Inflammatory tissue response in human soft tissue is caused by a higher particle load near carbon fiber-reinforced PEEK compared to titanium plates
Source abstract
Abstract
Titanium as the leading implant material in locked plating is challenged by polymers such as carbon fiber-reinforced polyetheretherketone (CFR-PEEK), which became the focus of interest of researchers and manufacturers in recent years. However, data on human tissue response to these new implant materials are rare. Osteosynthesis plates and peri‑implant soft tissue samples of 16 healed proximal humerus fractures were examined (n = 8 CFR-PEEK, n = 8 titanium). Soft tissue was analyzed by immunohistochemistry and µCT. The entrapped foreign bodies were further examined for their material composition by FTIR. To gain insight into their origin and formation mechanism, explanted and new plates were evaluated by SEM, EDX, profilometry and HR-CT. In the peri‑implant soft tissue of the CFR-PEEK plates, an inflammatory tissue reaction was detected. Tissues contained foreign bodies, which could be identified as tantalum wires, carbon fiber fragments and PEEK particles. Titanium particles were also found in the peri‑implant soft tissue of the titanium plates but showed a less intense surrounding tissue inflammation in immunohistochemistry. The surface of explanted CFR-PEEK plates was rougher and showed exposed and broken carbon fibers as well as protruding and deformed tantalum wires, especially in used screw holes, whereas scratches were identified on the titanium plate surfaces. Particles were present in the peri‑implant soft tissue neighboring both implant materials and could be clearly assigned to the plate material. Particles from both plate materials caused detectable tissue inflammation, with more inflammatory cells found in soft tissue over CFR-PEEK plates than over titanium plates. STATEMENT OF SIGNIFICANCE: Osteosynthesis plates are ubiquitously used in various medical specialties for the reconstruction of bone fractures and defects and are therefore indispensable for trauma surgeons, ENT specialists and many others. The leading implant material are metals such as titanium, but recently implants made of polymers such as carbon fiber-reinforced polyetheretherketone (CFR-PEEK) have become increasingly popular. However, little is known about human tissue reaction and particle generation related to these new implant types. To clarify this question, 16 osteosynthesis plates (n = 8 titanium and n = 8 CFR-PEEK) and the overlying soft tissue were analyzed regarding particle occurrence and tissue inflammation. Tissue inflammation is clinically relevant for the development of scar tissue, which is discussed to cause movement restrictions and thus contributes significantly to patient outcome.
Source: openalex_inverted_index · Quality: baseline
Research context
OpenAlex topics: Orthopaedic implants and arthroplasty · Bone fractures and treatments · Total Knee Arthroplasty Outcomes
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
References (34)
Show source-provided references
1. Open reduction internal fixation vs non-operative management in proximal humerus fractures: a prospective, randomized controlled trial protocol10.1186/s12891-018-2223-3 ↗
2. [Treatment of proximal humeral fractures with the PHILOS angular stable plate. Presentation of 225 cases of dislocated fractures]10.1007/s00113-006-1165-7 ↗
3. Long-term functional outcomes (median 10 years) after locked plating for displaced fractures of the proximal humerus10.1016/j.jse.2013.11.009 ↗
4. Thomsen, M. and P. Thomas, [Compatibility and allergies of osteosynthesis materials]. Unfallchirurg, 2017. 120(2): p. 116–121.10.1007/s00113-016-0309-7 ↗
5. Immuno-inflammatory tissue reaction to stainless-steel and titanium plates used for internal fixation of long bones10.1016/S0142-9612(02)00312-5 ↗
