Impact of nitinol stent surface processing on in-vivo nickel release and biological response

Srinidhi Nagaraja, Stacey J.L. Sullivan, Philip R. Stafford, Anne D. Lucas, Elon Malkin

Acta Biomaterialia· Volume 72
Publisher page

Abstract

Although nitinol is widely used in percutaneous cardiovascular interventions, a causal relationship between nickel released from implanted cardiovascular devices and adverse systemic or local biological responses has not been established. The objective of this study was to investigate the relationship between nitinol surface processing, in-vivo nickel release, and biocompatibility. Nitinol stents manufactured using select surface treatments were implanted into the iliac arteries of minipigs for 6 months. Clinical chemistry profile, complete blood count, serum and urine nickel analyses were performed periodically during the implantation period. After explant, stented arteries were either digested and analyzed for local nickel concentration or fixed and sectioned for histopathological analysis of stenosis and inflammation within the artery. The results indicated that markers for liver and kidney function were not different than baseline values throughout 180 days of implantation regardless of surface finish. In addition, white blood cell, red blood cell, and platelet counts were similar to baseline values for all surface finishes. Systemic nickel concentrations in serum and urine were not significantly different between processing groups and comparable to baseline values during 180 days of implantation. However, stents with non-optimized surface finishing had significantly greater nickel levels in the surrounding artery compared to polished stents. These stents had increased stenosis with potential for local inflammation compared to polished stents. These findings demonstrate that proper polishing of nitinol surfaces can reduce in-vivo nickel release locally, which may aid in minimizing adverse inflammatory reactions and restenosis. Statement of significance Nitinol is a commonly used material in cardiovascular medical devices. However, relationships between nitinol surface finishing, in-vivo metal ion release, and adverse biological responses have yet to be established. We addressed this knowledge gap by implanting single and overlapped nitinol stents with different surface finishes to assess systemic impact on minipigs (i.e. serum and urine nickel levels, liver and kidney function, immune and blood count) over the 6 month implantation period. In addition, nickel levels and histopathology in stented arteries were analyzed on explant to determine relationships between surface processing and local adverse tissue reactions. The findings presented here highlight the importance of surface processing on in-vivo nickel release and subsequent impact on local biological response for nitinol implants.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Bone Tissue Engineering Materials · Shape Memory Alloy Transformations · Coronary Interventions and Diagnostics

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (46)

