Influence of design and bearing material on polyethylene wear particle generation in total knee replacement

S. Utzschneider, A. Paulus, J.-C. Datz, C. Schroeder, B. Sievers, B. Wegener, V. Jansson

Acta Biomaterialia· Volume 5· Issue 7
Publisher page

Abstract

Periprosthetic osteolysis is one of the main reasons for revision of arthroplasty. The osteolytic reaction is influenced by the dose, size and shape of the wear particles. For arthroplasty, a low number and biologically less active particles are required. This is the first study which analyzes the impact of different knee designs, combined with crosslinked polyethylenes (sequentially irradiated and annealed as well as remelted techniques), on the amount, size and shape of particles. Overall, six material combinations, four of them with crosslinked polyethylene (XPE) and two of them with ultra-high molecular weight polyethylene (UHMWPE) inserts, including fixed and mobile bearings, were tested in a knee joint simulator. After isolation nearly 100,000 particles were analyzed in size, shape and number by scanning electron microscopy and image analysis. For all the designs, the wear was predominantly smooth and granular with few fibrillar particles. The Scorpio design with the X3 insert, the Natural Knee II design with the Durasul insert and the LCS design, also combined with a crosslinked polyethylene insert, generated statistically significant (P<0.05) lower particle numbers. The particle size was independent of the radiation dose. The wear generated by the LCS knee design (XPE and UHMWPE) had a higher percentage fraction of particles >1microm in size (equivalent circle diameter). The NexGen design, tested with the Prolong insert, showed a high number of particles in the biologically active size range compared with the other crosslinked designs, which could be a predictor for higher biological reactivity.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Total Knee Arthroplasty Outcomes · Orthopedic Infections and Treatments

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (53)

