A Gauge Study for Dynamic Light Scattering and Differential Mobility Analyzer Instruments
Source abstract
Abstract
Abstract This paper presents the performance of dynamic light scattering (DLS) and differential mobility analyzer (DMA) instruments in measuring nanoparticle size data using a gauge repeatability and reproducibility (GR&R) study based on the mixed model. The adequacy of the measurement process was evaluated using a measure of %GR&R with its confidence interval. The results show that the performance of the DMA instrument is superior to that of the DLS with regard to all variance components of a GR&R study. However, both instruments performed poorly in terms of %GR&R. Therefore, the part and measurement system require further improvement. Using these empirical study applications, this research promotes and maintains measurement quality through procedure analysis.
Source: crossref · Quality: baseline
Research context
OpenAlex topics: Surface Roughness and Optical Measurements · Tactile and Sensory Interactions
SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.
Affiliations
Dept. of Industrial Management, National Taiwan Univ. of Science and Technology 1 , No. 43 Keelung Rd., Sec. 4, Taipei106, TW
Bureau of Standards, Metrology and Inspection, Ministry of Economic Affairs (BSMI, MOEA) 2 , 4 Chinan Rd., Sec. 1, Taipei100, TW
Center for Measurement Standards, Industrial Technology Research Institute 3 , No. 321, Sec. 2, Kuang Fu Rd., Hsinchu300, TW
References (26)
Show source-provided references
1. Gauge Capability and Designed Experiments, Part I: Basic Methods10.1080/08982119308918710 ↗
2. Gauge Capability Analysis and Designed Experiments, Part II: Experimental Design Models and Variance Component Estimation10.1080/08982119308918725 ↗
3. Confidence Intervals on Measures of Variability in R&R Studies10.1080/00224065.1997.11979768 ↗
4. Analysis of a Two-factor R&R Study with Fixed Operators10.1080/00224065.1998.11979835 ↗
5. Measurement System Assessment via Generalized Inference10.1080/00224065.2000.11980000 ↗
6. A Simple General Method for Constructing Confidence Intervals for Functions of Variance Components10.1198/004017001316975943 ↗
7. Comparing Variability of Two Measurement Processes Using R&R Studies10.1080/00224065.2002.11980132 ↗
8. A Review of Methods for Measurement Systems Capability Analysis10.1080/00224065.2003.11980232 ↗
9. Confidence Intervals in a Gauge R&R Study with Fixed Operators10.1080/00224065.2005.11980319 ↗
10. Confidence Intervals for Unbalanced Two-factor Gauge R&R Studies10.1002/qre.682 ↗
11. Misclassification Rates in Gauge R&R Studies10.1080/00224065.2005.11980332 ↗
12. Burdick · 200510.1137/1.9780898718379 ↗
13. Modeling and Evaluating Repeatability and Reproducibility of Ordinal Classifications10.1198/TECH.2009.08052 ↗
14. On the Repeatability and Reproducibility of Experimental Two-chambered Microbial Fuel Cells10.1016/j.fuel.2009.04.026 ↗
15. An Assessment of the Robustness of Gauge Repeatability and Reproducibility Analysis in Automotive Components10.1177/0954407011401504 ↗
16. A Comparison Study on the Measurement of Nanoparticles
17. Preliminary Study on Nanoparticle Sizes under the APEC Technology Cooperative Framework10.1088/0957-0233/18/2/S23 ↗
18. A Measurement System Analysis Approach for Hard-to-Repeat Events10.1080/08982110902852344 ↗
19. Pitch Measurement by Traceable Atomic Force Microscope10.1142/S0219581X0300136X ↗
20. Form Measurements of Micro-holes10.1088/0957-0233/18/11/045 ↗
21. Quantitative Evaluation of Powder Spray Effects on Stereovision Measurements
22. A Review of Statistical Methods for Quality Improvement and Control in Nanotechnology10.1080/00224065.2009.11917770 ↗
23. Montgomery · 2005
24. Automotive Industry Action Group (AIAG) · 2002
25. Evaluating Measurement and Process Capabilities by GR&R with Four Quality Measures10.1016/j.measurement.2010.02.016 ↗
26. Two-dimensional Guidelines for Measurement System Indices10.1002/qre.1144 ↗