Show source-provided references
1. Nitinol medical devices and implants10.3109/13645700009063054 ↗
2. An overview of nitinol medical applications10.1016/S0921-5093(99)00294-4 ↗
3. A guide to shape memory and superelasticity in Nitinol medical devices10.1080/13645700410017236 ↗
4. Corroded nitinol wires in explanted aortic endografts: An important mechanism of failure?10.1177/152660280100800303 ↗
5. What can we learn from explanted endovascular devices?10.1053/ejvs.2002.1677 ↗
6. The corrosion of Nitinol by exposure to decontamination solutions. Shape Memory and Superelastic Technologies
7. Scanning electron microscopy analysis of corrosion of stainless steel and nitinol Stents from autopsy retrievals
8. In-vivo corrosion and local release of metallic ions from vascular stents into surrounding tissue
9. Surface characterization of explanted endovascular stents: evidence of in vivo corrosion10.1002/jbm.b.31698 ↗
10. Nickel release, a possible indicator for the duration of antiplatelet treatment, from a nickel cardiac device in vivo: a study in patients with atrial septal defects implanted with an Amplatzer occluder10.5414/CPP44107 ↗
11. Nickel release after implantation of the Amplatzer occluder10.1067/mhj.2003.7 ↗
12. Human biomonitoring reference values for metals and trace elements in blood and urine derived from the Canadian Health Measures Survey 2007–201310.1016/j.ijheh.2016.10.006 ↗
13. In vitro host response assessment of biomaterials for cardiovascular stent manufacture10.1023/B:JMSM.0000021123.51752.11 ↗
14. Relating nickel-induced tissue inflammation to nickel release in vivo10.1002/jbm.1052 ↗
15. Pericarditis associated with nickel hypersensitivity to the Amplatzer occluder device: a case report
16. Nickel allergy in interatrial shunt device-based closure patients10.1111/j.1747-0803.2007.00134.x ↗
17. Metal-induced generalized pruriginous dermatitis and endovascular surgery10.1034/j.1600-0536.2000.043001035.x ↗
18. Five-year experience with percutaneous closure of patent foramen ovale10.1016/j.amjcard.2006.12.054 ↗
19. Systemic allergic reaction to the percutaneous patent foramen ovale occluder10.1067/mtc.2003.125 ↗
20. A case of allergic contact dermatitis following transcatheter closure of patent ductus arteriosus using Amplatzer ductal occluder10.1016/j.ijcard.2007.04.090 ↗
21. Metal allergy to Amplatzer occluder device presented as severe bronchospasm10.1016/j.athoracsur.2009.05.044 ↗
22. Superficial femoral artery nitinol stent in a patient with nickel allergy10.1016/j.jvs.2013.01.041 ↗
23. Nickel and molybdenum contact allergies in patients with coronary in-stent restenosis10.1016/S0140-6736(00)03262-1 ↗
24. Evaluation of metal allergies in patients with coronary stents10.1034/j.1600-0536.2002.470607.x ↗
25. The impact of metallic allergy on stent implantation: metal allergy and recurrence of in-stent restenosis10.1016/j.ijcard.2004.12.034 ↗
26. Outcomes after coronary stent implantation in patients with metal allergy
27. Unusual presentation of nickel allergy requiring explantation of an Amplatzer atrial septal occluder device10.1002/clc.20427 ↗
28. Current practices in corrosion, surface characterization, and nickel leach testing of cardiovascular metallic implants10.1002/jbm.b.33630 ↗
29. Oxidation of nitinol and its effect on corrosion resistance
30. Effects of oxide layer composition and radial compression on nickel release in nitinol stents10.1007/s40830-015-0028-x ↗
31. Improvement of the corrosion resistance of NiTi stents by surface treatments
32. Effect of modification of oxide layer on NiTi stent corrosion resistance10.1002/(SICI)1097-4636(199824)43:4<433::AID-JBM11>3.0.CO;2-# ↗
33. Analysis of the effects of surface treatments on nickel release from nitinol wires and their impact on candidate gene expression in endothelial cells, Journal of materials science10.1007/s10856-006-0087-9 ↗
34. Influence of nitinol wire surface treatment on oxide thickness and composition and its subsequent effect on corrosion resistance and nickel ion release10.1002/jbm.a.30720 ↗
35. The effect of material removal on the corrosion resistance and biocompatibility of nitinol laser-cut and wire-form products10.1007/s11665-011-9882-7 ↗
36. The effects of surface processing on in-vivo corrosion of Nitinol stents in a porcine model10.1016/j.actbio.2017.08.029 ↗
37. Effect of surface treatment and sterilization processes on the corrosion behavior of NiTi shape memory alloy10.1002/1097-4636(20000915)51:4<685::AID-JBM17>3.0.CO;2-S ↗
38. A novel paclitaxel-eluting porous carbon–carbon nanoparticle coated, nonpolymeric cobalt–chromium stent: Evaluation in a porcine model
39. Safety and efficacy of bioabsorbable magnesium alloy stents in porcine coronary arteries
40. Optical coherence tomography and intravascular ultrasound imaging of bioabsorbable magnesium stent degradation in porcine coronary arteries10.1016/j.carrev.2008.04.001 ↗
41. Sirolimus-eluting stents for the treatment of obstructive superficial femoral artery disease: six-month results10.1161/01.CIR.0000029746.10018.36 ↗
42. Assessment of self-expanding nitinol stent deformation after chronic implantation into the femoropopliteal arteries10.4244/EIJV9I6A117 ↗
43. Late stent expansion and neointimal proliferation of oversized Nitinol stents in peripheral arteries10.1007/s00270-009-9601-z ↗
44. Effects of tissue digestion solutions on surface properties of nitinol stents
45. Polymer-free biolimus a9-coated stent demonstrates more sustained intimal inhibition, improved healing, and reduced inflammation compared with a polymer-coated sirolimus-eluting cypher stent in a porcine model
46. Comparison of the ovine and porcine animal models for biocompatibility testing of vascular prostheses10.1016/j.jss.2004.10.021 ↗