Cobalt from metal-on-metal hip replacements may be the clinically relevant active agent responsible for periprosthetic tissue reactions
Source abstract
Abstract
Some types of metal-on-metal (MOM) hip replacements have unacceptably high rates of failure, such as the Ultima TPS MOM hip, with 13.8% failure at 5 years. This has been attributed to an inflammatory reaction following the release of cobalt (Co) and chromium (Cr) from the bearing surfaces and modular junctions. There is in vitro evidence that Co is more important than Cr in the inflammatory process, but there are no reported human tissue studies of the analysis of implant-derived metals.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Advanced materials and composites · Orthopedic Infections and Treatments
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (23)
Show source-provided references
1. National Joint Registry. 6th Annual Report. Available at: http://www.njrcentre.org.uk/NjrCentre/Portals/0/Sixth%20annual%20NJR%20report.pdf, 2009.
2. BHS. Updated Guidance on Large Diameter Metal on Metal Bearing Total Hip Replacements. Available at: http://www.britishhipsociety.com/MoMUpdate.htm, 2011.
3. The John Charnley Award: a study of implant failure in metal-on-metal surface arthroplasties10.1097/01.blo.0000238777.34939.82 ↗
4. Contact stresses in a patient-specific unicompartmental knee replacement
5. Retrieval analysis of 240 metal-on-metal hip components, comparing modular total hip replacement with hip resurfacing10.1302/0301-620X.93B3.25551 ↗
6. Sensitivity and specificity of blood cobalt and chromium metal ions for predicting failure of metal-on-metal hip replacement10.1302/0301-620X.93B10.26249 ↗
7. Soluble and particulate Co–Cr–Mo alloy implant metals activate the inflammasome danger signaling pathway in human macrophages: a novel mechanism for implant debris reactivity10.1002/jor.20826 ↗
8. The chemical form of metallic debris in tissues surrounding metal-on-metal hips with unexplained failure10.1016/j.actbio.2010.06.006 ↗
9. Using synchrotron X-ray fluorescence microprobes in the study of metal homeostasis in plants10.1093/aob/mcn264 ↗
10. Early failure of the Ultima metal-on-metal total hip replacement in the presence of normal plain radiographs10.1302/0301-620X.92B11.24504 ↗
11. Three-dimensional tomographic imaging and characterization of iron compounds within Alzheimer’s plaque core material10.3233/JAD-2008-14211 ↗
12. I18 – the microfocus spectroscopy beamline at the diamond light source10.1107/S0909049509032282 ↗
13. Goldstein · 2002
14. Concentrations of Fe, Cu and Zn in breast tissue: a synchrotron XRF study10.1088/0031-9155/47/13/310 ↗
15. Iron, copper, zinc and bromine mapping in cirrhotic liver slices from patients with hemochromatosis studied by microscopic synchrotron radiation X-ray fluorescence analysis in continuous scanning mode10.1016/j.sab.2007.03.031 ↗
16. Tenderholt AHB, Hodgson KO. PySpline: a modern, cross-platform program for the processing of raw averaged XAS edge and EXAFS data. In: Hedman BP, editor. Proceedings of the 13th International Conference on X-ray Absorption Fine Structure – XAFS13. Melville, NY: American Institute of Physics; 2007, pp. 105–7.10.1063/1.2644442 ↗
17. 1988
18. ATHENA, ARTEMIS, HEPHAESTUS: data analysis for X-ray absorption spectroscopy using IFEFFIT10.1107/S0909049505012719 ↗
19. Cobalt speciation study in the cobalt–cysteine system by electrospray ionization mass spectrometry and anion-exchange chromatography inductively coupled plasma atomic emission spectrometry10.1366/0003702053945912 ↗
20. Does corrosion matter?10.1302/0301-620X.70B4.3403590 ↗
21. Wear particles from metal-on-metal total hip replacements: effects of implant design and implantation time10.1243/09544119JEIM112 ↗
22. In vivo corrosion of cobalt–chromium and titanium wear particles10.1302/0301-620X.77B6.7593115 ↗
23. Migration of corrosion products from modular hip prostheses. Particle microanalysis and histopathological findings10.2106/00004623-199409000-00009 ↗