journal article·2007-01

Factors influencing the reliability of ceramic femoral components

R. Morrell, S. Hughes

Publisher page

Abstract

The longevity of ceramic components implanted in the human body is controlled by a number of factors associated with their design, manufacture and interfacing with metallic components, as well as factors associated with the surgical implantation and patient use. While the technology can be considered to be mature, it requires maintenance of optimum materials and processes on the part of the ceramic manufacturer, as well as achieving optimum potential performance and failure risk reduction through design, commensurate with achieving practical goals. This paper reviews the issues of performance requirements, and the relevance of standardised tests for both the material and the components compared with in vivo loading. Long-term testing to determine resistance to static loading is described. Fractography is shown to be a key element for interpreting failure modes and mechanisms.
Source: openaire · Quality: candidate unverified · Source record ↗

Research context

OpenAlex topics: Orthopaedic implants and arthroplasty · Orthopedic Infections and Treatments · Dental materials and restorations

SciVX discovery tags are terminology-based. OpenAlex topics are source-provided classifications.

References (9)

Show source-provided references
1. A simulator study of friction in total replacement hip joints10.1243/PIME_PROC_1992_206_292_02 ↗
2. US Food and Drugs Administration, 510(k) Guidance Document for the Preparation of Pre-market Notifications For Ceramic Ball HIP Systems. [Website: http://www.fda.gov/cdrh/ode/355.html].
3. Axisymmetric mechanical analysis of ceramic heads for total hip replacement10.1243/PIME_PROC_1990_204_250_02 ↗
4. Fractography of ceramic femoral heads
5. Dragoni, E., Andrisano, A. and Bertuzzi, F., In Experimental Mechanics, Clinical aspects of biomedicine, Vol 3., ed. G. E. Little. Elsevier Science Publ. BV, Amsterdam, Netherlands, 1992, pp. 23–34, ISBN 0444905809.
6. Performance evaluation of zirconia femoral heads
7. Chevalier, J., Drouin, J. M. and Calès, B., Low-temperature ageing behaviour of zirconia hip joint heads. Bioceramics. Vol 10. In Proc. 10th. Int. Symposium on Ceramics in Medicine, Paris, France, October 1997, 1997, pp. 9–11.
8. Improved bearing systems for THP using zirconia ceramic
9. ASTM F2345-03, Standard test methods for determination of static and cyclic strength or fatigue of ceramic modular femoral heads.