Show source-provided references
1. The cellular and molecular biology of periprosthetic osteolysis10.1097/01.blo.0000238813.95035.1b ↗
2. Wear, debris, and biologic activity of cross-linked polyethylene in the knee: benefits and potential concerns10.1097/01.blo.0000148783.20469.4c ↗
3. Wear and morphology of ultra-high molecular weight polyethylene wear particles from total hip replacements10.1243/PIME_PROC_1996_210_409_02 ↗
4. Optical analysis of surface changes on early retrievals of highly cross-linked and conventional polyethylene tibial inserts10.1016/S0883-5403(03)00293-6 ↗
5. Polyethylene in knee arthroplasty: what is the future?10.1097/01.blo.0000187811.48717.9d ↗
6. Mechanism and clinical significance of wear debris-induced osteolysis10.1097/00003086-199203000-00003 ↗
7. A rat model of resorption of bone at the cement-bone interface in the presence of polyethylene wear particles10.2106/00004623-198870020-00015 ↗
8. Periprosthetic bone loss in total hip arthroplasty. Polyethylene wear debris and the concept of the effective joint space10.2106/00004623-199274060-00006 ↗
9. Composition and morphology of wear debris in failed uncemented total hip replacement10.1302/0301-620X.76B1.8300684 ↗
10. Isolation of predominantly submicron-sized UHMWPE wear particles from periprosthetic tissues10.1002/jbm.820290118 ↗
11. Highly cross-linked polyethylene in cemented THA: randomized study of 61 hips10.1097/01.blo.0000096802.78689.45 ↗
12. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements10.1002/jor.1100170203 ↗
13. Ex vivo wear of conventional and cross-linked polyethylene acetabular liners10.1097/01.blo.0000164407.02998.8d ↗
14. Surface-gradient cross-linked polyethylene acetabular cups: oxidation resistance and wear against smooth and rough femoral balls10.1097/01.blo.0000149999.17123.e8 ↗
15. Wear and toughness of crosslinked polyethylene for total knee replacements: a study using a simulator and small-punch testing10.1002/jor.20223 ↗
16. Dose effects of cross-linking polyethylene for total knee arthroplasty on wear performance and mechanical properties10.1002/jbm.b.30835 ↗
17. Comparison of wear, wear debris and functional biological activity of moderately crosslinked and non-crosslinked polyethylenes in hip prostheses10.1243/0954411021536333 ↗
18. Particle analysis for the determination of UHMWPE wear10.1002/jbm.b.30213 ↗
19. Effect of cross-linking on the microstructure and mechanical properties of ultra-high molecular weight polyethylene10.1097/01.blo.0000185310.59202.e5 ↗
20. Alpha-tocopherol-doped irradiated UHMWPE for high fatigue resistance and low wear10.1016/j.biomaterials.2003.12.048 ↗
21. Wear of fixed bearing and rotating platform mobile bearing knees subjected to high levels of internal and external tibial rotation10.1023/A:1012850224565 ↗
22. The influence of design, materials and kinematics on the in vitro wear of total knee replacements10.1016/j.jbiomech.2004.02.015 ↗
23. The influence of surgical malalignment on the contact pressures of fixed and mobile bearing knee prostheses—a biomechanical study10.1016/S0268-0033(02)00189-4 ↗
24. Mobile-bearing knees reduce rotational asymmetric wear10.1097/BLO.0b013e31806dba05 ↗
25. Micro-wear patterns on UHMWPE tibial inserts in total knee joint simulation10.1002/jbm.10027 ↗
26. International Standard ISO/FDIS 14243. Implants for surgery—wear of total knee-joint-prostheses—Part 1: Loading and displacement parameters for wear-testing machines with load control and corresponding enviromental conditions for tests, 2002.
27. International Standard ISO/FDIS 14243. Implants for surgery—wear of total knee-joint-prostheses—Part 2: Methods of measurement, 2000.
28. Comparison of three joint simulator wear debris isolation techniques: acid digestion, base digestion, and enzyme cleavage10.1002/1097-4636(200108)56:2<245::AID-JBM1091>3.0.CO;2-T ↗
29. International Standard ISO/FDIS 17853. Wear of implant particles—Polymer and metal wear particles—Isolation, characterization and quantification, 2002.
30. Morphological characteristics of total joint arthroplasty-derived ultra-high molecular weight polyethylene (UHMWPE) wear debris that provoke inflammation in a murine model of inflammation10.1002/jbm.a.10368 ↗
31. Investigation of wear of knee prostheses in a new displacement/force-controlled simulator10.1243/0954411021536289 ↗
32. Quantitative characterization of polyethylene debris isolated from periprosthetic tissue in early failure knee implants and early and late failure Charnley hip implants10.1002/jbm.1036 ↗
33. Knee-simulator testing of conventional and cross-linked polyethylene tibial inserts10.1016/j.arth.2004.03.019 ↗
34. Investigation on wear of knee prostheses under fixed kinematic conditions10.1111/j.1525-1594.2007.00455.x ↗
35. Shapes and dimensional characteristics of polyethylene wear particles generated in vivo by total knee replacements compared to total hip replacements10.1002/(SICI)1097-4636(199723)38:3<203::AID-JBM4>3.0.CO;2-T ↗
36. Wear of polyethylene in total joint replacements. Observations from retrieved PCA knee implants10.1097/00003086-199203000-00016 ↗
37. Transmission electron microscopy of intracellular particles of polyethylene from joint replacement prostheses: size distribution and cellular response10.1016/S0142-9612(01)00027-8 ↗
38. Wear of polyethylene cups in THR: Analysis of 159 cups retrieved at revision surgery or autopsy
39. Wear particle analysis of highly crosslinked polyethylene isolated from a failed total hip arthroplasty10.1002/jbm.b.31048 ↗
40. Characterization of a highly cross-linked ultrahigh molecular-weight polyethylene in clinical use in total hip arthroplasty10.1016/j.arth.2007.06.006 ↗
41. Sequentially crosslinked and annealed UHMWPE CR knee wear
42. Do current wear particle isolation procedures underestimate the number of particles generated by prosthetic bearing components?10.1016/S0043-1648(01)00762-1 ↗
43. Characteristics of polyethylene wear particles isolated from synovial fluid after mobile-bearing and posterior-stabilized total knee arthroplasties10.1002/jbm.b.30005 ↗
44. Evaluation of the response of primary human peripheral blood mononuclear phagocytes to challenge with in vitro generated clinically relevant UHMWPE particles of known size and dose10.1002/1097-4636(200011)52:2<296::AID-JBM8>3.0.CO;2-9 ↗
45. Comparison of the response of primary murine peritoneal macrophages and the U937 human histiocytic cell line to challenge with in vitro generated clinically relevant UHMWPE particles
46. Biological reactions to wear debris in total joint replacement10.1243/0954411001535219 ↗
47. Diverse cellular and apoptotic responses to variant shapes of UHMWPE particles in a murine model of inflammation10.1016/S0142-9612(02)00032-7 ↗
48. Tissue reactions to joint prostheses and the products of wear and corrosion
49. Average and peak contact stress distribution evaluation of total knee arthroplasties10.1016/S0883-5403(96)80137-9 ↗
50. Effect of size and dose on bone resorption activity of macrophages by in vitro clinically relevant ultra high molecular weight polyethylene particles10.1002/1097-4636(200009)53:5<490::AID-JBM7>3.0.CO;2-7 ↗
51. Biological reaction to debris in relation to joint prostheses10.1243/0954411971534304 ↗
52. Wear particle analyses of conventional and crosslinked UHMWPE tested in an anatomic hip simulator
53. Highly crosslinked polyethylene is safe for use in total knee arthroplasty10.1007/s11999-008-0472-4 ↗