journal article·2020-09

In vitro fretting crevice corrosion damage of CoCrMo alloys in phosphate buffered saline: Debris generation, chemistry and distribution

Dongkai Zhu, Yangping Liu, Jeremy L. Gilbert

Acta Biomaterialia· Volume 114
Publisher page

Abstract

Fretting crevice corrosion in modular tapers of total hip replacements has become a major concern in orthopedic medical devices. Solid and ionic debris arising from fretting crevice corrosion have been implicated in device failure and revision surgery. This study aims to use a 2D pin-on-disk fretting corrosion test system to visualize damage progression and debris generation during fretting corrosion of CoCrMo alloys in phosphate buffered saline (PBS). The results provide direct evidence of rapid debris generation during fretting corrosion (after only 12 min of testing). Debris was generated and either extruded from the contact region or impacted into adjacent crevice sites as long as fretting continued. After testing, the fretting region consisted of a damaged and plastically deformed contact region surrounded by a halo of fretting debris consisting entirely of oxides and phosphates within the crevice region. Evidence of pitting corrosion and grain boundary corrosion was observed. Solid debris consisted of chromium (Cr), phosphate (P) and oxygen (O). X-ray photoelectron spectroscopy analysis of the near-fretted metal surface area showed a thicker oxygen (O1s) containing film with the depth profile of O1s above 10% penetrating up to 5.75 nm while the O1s concentration on the unfretted area fell to below 10% after 1 nm depth. Ion concentration in the PBS, measured using inductively coupled mass spectrometry, showed cobalt (Co) ions were most prevalent (1.46 ppm) compared to chromium (Cr) (0.07 ppm) and molybdenum (Mo) (0.05 ppm) (p <0.05). All of these results are consistent with the analysis of in vivo modular taper corrosion processes. STATEMENT OF SIGNIFICANCE: CoCrMo alloys has been widely used as a metallic biomaterial for implant devices and can lose their durability and reliability due to wear, corrosion and tribocorrosion. Debris, as one of the major products of these reactions, is associated with implant device failure. In the first time, we developed a fretting corrosion testing system to visualize the debris generation process in real-time between CoCrMo alloy pin and disk samples. Debris was generated rapidly during fretting corrosion and some of the debris egressed from the crevice site while also accumulating within the crevice area as fretting continued. Our study opens a new method for future studies to advance understanding of debris generation processes during wear and tribocorrosion phenomenon.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Orthopedic Infections and Treatments · Advanced materials and composites

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (41)

