journal article·2021-03-01

Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices

Terry O. Woods, Jana G. Delfino, Sunder Rajan

Journal of Testing and Evaluation· Volume 49· Issue 2
Publisher page

Abstract

Abstract Millions of patients receive magnetic resonance imaging (MRI) scans each year, and thousands of healthcare professionals and other individuals also routinely enter the magnetic resonance (MR) environment. The electric and magnetic fields used by the MR system to create MR images produce hazards that must be considered to determine whether a medical device or other object may safely be used in or near an MR system. Magnetically induced deflection force and torque are two of the hazards that must be addressed to determine the MRI safety of any device or object that will be used in or near an MR system. This article presents magnetically induced force and torque test results for a variety of common medical implant alloys. For seven of the eleven tested alloys, magnetically induced force is less than the device weight in any current clinical 1.5, 3.0, and 7.0-T MR system. For the other alloys, the magnetically induced force is greater than the device weight in some MR systems. The torque test results for the tested alloys demonstrate that for static magnetic fields of 3.0 T or less, the magnetically induced torque will always be less than the worst case torque produced by gravity. These results may be leveraged in MRI safety assessments for medical devices and other objects composed of those alloys and conforming to the same ASTM specifications and may be used as part of the information needed to generate MR Conditional labeling for the devices.
Source: crossref · Quality: baseline

Research context

OpenAlex topics: Nuclear Physics and Applications · Non-Destructive Testing Techniques · Magnetic Properties of Alloys

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

Affiliations

Center for Devices and Radiological Health, Office of Science and Engineering Laboratories, U.S. Food and Drug Administration 1 , 10903 New Hampshire Ave., Silver Spring, MD 20993 ,

Center for Devices and Radiological Health, Office of In-Vitro Diagnostics and Radiological Health, U.S. Food and Drug Administration 2 , 10903 New Hampshire Ave., Silver Spring, MD 20993 ,

Center for Devices and Radiological Health, Office of Science and Engineering Laboratories, U.S. Food and Drug Administration 3 , 10903 New Hampshire Ave., Silver Spring, MD 20993 ,

References (36)

Show source-provided references
1. 2013
2. 2018
3. D. W. Chen, “Boy, 6, Dies of Skull Injury during M.R.I.,” New York Times, July 31, 2001, https://www.nytimes.com/2001/07/31/nyregion/boy-6-dies-of-skull-injury-during-mri.html
4. V. Doshi, “Indian Man Dies after Being Trapped in MRI Machine,” The Washington Post, January 29, 2018, https://www.washingtonpost.com/news/worldviews/wp/2018/01/29/indian-man-dies-after-being-trapped-in-mri-machine/?utm_term=.66d152ba45b2
5. U.S. Food & Drug Administration, “MAUDE Adverse Event Report: GE Medical Systems, LLC Signa HDX 1.5T,” Medical Device Reporting Database #1648230, 2010, https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfMAUDE/detail.cfm?mdrfoi__id=1648230%20
6. Placement of a Ferromagnetic Intracerebral Aneurysm Clip in a Magnetic Field with a Fatal Outcome10.1148/radiology.187.3.8497645 ↗
7. U.S. Food & Drug Administration, Medical Device Reporting Database #M175218, September 19, 1989, http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfMDR/Detail.CFM?ID=540787
8. U.S. Food & Drug Administration, “MAUDE Adverse Event Report: Flowonix Medical, Inc., Prometra Pump, Infusion, Implanted, Programmable,” Medical Device Reporting Database #5116276, 2015, https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfMAUDE/detail.cfm?mdrfoi__id=5116276
9. U.S. Food & Drug Administration, Medical Device Reporting Database #M297255, June 26, 1992, https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfMDR/Detail.cfm?ID=434943
10. Case Report: MRI Induced Fourth-Degree Burn in an Extremity, Leading to Amputation10.1016/j.burns.2007.11.008 ↗
11. U.S. Food & Drug Administration, “MAUDE Adverse Event Report: Siemens AG Magnetom Trio, A Tim System System, Nuclear Magnetic Resonance Imag.,” Medical Device Reporting Database #2240869-2014-02104, March 7, 2014, https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfMAUDE/detail.cfm?mdrfoi__id=3716520
12. U.S. Food & Drug Administration, “MAUDE Adverse Event Report: Philips Medical Systems Nederland B.V. Intera 1.5T,” Medical Device Reporting Database #3003768277-2013-00001, December 26, 2012, https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfMAUDE/detail.cfm?mdrfoi__id=2930476
13. U.S. Food & Drug Administration, “MAUDE Adverse Event Report: Philips Healthcare Panorama 1.0T,” Medical Device Reporting Database #3003768277-2011-00364, April 29, 2011, https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfMAUDE/detail.cfm?mdrfoi__id=2100792
14. U.S. Food & Drug Administration, “MAUDE Adverse Event Report: GE Medical Systems, LLC, HDI Echospeed 1.5T,” Medical Device Reporting Database #2183553-2010-00001, November 22, 2009, https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfMAUDE/detail.cfm?mdrfoi__id=1578579
15. Permanent Neurological Deficit Related to Magnetic Resonance Imaging in a Patient with Implanted Deep Brain Stimulation Electrodes for Parkinson’s Disease: Case Report10.1227/01.NEU.0000180810.16964.3E ↗
16. MRI Safety
17. 2015
18. Potential Hazards and Artifacts of Ferromagnetic and Nonferromagnetic Surgical and Dental Materials and Devices in Nuclear Magnetic Resonance Imaging10.1148/radiology.147.1.6828719 ↗
19. 2017
20. MRI and Implanted Medical Devices: Basic Interactions with an Emphasis on Heating10.1109/TDMR.2005.859033 ↗
21. 1985
22. Magnetic Displacement Force and Torque on Dental Keepers in the Static Magnetic Field of an MR Scanner10.1002/jmri.24500 ↗
23. MRI Safety and Compatibility of Implants and Medical Devices10.1520/STP11156S ↗
24. 2012
25. An Overview of Superelastic Stent Design
26. Magnetic Resonance Imaging of AO/ASIF Stainless Steel and Titanium Implants10.1016/S0020-1383(10)80001-4 ↗
27. The Effect of Mechanical Deformation on Magnetic Properties and MRI Artifacts of Type 304 and Type 316L Stainless Steel10.1002/jmri.1880070635 ↗
28. 2013
29. 2012
30. 2015
31. 2014
32. 2013
33. Magnetically Induced Force Measurements per ASTM F2052 of Active Implantable Medical Device Lead Materials
34. High-Field-Strength MR Imaging and Metallic Biomedical Implants: An Ex Vivo Evaluation of Deflection Forces10.2214/ajr.151.2.389 ↗
35. 2017
36. Solid Friction Appendix: Static and Kinetic Friction Coefficients for Selected Materials