Bone attachment to glass-fibre-reinforced composite implant with porous surface
Source abstract
Abstract
A method has recently been developed for producing fibre-reinforced composites (FRC) with porous surfaces, intended for use as load-bearing orthopaedic implants. This study focuses on evaluation of the bone-bonding behaviour of FRC implants. Three types of cylindrical implants, i.e. FRC implants with a porous surface, solid polymethyl methacrylate (PMMA) implants and titanium (Ti) implants, were inserted in a transverse direction into the intercondular trabeculous bone area of distal femurs and proximal tibias of New Zealand White rabbits. Animals were sacrificed at 3, 6 and 12 weeks post operation, and push-out tests (n=5-6 per implant type per time point) were then carried out. At 12 weeks the shear force at the porous FRC-bone interface was significantly higher (283.3+/-55.3N) than the shear force at interfaces of solid PMMA/bone (14.4+/-11.0 N; p<0.001) and Ti/bone (130.6+/-22.2N; p=0.001). Histological observation revealed new bone growth into the porous surface structure of FRC implants. Solid PMMA and Ti implants were encapsulated mostly with fibrous connective tissue. Finite element analysis (FEA) revealed that porous FRC implants had mechanical properties which could be tailored to smooth the shear stress distribution at the bone-implant interface and reduce the stress-shielding effect.
Source: europepmc · Quality: candidate unverified · Source record ↗
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Bone fractures and treatments · Bone Tissue Engineering Materials
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (34)
Show source-provided references
1. Strain distribution in the proximal femur with flexible composite and metallic femoral components under axial and torsional loads10.1002/jbm.820270504 ↗
2. The relationship between stress shielding and bone resoption around total hip stems and the effects of flexible material10.1097/00003086-199201000-00014 ↗
3. Composite technology in load-bearing orthopaedic implants10.1016/S0142-9612(97)00217-2 ↗
4. About composite materials and their use in bone surgery10.1016/S0020-1383(00)80023-6 ↗
5. Polymer-based composite hip prostheses10.1016/S0162-0134(99)00228-7 ↗
6. Biological evaluation of carbon-fibre-reinforced polybutyleneterephthalate (CFRPBT) employed in a novel acetabular cup10.1016/j.biomaterials.2003.10.064 ↗
7. The use of a carbon fiber plate for periprosthetic supracondylar femoral fractures10.1054/arth.2002.30291 ↗
8. Carbon fibre plates in the treatment of femoral periprosthetic fractures10.1016/j.injury.2003.10.014 ↗
9. Early failure of a carbon-fiber composite femoral component10.1016/S0883-5403(97)90038-3 ↗
10. Early failure of a press-fit carbon fiber hip prosthesis with a smooth surface10.1054/arth.2002.30285 ↗
11. Isolation of polyacetal wear particles from periprosthetic tissue of isoelastic femoral stems10.1302/0301-620X.83B8.12087 ↗
12. In vitro cytotoxicity of fiber-polymethyl methacrylate composite used in dentures10.1046/j.1365-2842.1999.00431.x ↗
13. Comparison of two different silane compounds used for improving adhesion between fibres and acrylic denture base material10.1111/j.1365-2842.1993.tb01640.x ↗
14. Anchorage of the femoral head prosthesis to the shaft of the femur10.1302/0301-620X.42B1.28 ↗
15. Fibre-reinforced composite implant: in vitro mechanical interlocking with bone model material and residual monomer analysis10.1007/s10853-006-7020-y ↗
16. Surface porous fibre-reinforced composite bulk bone substitute. In vitro studies and in vivo evaluation in segment defect10.1007/s10561-004-0214-0 ↗
17. Polymer scaffolds with interconnected spherical pores and controlled architecture for tissue engineering: fabrication, mechanical properties, and finite element modeling10.1002/jbm.b.30683 ↗
18. Analysis of push-out test data based on interfacial fracture energy10.1023/A:1008929201918 ↗
19. A high-modulus polymer for porous orthopedic implants: biomechanical compatibility of porous implants10.1002/jbm.820120508 ↗
20. Fixation of carbon fibre-reinforced carbon composite implanted into bone10.1023/A:1018526226382 ↗
21. Failed innovation in total hip replacement. Diagnosis and proposals for a cure10.3109/17453679308994602 ↗
22. Finite element modelling – predictor of implant survival?10.1007/BF00134322 ↗
23. Influence of the stiffness of bone defect implants on the mechanical conditions at the interface – a finite element analysis with contact10.1016/S0021-9290(03)00114-3 ↗
24. The effects of trabecular-bone loading variables on the surface signaling potential for bone remodeling and adaptation10.1007/s10439-005-8964-9 ↗
25. Numerical simulation of influence of chosen parameters on tensile stresses in bone cement layer in total hip arthroplasty
26. Mechanical properties and fracture surface morphologies in unnotched specimens of rubber-PMMA composites10.1007/s11665-007-9042-2 ↗
27. Christensen · 1979
28. Magnitudes of local stress and strain along bony surfaces predict the cource and type of fracture healing10.1016/S0021-9290(98)00153-5 ↗
29. Flexural properties of acrylic polymers reinforced with unidirectional and woven glass fibers10.1016/S0022-3913(99)70276-3 ↗
30. Composite hip prosthesis design. II. Simulation10.1002/(SICI)1097-4636(199801)39:1<102::AID-JBM13>3.0.CO;2-H ↗
31. Correlation of bony ingrowth to the distribution of stress and strain parameters surrounding a porous-coated implant10.1002/jor.1100140604 ↗
32. Review: if bone is the answer, then what is the question?10.1046/j.1469-7580.2000.19720145.x ↗
33. Effects of mechanical forces on maintenance and adaptation of form in trabecular bone10.1038/35015116 ↗
34. Bioactive glass fiber/polymeric composites bond to bone tissue10.1002/(SICI)1097-4636(199801)39:1<161::AID-JBM18>3.0.CO;2-I ↗