journal article·2013-11

Process of prototyping coronary stents from biodegradable Fe–Mn alloys

Hendra Hermawan, Diego Mantovani

Acta Biomaterialia· Volume 9· Issue 10
Publisher page

Abstract

Biodegradable stents are considered to be a recent innovation, and their feasibility and applicability have been proven in recent years. Research in this area has focused on materials development and biological studies, rather than on how to transform the developed biodegradable materials into the stent itself. Currently available stent technology, the laser cutting-based process, might be adapted to fabricate biodegradable stents. In this work, the fabrication, characterization and testing of biodegradable Fe-Mn stents are described. A standard process for fabricating and testing stainless steel 316L stents was referred to. The influence of process parameters on the physical, metallurgical and mechanical properties of the stents, and the quality of the produced stents, were investigated. It was found that some steps of the standard process such as laser cutting can be directly applied, but changes to parameters are needed for annealing, and alternatives are needed to replace electropolishing.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Corrosion Behavior and Inhibition · Hydrogen embrittlement and corrosion behaviors in metals · Electrodeposition and Electroless Coatings

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (40)

Show source-provided references
1. A randomized comparison of coronary-stent placement and balloon angioplasty in the treatment of coronary artery disease. Stent Restenosis Study Investigators10.1056/NEJM199408253310802 ↗
2. A comparison of balloon-expandable-stent implantation with balloon angioplasty in patients with coronary artery disease. Benestent Study Group10.1056/NEJM199408253310801 ↗
3. Degradable metallic biomaterials: the concept, current developments and future directions
4. Developments in metallic biodegradable stents10.1016/j.actbio.2009.10.006 ↗
5. Update on in-stent restenosis
6. Four-year experience with Palmaz–Schatz stenting in coronary angioplasty complicated by dissection with threatened or present vessel closure10.1161/01.CIR.90.6.2716 ↗
7. Drug-eluting stents: caution and concerns for long-term outcome10.1097/00019501-200409000-00003 ↗
8. Patterns and mechanisms of in-stent restenosis. A serial intravascular ultrasound study10.1161/01.CIR.94.6.1247 ↗
9. Update on bioabsorbable stents: from bench to clinical10.1111/j.1540-8183.2006.00187.x ↗
10. A novel approach to temporary stenting: degradable cardiovascular stents produced from corrodible metal – results 6–18 months after implantation into New Zealand white rabbits10.1136/heart.86.5.563 ↗
11. Biocorrosion of magnesium alloys: a new principle in cardiovascular implant technology?10.1136/heart.89.6.651 ↗
12. Interventional cardiac catheterization at Duke Medical Center10.1016/0002-9149(88)91529-9 ↗
13. Temporary scaffolding of coronary arteries with bioabsorbable magnesium stents: a prospective, non-randomised multicentre trial10.1016/S0140-6736(07)60853-8 ↗
14. A bioabsorbable everolimus-eluting coronary stent system for patients with single de-novo coronary artery lesions (ABSORB): a prospective open-label trial10.1016/S0140-6736(08)60415-8 ↗
15. AMS INSIGHT – absorbable metal stent implantation for treatment of below-the-knee critical limb ischemia: 6-month analysis10.1007/s00270-009-9530-x ↗
16. First successful implantation of a biodegradable metal stent into the left pulmonary artery of a preterm baby10.1002/ccd.20520 ↗
17. Bioabsorbable metal stents for percutaneous treatment of critical recoarctation of the aorta in a newborn10.1002/ccd.20756 ↗
18. Preliminary results after application of absorbable metal stents in patients with critical limb ischemia10.1583/04-1349R.1 ↗
19. Initial and 6-month results of biodegradable poly-l-lactic acid coronary stents in humans10.1161/01.CIR.102.4.399 ↗
20. Marked inflammatory sequelae to implantation of biodegradable and nonbiodegradable polymers in porcine coronary arteries10.1161/01.CIR.94.7.1690 ↗
21. Investigation of corrosion behaviour of magnesium alloy AM60B-F under pseudo-physiological conditions10.4028/www.scientific.net/MSF.426-432.521 ↗
22. In vivo corrosion of four magnesium alloys and the associated bone response10.1016/j.biomaterials.2004.09.049 ↗
23. Corrosion behavior of biomedical AZ91 magnesium alloy in simulated body fluids10.1557/jmr.2007.0233 ↗
24. Drug-eluting bioabsorbable magnesium stent10.1111/j.1540-8183.2004.04081.x ↗
25. Safety and efficacy of bioabsorbable magnesium alloy stents in porcine coronary arteries10.1002/ccd.20727 ↗
26. Design strategy for microalloyed ultra-ductile magnesium alloys for medical applications10.4028/www.scientific.net/MSF.618-619.75 ↗
27. Microstructure, mechanical properties and bio-corrosion properties of Mg–Zn–Mn–Ca alloy for biomedical application10.1016/j.msea.2008.06.019 ↗
28. The development of binary Mg–Ca alloys for use as biodegradable materials within bones10.1016/j.biomaterials.2007.12.021 ↗
29. Long-term biocompatibility of a corrodible peripheral iron stent in the porcine descending aorta10.1016/j.biomaterials.2006.05.029 ↗
30. Iron-manganese: new class of degradable metallic biomaterials prepared by powder metallurgy10.1179/174329008X284868 ↗
31. Design strategy for biodegradable Fe-based alloys for medical applications10.1016/j.actbio.2009.07.039 ↗
32. Fe–Mn alloys for metallic biodegradable stents: degradation and cell viability studies10.1016/j.actbio.2009.11.025 ↗
33. Patents on metallic biodegradable stents
34. Degradable metallic biomaterials: design and development of Fe–Mn alloys for stents10.1002/jbm.a.32224 ↗
35. ASTM. ASTM E 8: Standard test methods for tension testing of metallic materials. West Conshohocken, PA: ASTM International; 2004.
36. Poncin P, Proft J. Stent tubing: understanding the desired attributes. Materials & processes for medical devices conference, California, USA, Materials Park, OH: ASM International; 2003.
37. The importance of annealing 316 LVM stent
38. Brandes · 1992
39. Damping capacity in Fe–Mn binary alloys10.2355/isijinternational.37.1023 ↗