journal article·2019-10Open access

Improving our understanding of metal implant failures: Multiscale chemical imaging of exogenous metals in ex-vivo biological tissues

Alexander P. Morrell, Hayley Floyd, J. Frederick W. Mosselmans, Liam M. Grover, Hiram Castillo-Michel, Edward T. Davis, Julia E. Parker, Richard A. Martin, Owen Addison

Acta Biomaterialia· Volume 98

Abstract

Biological exposures to micro- and nano-scale exogenous metal particles generated as a consequence of in-service degradation of orthopaedic prosthetics can result in severe adverse tissues reactions. However, individual reactions are highly variable and are not easily predicted, due to in part a lack of understanding of the speciation of the metal-stimuli which dictates cellular interactions and toxicity. Investigating the chemistry of implant derived metallic particles in biological tissue samples is complicated by small feature sizes, low concentrations and often a heterogeneous speciation and distribution. These challenges were addressed by developing a multi-scale two-dimensional X-ray absorption spectroscopic (XAS) mapping approach to discriminate sub-micron changes in particulate chemistry within ex-vivo tissues associated with failed CoCrMo total hip replacements (THRs). As a result, in the context of THRs, we demonstrate much greater variation in Cr chemistry within tissues compared with previous reports. Cr compounds including phosphate, hydroxide, oxide, metal and organic complexes were observed and correlated with Co and Mo distributions. This variability may help explain the lack of agreement between biological responses observed in experimental exposure models and clinical outcomes. The multi-scale 2D XAS mapping approach presents an essential tool in discriminating the chemistry in dilute biological systems where speciation heterogeneity is expected. SIGNIFICANCE: Metal implants are routinely used in healthcare but may fail following degradation in the body. Although specific implants can be identified as 'high-risk', our analysis of failures is limited by a lack of understanding of the chemistry of implant metals within the peri-prosthetic milieu. A new approach to identify the speciation and variability in speciation at sub-micron resolution, of dilute exogenous metals within biological tissues is reported; applied to understanding the failure of metallic (CoCrMo) total-hip-replacements widely used in orthopedic surgery. Much greater variation in Cr chemistry was observed compared with previous reports and included phosphate, hydroxide, oxide, metal and organic complexes. This variability may explain lack of agreement between biological responses observed in experimental exposure models and clinical outcomes.
Source: openalex_inverted_index · Quality: baseline

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Advanced X-ray and CT Imaging · Radioactive element chemistry and processing

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (64)

