journal article·2016-11

Visual H2 sensor for monitoring biodegradation of magnesium implants in vivo

Daoli Zhao, Tingting Wang, William Hoagland, David Benson, Zhongyun Dong, Shuna Chen, Da-Tren Chou, Daeho Hong, Jingyao Wu, Prashant N. Kumta, William R. Heineman

Acta Biomaterialia· Volume 45
Publisher page

Abstract

A visual sensor for H2 was used to transdermally monitor H2 that originated from biodegrading magnesium (Mg) alloys implanted subcutaneously in mice. The visual sensor consisted of a thin film of H2-sensitive material (MoO3 and Pd catalyst) coated on a flexible plastic sheet that was pressed against the mouse skin directly above the implant. Although the H2 levels permeating through the skin during the degradation process were very low, the sensor changed color to give a three dimensional (3D) visualization of H2 permeation. The correlation between the visual sensor response and measurements made with an electrochemical H2 microsensor on several magnesium alloys demonstrates that the visual sensor has the capability to monitor in real-time the dissolution rate of implants in vivo. This detection method is noninvasive, easy to implement, effective and potentially low cost compared to electrochemical detection. Statement of significance Biodegradable Mg implants offer advantages over permanent implants such as stainless steel that are used for broken bone repair. Mg alloys gradually dissolve, avoiding the need for removal by a later surgery if complications arise. Here we report a visual H2 sensor that can be used in the research laboratory to monitor the corrosion process in vivo during animal testing of different Mg alloys. The sensor consists of a plastic sheet with a thin coating that changes color in the presence of H2 gas. The sensor is easily used by taping it on the skin over the Mg implant. The color change gives a map of the H2 level permeating from the degrading Mg through the skin above it. This low cost, simple method of monitoring the dissolution of biodegradable implants would greatly facilitate the development of the biodegradable materials, especially in animal studies where in vivo biodegradation is tested.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Magnesium Alloys: Properties and Applications · Hydrogen Storage and Materials · Corrosion Behavior and Inhibition

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (43)

Show source-provided references
1. Magnesium and its alloys as orthopedic biomaterials: a review10.1016/j.biomaterials.2005.10.003 ↗
2. Recent metallic materials for biomedical applications10.1007/s11661-002-0109-2 ↗
3. Stress shielding and bone resorption in shoulder arthroplasty10.1067/mse.2003.22 ↗
4. The response of cancellous and cortical canine bone to hydroxylapatite-coated and uncoated titanium rods10.1002/jab.770060404 ↗
5. Variation in cytokines induced by particles from different prosthetic materials
6. Four-year study of cobalt and chromium blood levels in patients managed with two different metal-on-metal total hip replacements10.1016/S0736-0266(02)00152-3 ↗
7. Titanium, chromium and cobalt ions modulate the release of bone-associated cytokines by human monocytes/macrophages in vitro10.1016/0142-9612(96)00072-5 ↗
8. Metal ions induce bone-resorbing cytokine production through the redox pathway in synoviocytes and bone marrow macrophages10.1016/S0142-9612(02)00531-8 ↗
9. Biodegradability engineering of biodegradable Mg alloys: tailoring the electrochemical properties and microstructure of constituent phases10.1038/srep02367 ↗
10. Microstructure, mechanical properties and bio-corrosion evaluation of biodegradable AZ91-FA nanocomposites for biomedical applications10.1016/j.msea.2010.07.063 ↗
11. The microstructure, mechanical and corrosion properties of calcium polyphosphate reinforced ZK60A magnesium alloy composites10.1016/j.jallcom.2010.06.013 ↗
12. Formation by ion plating of Ti-coating on pure Mg for biomedical applications10.1016/j.scriptamat.2005.05.009 ↗
13. Corrosion mechanisms of magnesium alloys10.1002/(SICI)1527-2648(199909)1:1<11::AID-ADEM11>3.0.CO;2-N ↗
14. Recent progress in corrosion and protection of magnesium alloys10.1002/adem.200500013 ↗
15. Degradable biomaterials based on magnesium corrosion10.1016/j.cossms.2009.04.001 ↗
16. Understanding magnesium corrosion – a framework for improved alloy performance10.1002/adem.200310405 ↗
17. In vitro and in vivo corrosion measurements of magnesium alloys10.1016/j.biomaterials.2005.07.037 ↗
18. Control of biodegradation of biocompatable magnesium alloys10.1016/j.corsci.2007.01.001 ↗
19. A possible biodegradable magnesium implant material10.1002/adem.200600252 ↗
20. Influence of different surface machining treatments of magnesium-based resorbable implants on the degradation behavior in rabbits
21. The effects of magnesium particles in posterolateral spinal fusion: an experimental in vivo study in a sheep model10.3171/spi.2007.6.2.141 ↗
22. Fast escape of hydrogen from gas cavities around corroding magnesium implants10.1016/j.actbio.2012.10.008 ↗
23. In vivo corrosion of four magnesium alloys and the associated bone response10.1016/j.biomaterials.2004.09.049 ↗
24. Corrosion of magnesium10.1179/imr.1993.38.3.138 ↗
25. Corrosion studies of rapidly solidified magnesium alloys10.1149/1.2086455 ↗
26. In vivo corrosion and corrosion protection of magnesium alloy LAE44210.1016/j.actbio.2009.10.012 ↗
27. Measurement of the corrosion rate of magnesium alloys using Tafel extrapolation10.1016/j.corsci.2009.10.016 ↗
28. Distinct functions for Wnt/beta-Catenin in hair follicle stem cell proliferation and survival and interfollicular epidermal homeostasis10.1016/j.stem.2013.10.003 ↗
29. GDNF stimulates the proliferation of cultured mouse immature Sertoli cells via its receptor subunit NCAM and ERK1/2 signaling pathway10.1186/1471-2121-11-78 ↗
30. Monitoring biodegradation of magnesium implants with sensors10.1007/s11837-015-1775-z ↗
31. In vivo monitoring the biodegradation of magnesium alloys with an electrochemical H2 sensor10.1016/j.actbio.2016.03.039 ↗
32. W. Hoagland, Hydrogen Gas Indicator System, U.S. patent 6895805, USA, 2005.
33. In vitro degradation and cytotoxicity response of Mg-4% Zn-0.5% Zr (ZK40) alloy as a potential biodegradable material10.1016/j.actbio.2013.07.001 ↗
34. In vitro and in vivo corrosion, cytocompatibility and mechanical properties of biodegradable Mg-Y-Ca-Zr alloys as implant materials10.1016/j.actbio.2013.06.025 ↗
35. UNISENSE, Hydrogen Sensor User Manual, 2014.
36. A convenient method for determining the concentration of hydrogen in water: use of methylene blue with colloidal platinum10.1186/2045-9912-2-1 ↗
37. A system for characterizing Mg corrosion in aqueous solutions using electrochemical sensors and impedance spectroscopy10.1016/j.actbio.2013.07.011 ↗
38. Comparative study on MoO3 and HxMoO3 nanobelts: structure and electric transport10.1021/cm702942y ↗
39. Concerning the origin of superstructures in hydrogen molybdenum bronzes HxMoO310.1016/j.ssi.2002.11.002 ↗
40. In Situ Raman spectroscopy of H2 gas interaction with layered MoO310.1021/jp202123a ↗
41. Brimblecombe · 1996
42. Molecular hydrogen alleviates nephrotoxicity induced by an anti-cancer drug cisplatin without compromising anti-tumor activity in mice10.1007/s00280-008-0924-2 ↗
43. Advances in H2 sensors for bioanalytical applications10.1016/j.trac.2016.01.015 ↗