6. Titanium and zirconium release from titanium- and zirconia implants in mini pig maxillae and their toxicity in vitro10.1016/j.dental.2020.01.013 ↗
7. PEEK biomaterials in trauma, orthopedic, and spinal implants10.1016/j.biomaterials.2007.07.013 ↗
8. The use of carbon-fiber-reinforced (CFR) PEEK material in orthopedic implants: a systematic review10.4137/CMAMD.S20354 ↗
9. Mixed modification of the surface microstructure and chemical state of polyetheretherketone to improve its antimicrobial activity, hydrophilicity, cell adhesion, and bone integration10.1021/acsbiomaterials.9b01148 ↗
10. Soft tissue response to dental implant closure caps made of either polyetheretherketone (PEEK) or titanium10.1111/clr.13487 ↗
11. Bone graft incorporation in radiographically successful human intervertebral body fusion cages10.1097/00007632-200112150-00025 ↗
12. Failure of a carbon fiber implant. A case report10.1097/00007632-199808150-00016 ↗
13. Debris of carbon-fibers originated from a CFRP (pEEK) wrist-plate triggered a destruent synovitis in human10.1007/s10856-015-5664-3 ↗
14. Simultaneous phase and amplitude extraction from a single defocused image of a homogeneous object10.1046/j.1365-2818.2002.01010.x ↗
15. Optimizing and applying high-resolution, in-line laboratory phase-contrast X-ray imaging for low-density material samples10.1111/jmi.12986 ↗
16. Practical cone-beam algorithm10.1364/JOSAA.1.000612 ↗
17. Surface polishing positively influences ease of plate and screw removal10.22203/eCM.v019a12 ↗
18. Influence of plate material and screw design on stiffness and ultimate load of locked plating in osteoporotic proximal humeral fractures10.1016/j.injury.2016.01.004 ↗
19. PEEK versus titanium locking plates for proximal humerus fracture fixation: a comparative biomechanical study in two- and three-part fractures10.1007/s00402-016-2620-8 ↗
20. Hardware removal: indications and expectations10.5435/00124635-200602000-00006 ↗
21. Computational contact pressure prediction of CoCrMo, SS 316L and Ti6Al4V femoral head against UHMWPE acetabular cup under gait cycle10.3390/jfb13020064 ↗
22. Isolation and characterisation of wear debris surrounding failed total ankle replacements10.1016/j.actbio.2023.01.051 ↗
23. Pre-clinical studies to validate the MITCH PCR Cup: a flexible and anatomically shaped acetabular component with novel bearing characteristics10.1007/s10856-007-3256-6 ↗
24. Macrophage polarization by titanium dioxide (TiO2) particles: size matters10.1021/acsbiomaterials.6b00006 ↗
25. Amorphous and crystalline polyetheretherketone: mechanical properties and tissue reactions during a 3-year follow-up10.1002/jbm.a.31310 ↗
26. Immune cell response to orthopedic and craniofacial biomaterials depends on biomaterial composition10.1016/j.actbio.2023.03.007 ↗
27. Macrophage polarization in response to wear particles in vitro10.1038/cmi.2013.39 ↗
28. Tauro, J.A., R. L., The shoulder. AANA Advanced Arthroscopy Series, ed. R.K.N. Ryu 2010: Elsevier. 10.
29. Effects of wear particles of polyether-ether-ketone and cobalt-chromium-molybdenum on CD4- and CD8-T-cell responses10.18632/oncotarget.23757 ↗
30. Surface analysis of titanium maxillofacial plates and screws retrieved from patients10.1054/ijom.2002.0283 ↗
31. Metallic fragments on the surface of miniplates and screws before insertion10.1054/bjom.1998.0264 ↗
32. Influences of tantalum pentoxide and surface coarsening on surface roughness, hydrophilicity, surface energy, protein adsorption and cell responses to PEEK based biocomposite10.1016/j.colsurfb.2018.10.081 ↗
33. Finite element investigation into the use of carbon fibre reinforced PEEK laminated composites for distal radius fracture fixation implants10.1016/j.medengphy.2019.03.006 ↗
34. Optimizing surface characteristics for cell adhesion and proliferation on titanium plasma spray coatings on polyetheretherketone10.1016/j.spinee.2016.05.017 ↗