)?-arguments against avoiding RMSE in the literature10.5194/gmd-7-1247-2014 ↗
67. The effects of chronic intracortical microstimulation on neural tissue and fine motor behavior10.1088/1741-2560/12/6/066018 ↗
68. M.S. Fifer, D.P. McMullen, T.M. Thomas, L.E. Osborn, R.W. Nickl, D.N. Candrea, E.A. Pohlmeyer, M.C. Thompson, M. Anaya, W. Schellekens, N.F. Ramsey, S.J. Bensmaia, W.S. Anderson, B.A. Wester, N.E. Crone, P.A. Celnik, G.L. Cantarero, F.V. Tenore, Intracortical microstimulation elicits Human fingertip sensations, MedRxiv (2020) 2020.05.29.20117374. doi:10.1101/2020.05.29.20117374.10.1101/2020.05.29.20117374 ↗
69. Intracortical microstimulation of somatosensory cortex generates evoked responses in motor cortex
70. Perceived timing of cutaneous vibration and intracortical microstimulation of human somatosensory cortex10.1016/j.brs.2022.05.015 ↗
71. Notes on continuous stochastic phenomena10.1093/biomet/37.1-2.17 ↗
72. Long-term measurement of impedance in chronically implanted depth and subdural electrodes during responsive neurostimulation in humans10.1016/j.brs.2013.02.001 ↗
73. Fast, three-dimensional super-resolution imaging of live cells10.1038/nmeth.1605 ↗
74. An optimization-based study of equivalent circuit models for representing recordings at the neuron-electrode interface10.1109/TBME.2012.2203820 ↗
75. A physical-based equivalent circuit model for an organic/electrolyte interface10.1016/j.orgel.2016.05.018 ↗
76. Relationship between microelectrode array impedance and chronic recording quality of single units and local field potentials
77. Unique electrophysiological and impedance signatures between encapsulation types: an analysis of biological Utah array failure and benefit of a biomimetic coating in a rat model10.1016/j.biomaterials.2018.01.025 ↗
78. Freiman · 2012
79. Silicon corrosion in neutral Media: the influence of confined geometries and crevice corrosion in simulated physiological solutions10.1149/2.0241906jes ↗
80. Electrical performance of penetrating microelectrodes chronically implanted in cat cortex10.1109/TBME.2013.2248152 ↗
81. Acute microelectrode array implantation into human neocortex: preliminary technique and histological considerations10.3171/foc.2006.20.5.5 ↗
82. Approaches to a cortical vision prosthesis: implications of electrode size and placement10.1088/1741-2560/13/2/025003 ↗