journal article·2010-03

Osmotic cavitation of elastomeric intraocular lenses

David M. Saylor, D. Coleman Richardson, Benita J. Dair, Steven K. Pollack

Acta Biomaterialia· Volume 6· Issue 3
Publisher page

Abstract

In recent years, traditional rigid materials have been replaced with softer elastomers in intraocular lenses to minimize the size of the required surgical incision, thereby reducing patient recuperation time. However, water-filled cavities that may impact visual acuity are found in many of these new implants. We demonstrate that the cavitation observed in vivo can occur due to an osmotic pressure difference between the aqueous solution within the cavity and the external media in which the lens is immersed. By reducing the osmolarity of the external solution from 300 to 0mM, we observe an increase in cavity volume of almost a factor of 30. Further, we have developed a model for cavity growth assuming the controlling factor is diffusion of hydrophilic molecules from the polymer matrix into the cavity. We find that the experimental observations are consistent with the model and suggest that oligomeric species generated during polymerization are responsible for the observed cavitation.
Source: europepmc · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Intraocular Surgery and Lenses · Corneal surgery and disorders · Ophthalmology and Visual Impairment Studies

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (40)

Show source-provided references
1. Practice styles and preferences of ASCRS members – 1994 survey10.1016/S0886-3350(13)80524-X ↗
2. Tadini, the man who invented the artificial lens10.1016/S0146-2776(79)80009-9 ↗
3. Results of intraocular lens insertion
4. The effect of polymethylmethacrylate, silicone, and polyacrylic intraocular lenses on posterior capsular opacification 3 years after cataract surgery10.1016/S0161-6420(99)90047-7 ↗
5. Thirty years of intraocular lens implantation: the way it was and the way it is10.1016/S0886-3350(99)80028-5 ↗
6. 12 year results of a prospective trial comparing poly(methyl methacrylate) and poly(hydroxyethyl methacrylate) intraocular lenses10.1016/S0886-3350(99)00199-6 ↗
7. Alcon AcrySof acrylic intraocular lens
8. SI-18 three-piece foldable silicone IOL for small-incision cataract surgery
9. Multicenter trial of an intraocular hydrogel lens implant10.1016/S0886-3350(87)80150-5 ↗
10. Keratometric cylinder and visual performance following phacoemulsification and implantation with silicone small-incision or poly(methyl methacrylate) intraocular lenses10.1016/S0886-3350(13)80981-9 ↗
11. Effect of small incision intraocular lens surgery on postoperative inflammation and astigmatism: a study of the amo si-18nb small incision lens10.1016/S0886-3350(13)80383-5 ↗
12. Effect of incision size on early postoperative visual rehabilitation after cataract surgery and intraocular lens implantation10.1016/S0886-3350(13)80613-X ↗
13. Visual significance of glistenings seen in the acrysof intraocular lens10.1016/S0886-3350(96)80041-1 ↗
14. Optical and atomic force microscopy of an explanted acrysof intraocular lens with glistenings10.1016/S0886-3350(99)00342-9 ↗
15. Clinical and experimental observation of glistening in acrylic intraocular lenses10.1016/S0021-5155(01)00429-4 ↗
16. Effects on visual function of glistenings and folding marks in acrysof intraocular lenses10.1016/S0886-3350(01)00995-6 ↗
17. Influence of glistenings on the optical quality of acrylic foldable intraocular lens10.1136/bjo.85.9.1034 ↗
18. Transient fogging of acrylic (acrysof) intraocular lenses10.1038/eye.1999.184 ↗
19. Post cataract-intraocular lens (iol) surgery opacification10.1038/sj.eye.6700066 ↗
20. Equilibrium water content and glistenings10.1016/j.jcrs.2003.12.038 ↗
21. Water damage in polyester resins10.1098/rspa.1967.0179 ↗
22. Microcavity formation in engineering polymers exposed to hot water10.1002/app.1983.070280920 ↗
23. A new mechanism of failure in polymers10.1002/pol.1974.130120204 ↗
24. Osmotic effects in water absorption by polymers10.1016/0032-3861(80)90062-2 ↗
25. Cracking in a glassy epoxy resin induced by water absorption10.1016/0032-3861(80)90332-8 ↗
26. Adsorption and desorption of water in rubbers10.1016/0032-3861(87)90193-5 ↗
27. Osmotic cracking in unsaturated polyester matrices under humid environments10.1002/1097-4628(20010401)79:14<2517::AID-APP1060>3.0.CO;2-P ↗
28. Image analysis of implanted rigid and foldable intraocular lenses in human eyes using Scheimpflug photography10.1159/000055524 ↗
29. Dieter · 1986
30. Determining the gaussian probability distribution of the best-fit ellipsoid of revolution for a polymer chain from planar projections10.1088/0266-5611/19/3/315 ↗
31. Hill · 1950
32. Mechanics and mechanism of environmental crazing in a polymeric glass10.1016/0032-3861(72)90101-2 ↗
34. The kinetics of precipitation from supersaturated solid solutions10.1016/0022-3697(61)90054-3 ↗
35. Freeman C, Schlueter D. Foldable ophthalmic and otorhinolaryngological device materials. US Patent 6528602; 2003.
36. In vitro comparison of glistening formation among hydrophobic acrylic intraocular lenses10.1016/S0886-3350(02)01224-5 ↗
37. Ultrastructural analysis of opacities seen in a hydrophilic acrylic intraocular lens10.1038/eye.2001.204 ↗
38. Surface scattering in implanted hydrophobic intraocular lenses10.1016/S0886-3350(02)01994-6 ↗
39. Moad · 1995
40. NMR measurement of small-molecule diffusion in PVA hydrogels: a comparison of convex and standard pgse methods10.62721/diffusion-fundamentals.4.34 ↗