Show source-provided references
1. Wear analysis of a retrieved hip implant with titanium nitride coating10.1016/S0883-5403(97)90164-9 ↗
2. Method of characterizing fretting and corrosion at the various taper connections of retrieved modular components from metal-on-metal total hip arthroplasty
3. Modular taper junction corrosion and failure: how to approach a recalled total hip arthroplasty implant10.1016/j.arth.2013.08.026 ↗
4. In vivo oxide‐induced stress corrosion cracking of Ti‐6Al‐4V in a neck–stem modular taper: emergent behavior in a new mechanism of in vivo corrosion10.1002/jbm.b.31943 ↗
5. In vivo severe corrosion and hydrogen embrittlement of retrieved modular body titanium alloy hip‐implants10.1002/jbm.b.31171 ↗
6. In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations. The effect of crevice, stress, motion, and alloy coupling10.1002/jbm.820271210 ↗
7. Differences in the fretting corrosion of metal–metal and ceramic–metal modular junctions of total hip replacements10.1016/S0736-0266(03)00186-4 ↗
8. Wear mapping of CoCrMo alloy in simulated bio-tribocorrosion conditions of a hip prosthesis bearing in calf serum solution10.1016/j.msec.2015.01.004 ↗
9. Fretting corrosion of Si3N4 vs CoCrMo femoral heads on Ti‐6Al‐V trunnions10.1002/jor.24681 ↗
10. Study on torsional fretting wear behavior of a ball-on-socket contact configuration simulating an artificial cervical disk10.1016/j.msec.2015.05.056 ↗
11. Contact damage failure analyses of fretting wear behavior of the metal stem titanium alloy–bone cement interface10.1016/j.jmbbm.2015.06.025 ↗
12. Processing, characterization and fretting wear of zinc oxide and silver nanoparticles reinforced ultra high molecular weight polyethylene biopolymer nanocomposite10.1007/s11837-015-1358-z ↗
13. A metallic biomaterial tribocorrosion model linking fretting mechanics, currents, and potentials: model development and experimental comparison
14. Influence of fretting regimes on the tribocorrosion behaviour of Ti6Al4V in 0.9wt.% sodium chloride solution10.1016/j.wear.2003.11.003 ↗
15. A tribo-electrochemical apparatus for in vitro investigation of fretting–corrosion of metallic implant materials10.1016/S0043-1648(02)00027-3 ↗
16. Fretting corrosion of CoCrMo and Ti6Al4V interfaces10.1016/j.biomaterials.2012.04.015 ↗
17. Fretting-corrosion of materials used as orthopaedic implants10.1016/j.wear.2004.11.027 ↗
18. Fretting corrosion damage of total hip prosthesis: friction coefficient and damage rate constant approach10.1016/j.triboint.2012.10.008 ↗
19. Effect of the environment on wear ranking and corrosion of biomedical CoCrMo alloys10.1007/s10856-010-4224-0 ↗
20. Interactive effects of albumin and phosphate ions on the corrosion of CoCrMo implant alloy10.1149/1.2764238 ↗
21. A new antibacterial Co-Cr-Mo-Cu alloy: preparation, biocorrosion, mechanical and antibacterial property10.1016/j.msec.2016.05.028 ↗
22. The effects of cobalt–chromium–molybdenum wear debris in vitro on serum cytokine profiles and T cell repertoire10.1016/j.biomaterials.2015.07.034 ↗
23. Characterization of solid products of corrosion generated by modular-head femoral stems of different designs and materials
24. Corrosion and fretting of a modular hip system: a retrieval analysis of 60 rejuvenate stems10.1016/j.arth.2015.03.010 ↗
25. Corrosion at the neck-stem junction as a cause of metal ion release and pseudotumour formation10.1302/0301-620X.94B7.29122 ↗
26. Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty10.2106/JBJS.K.01352 ↗
27. Zhu D.Development of a “2-d” test system for visualizing fretting corrosion: a study of the fretting corrosion behavior of Co-Cr-Mo alloy" (2016). Dissertations - ALL. 497. https://surface.syr.edu/etd/497
28. ASTM F1537-20, Standard Specification for Wrought Cobalt-28Chromium-6Molybdenum Alloys for Surgical Implants (UNS R31537, UNS R31538, and UNS R31539), ASTM International, West Conshohocken, PA, 2020, 3.
29. Simple isolation method for the bulk isolation of wear particles from metal on metal bearing surfaces generated in a hip simulator test10.1007/s10856-012-4573-y ↗
30. A screening method for bacterial corrosion of metals10.1016/0167-7012(88)90001-2 ↗
31. The localized corrosion of Al at engineered Cu islands10.1007/s11837-001-0085-9 ↗
32. The development of a small-scale wear test for CoCrMo specimens with human synovial fluid10.1016/j.biotri.2018.04.001 ↗
33. The clinical significance of metal ion release from cobalt-chromium metal-on-metal hip joint arthroplasty10.1243/09544119JEIM78 ↗
34. Metal ion release with large-diameter metal-on-metal hip arthroplasty10.1016/j.arth.2009.12.013 ↗
35. The composition of the surface passive film formed on CoCrMo alloy in simulated physiological solution10.1016/S0013-4686(03)00396-7 ↗
36. Biotribocorrosion—an appraisal of the time dependence of wear and corrosion interactions: II. Surface analysis10.1088/0022-3727/39/15/S11 ↗
37. Influence of grain boundaries on dissolution behavior of a biomedical CoCrMo alloy: in-situ electrochemical-optical, AFM and SEM/TEM studies10.1149/2.056209jes ↗
38. Grain boundary assisted crevice corrosion in CoCrMo alloys10.5006/2108 ↗
39. Intergranular pitting corrosion of CoCrMo biomedical implant alloy10.1002/jbm.b.33067 ↗
40. Effect of simulated inflammatory conditions and potential on dissolution and surface oxide of CoCrMo alloy: in situ electrochemical atomic force microscopy study10.1016/j.electacta.2017.12.151 ↗
41. Fretting initiated crevice corrosion of 316LVM stainless steel in physiological phosphate buffered saline: potential and cycles to initiation10.1016/j.actbio.2019.07.051 ↗