journal article·2005-01

Characterisation of thin layers of parylene at high frequency using PZT thick film resonators

Franck Levassort, Louis-Pascal Tran-Huu-Hue, Pierre Marechal, Erling Ringgaard, Marc Lethiecq

Publisher page

Abstract

Abstract High frequency electrical impedance measurements on PZT thick film structures were used to characterise thin layers of parylene for acoustic matching applications. The parylene properties (i.e. longitudinal wave velocity and acoustic impedance) were obtained in a realistic configuration for transducer applications. The measured parylene properties were compatible with medical imaging requirements, in particular for high frequency, since the deposited thickness can be controlled with high accuracy (around 1 μm). The mean longitudinal wave velocity was measured at 2135 ms −1 and corresponding acoustic impedance was 2.75 MRa. Three high frequency single element transducers were simulated to show that using parylene as a matching layer is a good trade-off between transducer performance and the technical process.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Neuroscience and Neural Engineering · Analytical Chemistry and Sensors · Electrical and Bioimpedance Tomography

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (8)

Show source-provided references
1. Polymer films as acoustic matching layers10.1109/ULTSYM.1990.171581 ↗
2. Dual layer matching (20MHz) piezoelectric transducer with glass and parylene10.1109/58.655643 ↗
3. Characterizing ultra-thin matching layers of high-frequency ultrasonic transducer based on impedance matching principle10.1109/TUFFC.2004.1320769 ↗
4. COMELEC_SA, Switzerland, La Chaux-de-Fonds. www.comelec.ch.
5. New equivalent circuit for elementary piezoelectric transducers
6. Onda Corporation Sunnycale, C., USA, Acoustical properties of plastics. www.ondacorp.com.
7. High performance piezoceramic films on substrates for high frequency imaging
8. Optimisation method for ultrasonic transducers used in medical imaging