journal article·2017-06

A strain-mediated corrosion model for bioabsorbable metallic stents

E. Galvin, D. O'Brien, C. Cummins, B.J. Mac Donald, C. Lally

Acta Biomaterialia· Volume 55
Publisher page

Abstract

This paper presents a strain-mediated phenomenological corrosion model, based on the discrete finite element modelling method which was developed for use with the ANSYS Implicit finite element code. The corrosion model was calibrated from experimental data and used to simulate the corrosion performance of a WE43 magnesium alloy stent. The model was found to be capable of predicting the experimentally observed plastic strain-mediated mass loss profile. The non-linear plastic strain model, extrapolated from the experimental data, was also found to adequately capture the corrosion-induced reduction in the radial stiffness of the stent over time. The model developed will help direct future design efforts towards the minimisation of plastic strain during device manufacture, deployment and in-service, in order to reduce corrosion rates and prolong the mechanical integrity of magnesium devices. Statement of significance The need for corrosion models that explore the interaction of strain with corrosion damage has been recognised as one of the current challenges in degradable material modelling (Gastaldi et al., 2011). A finite element based plastic strain-mediated phenomenological corrosion model was developed in this work and was calibrated based on the results of the corrosion experiments. It was found to be capable of predicting the experimentally observed plastic strain-mediated mass loss profile and the corrosion-induced reduction in the radial stiffness of the stent over time. To the author's knowledge, the results presented here represent the first experimental calibration of a plastic strain-mediated corrosion model of a corroding magnesium stent.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Magnesium Alloys: Properties and Applications · Elasticity and Material Modeling · Metal Forming Simulation Techniques

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (33)

Show source-provided references
1. Baboian · 1995
2. Continuum damage model for bioresorbable magnesium alloy devices – application to coronary stents10.1016/j.jmbbm.2010.11.003 ↗
3. Formulation of a macroscale corrosion damage internal state variable model10.1016/j.ijsolstr.2013.12.007 ↗
4. Fibre laser cutting and chemical etching of AZ31 for manufacturing biodegradable stents
5. Testing and mesoscale modelling of hydrogen assisted cracking of magnesium10.1016/j.engfracmech.2009.07.009 ↗
6. Microstructure-based multistage fatigue modeling of a cast AE44 magnesium alloy10.1016/j.ijfatigue.2006.07.005 ↗
7. An arbitrary Lagrangian-Eulerian model for modelling the time-dependent evolution of crevice corrosion10.1016/j.corsci.2013.10.003 ↗
8. Finite element shape optimization for biodegradable magnesium alloy stents10.1007/s10439-010-0057-8 ↗
9. Finite element analyses for design evaluation of biodegradable magnesium alloy stents in arterial vessels10.1016/j.mseb.2011.03.013 ↗
10. Experimental data confirm numerical modeling of the degradation process of magnesium alloys stents10.1016/j.actbio.2012.10.035 ↗
11. A corrosion model for bioabsorbable metallic stents10.1016/j.actbio.2011.05.032 ↗
12. A physical corrosion model for bioabsorbable metal stents10.1016/j.actbio.2013.12.059 ↗
13. Stress corrosion cracking of rare-earth containing magnesium alloys ZE41, QE22 and Elektron 21 (EV31A) compared with AZ8010.1016/j.msea.2007.07.070 ↗
14. Characterisation of stress corrosion cracking (SCC) of Mg-Al alloys10.1016/j.msea.2007.11.064 ↗
15. Effect of compression deformation on the microstructure and corrosion behavior of magnesium alloys10.1016/j.jallcom.2012.03.010 ↗
16. Corrosion fatigue behaviour of creep-resistant magnesium alloy Mg-4Al-2Ba-2Ca10.1016/j.proeng.2014.06.228 ↗
17. Influence of strain on the corrosion of magnesium alloys and zinc in physiological environments10.1016/j.actbio.2016.10.030 ↗
18. The stress-strain behavior of coronary stent struts is size dependent10.1114/1.1569268 ↗
19. Susceptibility of rare-earth-magnesium alloys to micro-galvanic corrosion10.1016/j.corsci.2010.08.006 ↗
20. Research on an Mg-Zn alloy as a degradable biomaterial10.1016/j.actbio.2009.06.028 ↗
21. A new in vitro-in vivo correlation for bioabsorbable magnesium stents from mechanical behavior10.1016/j.msec.2013.08.042 ↗
22. Plastic strains during stent deployment have a critical influence on the rate of corrosion in absorbable magnesium stents
23. Numerical modelling of convection interacting with a melting and solidification front: application to the thermal evolution of the basal magma ocean10.1016/j.pepi.2012.06.008 ↗
24. On the finite element modelling of balloon-expandable stents10.1016/j.jmbbm.2007.07.002 ↗
25. A general finite element analysis method for balloon expandable stents based on repeated unit cell (RUC) model10.1016/j.finel.2007.01.001 ↗
26. Shape optimization of coronary artery stent based on a parametric model10.1016/j.finel.2009.01.001 ↗
27. Stresses in peripheral arteries following stent placement: a finite element analysis10.1080/10255840802136135 ↗
28. Simulation of a balloon expandable stent in a realistic coronary artery–determination of the optimum modelling strategy10.1016/j.jbiomech.2010.03.050 ↗
29. Continuum models of ductile fracture: a review10.1177/1056789509103482 ↗
30. Performance comparison between a statistical model a deterministic model, and an artificial neural network model for predicting damage from pitting corrosion
31. Degradation behaviour and mechanical properties of magnesium implants in rabbit tibiae10.1007/s10853-009-3936-3 ↗
32. Bioresorbable metal scaffold for cardiovascular application: current knowledge and future perspectives10.1016/j.carrev.2014.01.011 ↗
33. Surface modification of absorbable magnesium stents by reactive ion etching10.1007/s11090-013-9477-1 ↗