Show source-provided references
1. Exposure-related health effects of silver and silver compounds: a review
2. Nanoparticles: a review of particle toxicology following inhalation exposure10.3109/08958378.2010.642021 ↗
3. An overview of the abundance, relative mobility, bioavailability, and human toxicity of metals
4. Orthopaedic metals and their potential toxicity in the arthroplasty patient: a review of current knowledge and future strategies10.1302/0301-620X.89B5.18903 ↗
5. Synchrotron-based ν-XRF mapping and μ-FTIR microscopy enable to look into the fate and effects of tattoo pigments in human skin10.1038/s41598-017-11721-z ↗
6. Gold nanoparticles: interesting optical properties and recent applications in cancer diagnostics and therapy10.2217/17435889.2.5.681 ↗
7. Human exposure to toxic metals: factors influencing interpretation of biomonitoring results10.1016/0048-9697(95)04478-J ↗
8. Human exposure to toxic metals: factors influencing interpretation of biomonitoring results10.1016/j.biomaterials.2007.02.034 ↗
9. Surface characteristics, copper release, and toxicity of nano-and micrometer-sized copper and copper (II) oxide particles: a cross-disciplinary study10.1002/smll.200801220 ↗
10. Metal wear particles: What we know, what we do not know, and why
11. Metal in total hip arthroplasty: wear particles, biology, and diagnosis10.3928/01477447-20160719-06 ↗
12. Comparative study of cytotoxicity, oxidative stress and genotoxicity induced by four typical nanomaterials: the role of particle size, shape and composition10.1002/jat.1385 ↗
13. Comparison of the cytotoxicity of clinically relevant cobalt–chromium and alumina ceramic wear particles in vitro10.1016/S0142-9612(02)00360-5 ↗
14. Surface charge-dependent toxicity of silver nanoparticles10.1021/es1034188 ↗
15. Concentration-and composition-dependent effects of metal ions on human MG-63 osteoblasts10.1002/jbm.10106 ↗
16. Metal ion release from metal implants10.1016/j.msec.2004.08.018 ↗
17. The operation of the century: total hip replacement10.1016/S0140-6736(07)60457-7 ↗
18. Revision for adverse local tissue reaction following metal-on-polyethylene total hip arthroplasty is associated with a high risk of early major complications10.1302/0301-620X.100B6.BJJ-2017-1466.R1 ↗
19. Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty10.2106/JBJS.K.01352 ↗
20. Elevated intra-articular cobalt and chromium levels in mechanically assisted crevice corrosion in metal-on-polyethylene total hip arthroplasty10.1016/j.arth.2016.11.056 ↗
21. Speciation in metal toxicity and metal-based therapeutics10.3390/toxics3020170 ↗
22. IARC Monographs on the Evaluation of Carcinogenic Risks to Humans. 1990. 49, 49.
23. P. Apostoli, J. D. Cornelis, P. Hoet, D. Lison, D. Templeton, World Health Organization, Environmental Health Criteria, 234. 2006.
24. Role of the crystalline form of titanium dioxide nanoparticles: Rutile, and not anatase, induces toxic effects in Balb/3T3 mouse fibroblasts10.1016/j.tiv.2015.11.005 ↗
25. J. H. Kim, H. J. Gibb, P. D. Howe, World Health Organization, Concise International Chemical Assessment Document, 69. 2006.
26. Cobalt chloride speciation, mechanisms of cytotoxicity on human pulmonary cells, and synergistic toxicity with zinc10.1039/c3mt20196a ↗
27. Mapping the surface adsorption forces of nanomaterials in biological systems10.1021/nn203303c ↗
28. Nano-analyses of wear particles from metal-on-metal and non-metal-on-metal dual modular neck hip arthroplasty10.1016/j.nano.2016.11.003 ↗
29. Lymphoid aggregates that resemble tertiary lymphoid organs define a specific pathological subset in metal-on-metal hip replacements10.1371/journal.pone.0063470 ↗
30. Trace metal imaging with high spatial resolution: applications in biomedicine10.1039/C0MT00048E ↗
31. Imaging and speciation of trace elements in biological environment10.1016/j.biochi.2006.10.003 ↗
32. Bio-metals imaging and speciation in cells using proton and synchrotron radiation X-ray microspectroscopy10.1098/rsif.2009.0166.focus ↗
33. Do ‘passive’medical titanium surfaces deteriorate in service in the absence of wear?10.1098/rsif.2012.0438 ↗
34. Fundamentals of XAFS10.2138/rmg.2014.78.2 ↗
35. The chemical form of metallic debris in tissues surrounding metal-on-metal hips with unexplained failure10.1016/j.actbio.2010.06.006 ↗
36. Cobalt from metal-on-metal hip replacements may be the clinically relevant active agent responsible for periprosthetic tissue reactions10.1016/j.actbio.2012.05.003 ↗
37. Practical review on the use of synchrotron based micro-and nano-X-ray fluorescence mapping and X-ray absorption spectroscopy to investigate the interactions between plants and engineered nanomaterials10.1016/j.plaphy.2016.07.018 ↗
38. Speciation mapping of environmental samples using XANES imaging10.1071/EN13189 ↗
39. Fast XANES fluorescence imaging using a Maia detector10.1107/S1600577518004940 ↗
40. Overview of chemical imaging methods to address biological questions10.1016/j.micron.2016.02.005 ↗
41. Ong · 2014
42. Dynamically figured Kirkpatrick Baez X-ray microfocusing optics
43. I18–the microfocus spectroscopy beamline at the Diamond Light Source10.1107/S0909049509032282 ↗
44. The ID21 scanning X-ray microscope at ESRF
45. A multiplatform code for the analysis of energy-dispersive X-ray fluorescence spectra10.1016/j.sab.2006.12.002 ↗
46. M. D. Bethesda. ICRU Reports. International Commission on Radiation Units and Measurements, 1989, 44, 35.
47. MANTiS: a program for the analysis of X-ray spectromicroscopy data10.1107/S1600577514013964 ↗
48. SimpleElastix: A user-friendly, multi-lingual library for medical image registration
49. ATHENA, ARTEMIS, HEPHAESTUS: data analysis for X-ray absorption spectroscopy using IFEFFIT10.1107/S0909049505012719 ↗
50. Chemical speciation of nanoparticles surrounding metal-on-metal hips10.1039/c2cc33016d ↗
51. X-ray fluorescence mapping and micro-XANES spectroscopic characterization of exhaust particulates emitted from auto engines burning MMT-added gasoline10.1021/es0010958 ↗
52. The hard X-ray nanoprobe beamline at diamond-current status10.1017/S1431927618013569 ↗
53. Microfocus study of metal distribution and speciation in tissue extracted from revised metal on metal hip implants10.1088/1742-6596/190/1/012208 ↗
54. Quantification procedures in micro X-ray fluorescence analysis10.1016/S0584-8547(02)00281-1 ↗
55. Quantification estimate methods for synchrotron radiation X-ray fluorescence spectroscopy10.1016/j.nimb.2015.01.041 ↗
56. X-ray Spectroscopy
57. The fundamental algorithm: a natural extension of the Sherman equation, part I: theory
58. A general Monte Carlo simulation of energy-dispersive X-ray fluorescence spectrometers—Part 6. Quantification through iterative simulations10.1016/j.sab.2012.12.011 ↗
59. Synchrotron analysis of human organ tissue exposed to implant material10.1016/j.jtemb.2017.12.007 ↗
60. Understanding the reactivity of CoCrMo-implant wear particles10.1038/s41529-018-0029-2 ↗
61. Clinical usefulness of blood metal measurements to assess the failure of metal-on-metal hip implants10.1258/acb.2011.011141 ↗
62. Total cobalt determination in human blood and synovial fluid using inductively coupled plasma-mass spectrometry: method validation and evaluation of performance variables affecting metal hip implant patient samples10.1080/02772248.2015.1092735 ↗
63. Changes in the surface oxide composition of Co–Cr–Mo implant alloy by macrophage cells and their released reactive chemical species10.1016/j.biomaterials.2003.08.016 ↗
64. Surface chemistry and microstructure of metallic biomaterials for hip and knee endoprostheses10.1016/j.apsusc.2017.08.007 ↗